Is the use of an additional pace/sense lead the optimal strategy for the avoidance of lead extraction in defibrillation lead failure? A single-centre experience

被引:7
|
作者
Scott, Paul A. [1 ]
Chungh, Aman [1 ]
Zeb, Mehmood [1 ]
Yue, Arthur M. [1 ]
Roberts, Paul R. [1 ]
Morgan, John M. [1 ]
机构
[1] Univ Southampton, Hosp NHS Trust, Wessex Cardiothorac Unit, Southampton SO16 6YD, Hants, England
来源
EUROPACE | 2010年 / 12卷 / 04期
关键词
Implantable cardioverter defibrillators; Lead; Failure; Complications; Defibrillation; PERFORMANCE; MANAGEMENT; DEFECTS;
D O I
10.1093/europace/eup406
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims The implantation of an additional pace-sense (P/S) lead is a standard treatment option in the management of an isolated pace-sense problem in a defibrillation (HV-P/S) lead. However, the safety of this management strategy is unclear. We performed a retrospective single-centre study to assess this. Methods and results We studied all patients with an isolated P/S problem in an HV-P/S lead, treated with an additional P/S lead, in our institution. The need for further invasive intervention for a lead-related complication, or death during follow-up, was assessed. From 2000 to 2008, 45 patients were treated with an additional P/S lead. Mean follow-up was 78 +/- 38 months from original device implantation and 28 +/- 17 months following implantation of the additional lead. During follow-up, three patients required an invasive intervention for a lead-related problem. All were successfully treated with lead extraction and device re-implantation. There were five deaths. Following implant of an additional lead, cumulative survival from further lead defects after 6 months, 1, 2, and 3 years was 100, 100, 93, and 87%, respectively. Conclusion In the treatment of an isolated P/S problem in an HV-P/S lead, the placement of an additional P/S lead is a safe management strategy, at least in the short term.
引用
收藏
页码:522 / 526
页数:5
相关论文
共 50 条
  • [1] PM/ICD lead extraction: A new single-centre experience
    Golzio, P. G.
    Bongiorni, M. G.
    Vinci, M.
    Trevi, G. P.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2007, 18 : S65 - S66
  • [2] Lead extraction for device related infections: a single-centre experience
    Bracke, FALE
    Meijer, A
    van Gelder, LM
    EUROPACE, 2004, 6 (03): : 243 - 247
  • [3] Transvenous lead extraction of pacing and defibrillator leads: a single-centre experience
    Geselle, P-J.
    Poesen, R.
    Rega, F.
    Koopman, P.
    Nuyens, D.
    Heidbuechel, H.
    Willems, R.
    ACTA CARDIOLOGICA, 2011, 66 (05) : 678 - 678
  • [4] Deep sedation for transvenous lead extraction: a large single-centre experience
    Bode, Kerstin
    Whittaker, Peter
    Lucas, Johannes
    Muessigbrodt, Andreas
    Hindricks, Gerhard
    Richter, Sergio
    Doering, Michael
    EUROPACE, 2019, 21 (08): : 1246 - 1253
  • [5] Lead extraction with high frequency laser sheaths: a single-centre experience
    Pecha, Simon
    Linder, Matthias
    Gosau, Nils
    Castro, Liesa
    Vogler, Julia
    Willems, Stephan
    Reichenspurner, Hermann
    Hakmi, Samer
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2017, 51 (05) : 902 - 905
  • [6] ENDOVASCULAR LEAD EXTRACTION : A SINGLE CENTRE EXPERIENCE
    Bontempi, L.
    Curnis, A.
    Cerini, M.
    Lipari, A.
    Pagnoni, C.
    Ashofair, N.
    Mutti, M. G.
    Latini, M. G.
    Vassanelli, F.
    Dei Cas, L.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 : S132 - S132
  • [7] A single-centre experience concerning the safety of Sprint Fidelis defibrillator lead extraction at the time of pulse generator replacement or in case of evidence of lead failure
    Da Costa, Antoine
    Da Cruz, Carine
    Romeyer-Bouchard, Cecile
    Abdellaoui, Loucif
    Nadrouss, Abdallah
    Bisch, Laurence
    Chometon, Frederic
    Afif, Zai
    Gate-Martinet, Alexie
    Combier, Michel
    Isaaz, Karl
    ARCHIVES OF CARDIOVASCULAR DISEASES, 2012, 105 (04) : 203 - 210
  • [8] Single sheath lead extraction, a single centre experience of more than 900 lead extractions
    Platou, E. S.
    Knutsen, T. M.
    Steen, T.
    EUROPEAN HEART JOURNAL, 2011, 32 : 306 - 306
  • [9] Efficacy and safety of transvenous lead extraction in 108 consecutive patients: a single-centre experience
    Monsefi, Nadejda
    Waraich, Harmeet Singh
    Vamos, Mate
    Erath, Julia
    Sirat, Sami
    Moritz, Anton
    Hohnloser, Stefan H.
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2019, 28 (05) : 704 - 708
  • [10] EXPERIENCE WITH THE SPRINT FIDELIS LEAD RECALL IN A SINGLE CANADIAN LASER LEAD EXTRACTION CENTRE
    Bashir, J.
    Yamashita, M.
    Karim, S. S.
    Tung, S.
    Yeung-Lai-Wah, J.
    Chakrabarti, S.
    Hahn, E.
    Sanatani, S.
    Kerr, C.
    CANADIAN JOURNAL OF CARDIOLOGY, 2011, 27 (05) : S281 - S282