Design of a conduction-cooled 9.4T REBCO magnet for whole-body MRI systems

被引:47
|
作者
Miyazaki, Hiroshi [1 ]
Iwai, Sadanori [1 ]
Otani, Yasumi [1 ]
Takahashi, Masahiko [1 ]
Tosaka, Taizo [1 ]
Tasaki, Kenji [1 ]
Nomura, Shunji [1 ]
Kurusu, Tsutomu [1 ]
Ueda, Hiroshi [2 ,4 ]
Noguchi, So [3 ,4 ]
Ishiyama, Atsushi [4 ]
Urayama, Shinichi [5 ]
Fukuyama, Hidenao [5 ]
机构
[1] Toshiba Co Ltd, Yokohama, Kanagawa 2300045, Japan
[2] Okayama Univ, Okayama 7708530, Japan
[3] Hokkaido Univ, Sapporo, Hokkaido 0600814, Japan
[4] Waseda Univ, Tokyo 1698555, Japan
[5] Kyoto Univ, Kyoto 6068507, Japan
来源
SUPERCONDUCTOR SCIENCE & TECHNOLOGY | 2016年 / 29卷 / 10期
关键词
ultra-high-field MRI; REBCO; conduction-cooled;
D O I
10.1088/0953-2048/29/10/104001
中图分类号
O59 [应用物理学];
学科分类号
摘要
A project on the development of REBa2Cu3O7-delta (REBCO) magnets for ultra-high-field magnetic resonance imaging (MRI) was started in 2013. Since REBCO-coated conductors feature high mechanical strength under tensile stress and high critical current density, use of REBCO coils would allow superconducting magnets to be made smaller and lighter than conventional ones. In addition, a conduction-cooled superconducting magnet is simpler to use than one cooled by a liquid helium bath because the operation and maintenance of the cryogenic system become simpler, without the need to handle cryogenic fluid. Superconducting magnets for MRI require homogeneous, stable magnetic fields. The homogeneity of the magnetic field is highly dependent on the coil shape and position. Moreover, in REBCO magnets, the screening-current- induced magnetic field, which changes the magnetic field distribution of the magnet, is one of the critical issues. In order to evaluate the magnetic field homogeneity and the screening-current- induced magnetic field, a 1 T model magnet and some test coils were fabricated. From an evaluation of the 1 T model magnet, it was found that the main reason for magnetic field inhomogeneity was the tolerances in the z-axis positions of the coils, and therefore, it is important to control the gap between the single pancakes. In addition, we have already demonstrated the generation of an 8.27 T central magnetic field at 10 K with a small test coil. The screening-current-induced magnetic field was 0.43 T and was predictable by using an electromagnetic field simulation program. These results were reflected in the design of a conduction-cooled 9.4 T REBCO magnet for whole-body MRI systems. The magnet was composed of six main coils and two active shield coils. The total conductor length was 581 km, and the stored energy was 293 kJ. The field inhomogeneity was 24 ppm peak to peak and 3 ppm volume-root-mean-square (VRMS) for a 500 mm diameter spherical volume (DSV). The axial and radial 5 gauss line locations were less than 5 m and 4.2 m respectively.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Progress of the 9.4-T Whole-Body MRI Superconducting Coils Manufacturing
    Cheng, Junsheng
    Li, Lankai
    Wang, Hui
    Li, Yi
    Sun, Wanshuo
    Chen, Shunzhong
    Zhao, Baozhi
    Zhu, Xucheng
    Wang, Lei
    Dai, Yinming
    Yan, Luguang
    Wang, Qiuliang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2018, 28 (04)
  • [22] Systematic analysis of the quench process performance and simulation of the 9.4 T-800 mm whole-body MRI magnet
    Chen, Shunzhong
    Wang, Lei
    Zhang, Zili
    Wang, Yaohui
    Wang, Hui
    Cheng, Junsheng
    Dai, Yinming
    Wang, Qiuliang
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2023, 36 (04):
  • [23] Design and simulation of 1.5 T conduction-cooled superconducting magnet with flip-over capability
    Wang, Donghu
    Qin, Jinggang
    Zhang, Zhan
    Li, Zuoguang
    Xue, Shengquan
    Li, Bo
    Ma, Shuliang
    Zhou, Chao
    FUSION ENGINEERING AND DESIGN, 2023, 193
  • [24] Design and Fabrication of a Conduction-Cooled Superconducting Magnet for Hall Sensor Calibration
    Hu, Xinning
    Dai, Yinming
    Wang, Hui
    Chen, Shunzhong
    Cheng, Junsheng
    Li, Yi
    Cui, Chunyan
    Li, Lankai
    Yan, Luguang
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2016, 26 (04)
  • [25] Quench characteristics of 6 T conduction-cooled NbTi magnet system
    Kar, S.
    Soni, V.
    Konduru, P.
    Sharma, R. G.
    Kanjilal, D.
    Datta, T. S.
    ADVANCES IN CRYOGENIC ENGINEERING, 2015, 101
  • [26] A Conceptual Design of a Conduction-Cooled Magnet for a Superconducting Property Measurement System
    Choi, Sukjin
    Lee, Seyeon
    Bae, Joon-Han
    Sohn, Myung-Hwan
    Kim, Woo-Seok
    Park, Chan
    Lee, Ji-Kwang
    Lee, Sangjin
    Choi, Kyeongdal
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2009, 19 (03) : 1249 - 1252
  • [27] Design of a Conduction-Cooled Open Gradient Superconducting Magnetic Separation Magnet
    Xiao, Mingming
    Du, Junjie
    Han, Zhengnan
    Chang, Haoran
    Guo, Wei
    Liu, Liming
    Wu, Wei
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2023, 33 (06)
  • [28] The Design of Decoupled Even-Order Zonal Superconducting Shim Coils for a 9.4 T Whole-Body MRI
    Zhu, Xuchen
    Wang, Qiuliang
    Li, Yi
    Hu, Yang
    Chen, Wenbo
    Liu, Feng
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2016, 26 (08)
  • [29] A Multiple Layers Superconducting Magnet Design for 9.4T Magnetic Resonance Imaging
    Zhang, Guoqing
    Zhu, Zian
    Zhao, Ling
    Hou, Zhilong
    Ning, Feipeng
    Yao, Weichao
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2014, 24 (03)
  • [30] Conduction cooled magnet design for 1.5T, 3.0T and 7.0T MRI systems
    Baig, Tanvir
    Yao, Zhen
    Doll, David
    Tomsic, Michael
    Martens, Michael
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2014, 27 (12):