Oxygen contamination of the nitrous oxide pipeline supply

被引:7
|
作者
Lye, A [1 ]
Patrick, R [1 ]
机构
[1] Royal Prince Alfred Hosp, Dept Anaesthet, Camperdown, NSW 2050, Australia
关键词
equipment; failure; nitrous oxide oxygen blender; Midogas; oxygen alarm;
D O I
10.1177/0310057X9802600215
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Midogas(R) (CIG Australia) nitrous oxide/oxygen blenders are commonly used in delivery wards in Australia. In this ease report we describe an incident where a small hole in the diaphragm of the oxygen failure alarm in a Midogas nitrous oxide/oxygen blender led to retrograde flow of oxygen contaminating the nitrous oxide pipeline supply to the operating theatres and the delivery ward. This caused a reduced level of nitrous oxide to the patient in theatre, but there was no adverse outcome, However if the oxygen pipeline pressure had been lower than the nitrous oxide pipeline pressure, the contamination would have been reversed potentially resulting in a hypoxic mixture being delivered to many patients, Suggestions are made as to a method of prevention of this problem.
引用
收藏
页码:207 / 209
页数:3
相关论文
共 50 条
  • [41] CONTAMINATION OF A HOSPITAL OXYGEN-SUPPLY
    GILMOUR, IJ
    MCCOMB, RC
    PALAHNIUK, RJ
    ANESTHESIA AND ANALGESIA, 1990, 71 (03): : 302 - 304
  • [42] NITROUS-OXIDE CONTAMINATION DURING OUTPATIENT CRYOSURGERY
    SHEPHERD, JP
    JONES, DC
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1988, 17 (01) : 4 - 5
  • [43] Causes of nitrous oxide contamination in operating rooms - Reply
    Kanmura, Y
    ANESTHESIOLOGY, 1999, 91 (06) : 1961 - 1962
  • [44] REDUCTION OF NITROUS-OXIDE CONTAMINATION IN A PEDIATRIC HOSPITAL
    FLOWERDEW, RMM
    BRUMMITT, WM
    CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1979, 26 (05) : 370 - 374
  • [45] Oxidation of propane with oxygen, nitrous oxide and oxygen/nitrous oxide mixture over Co- and Fe-zeolites
    Novoveská, K
    Bulánek, R
    Wichterlová, B
    CATALYSIS TODAY, 2005, 100 (3-4) : 315 - 319
  • [46] OXYGEN PIPELINE SUPPLY FAILURE - A COPING STRATEGY
    ANDERSON, WR
    BROCKUTNE, JG
    JOURNAL OF CLINICAL MONITORING, 1991, 7 (01): : 39 - 41
  • [47] An unusual incident of oxygen pipeline supply failure
    Dias, L.
    Djabatey, E.
    ANAESTHESIA, 2019, 74 : 13 - 13
  • [48] OXYGEN ELECTRODES SENSITIVE TO NITROUS-OXIDE
    EVANS, MC
    CAMERON, IR
    LANCET, 1978, 2 (8104): : 1371 - 1371
  • [49] Nitrous oxide-oxygen anaesthesia.
    Robinson, WE
    BRITISH MEDICAL JOURNAL, 1916, 1916 : 669 - 669
  • [50] Role of oxygen in the nitrous oxide/carbon reaction
    Byrne, CP
    Yang, RT
    Zhu, ZH
    Lu, GQ
    JOURNAL OF PHYSICAL CHEMISTRY B, 2002, 106 (10): : 2592 - 2596