Feedback and diagnostic signals in direct sample insertion inductively coupled plasma-atomic emission spectrometry (DSI-ICP-AES)

被引:0
|
作者
Wood, TJ [1 ]
Karanassios, V [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
关键词
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Diagnostic signals are discussed in conjunction with an automated direct sample insertion-inductively coupled plasma-atomic emission spectrometry (DSI-ICP-AES) system and a computer-controlled feedback loop. Such signals were obtained from computer-controlled mass flow controllers and by monitoring emission from Ar and C spectral lines in real-time. Use of feedback and diagnostic signals proved to be a particularly attractive way of enhancing the operational reliability of the automated DSI-ICP-AES system used for this work and at improving the overall quality of analytical data obtained with this system.