Low-noise tunneling-magnetoresistance vector magnetometers with flux chopping technique

被引:26
|
作者
Van Su Luong [1 ,3 ]
Jeng, Jen-Tzong [2 ]
Lu, Chih-Cheng [1 ]
Hsu, Hua-Yi [1 ]
机构
[1] Natl Taipei Univ Technol, Grad Inst Mechatron Engn, Dept Mech Engn, Taipei 10608, Taiwan
[2] Natl Kaohsiung Univ Appl Sci, Dept Mech Engn, Kaohsiung 80778, Taiwan
[3] Hanoi Univ Sci & Technol, Int Training Inst Mat Sci, Hanoi 100000, Vietnam
关键词
3-axis magnetometers; Low-frequency noise; Magnetic noise; Magnetic shielding; THERMAL AGITATION; COMPASS;
D O I
10.1016/j.measurement.2017.05.062
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A concept for a low-noise three-axis magnetometer consisting of tunneling magnetoresistance (TMR) sensors and a flux chopper was designed, implemented, and characterized in this work. The TMR sensors used in this study were the model of TMR2102D from Multidimension Technology Inc. Three TMR sensors were aligned orthogonally on a printed circuit board (PCB) and mounted inside a cylindrical flux chopper. The cylindrical flux chopper including a soft magnetic shielding tube and enameled copper wires was 16 mm in length and 8 mm in diameter. The shielding tube was made of a cobalt-based soft magnetic ribbon, Metglas-2714A, from Metglas Inc. The flux chopper modulated the external magnetic flux density using the fluxgate effect, which made the sensors respond to the quasi-static field at the chopping frequency. The demodulated output showed a reduction in low-frequency noise to the level of 0.17 nT/root Hz@1 Hz. To demonstrate the technical feasibility for electronic compass application, the demodulated output of the vector magnetometer prototype was recorded by rotating the device in Earth's magnetic fields about a fixed axis. The Cartesian components B-x, B-y, and B-z of the Earth's fields at various azimuth angles were retrieved by performing a calibration algorithm to correct the non orthogonality caused by the misalignment of TMR sensors and the chopper. The calibrated outputs were linear and orthogonal to each other with an angle error less than 1 and nonlinearity of 0.7%, indicating that the chopping technique is useful to realize an extremely low-noise three-dimensional magnetometer for the geomagnetic application. (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:297 / 303
页数:7
相关论文
共 50 条
  • [31] LOW-NOISE SPECTROMETER FOR THE DAILY MEASUREMENT OF SOLAR CHROMOSPHERIC FLUX
    LIVINGSTON, WC
    MAHAFFEY, C
    YE, B
    PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS, 1982, 331 : 249 - 251
  • [32] Low-noise power supply technique for RF amplifiers
    Koukab, A
    Deval, P
    Dehollain, C
    Declercq, M
    IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS II-ANALOG AND DIGITAL SIGNAL PROCESSING, 2001, 48 (03): : 291 - 293
  • [33] Simplified inelastic electron tunneling spectroscopy based on low-noise derivatives
    Shankar Kesarwani
    Shobhna Misra
    Dipankar Saha
    Maria Luisa Della Rocca
    Indrajit Roy
    Swaroop Ganguly
    Ashutosh Mahajan
    Scientific Reports, 12
  • [34] DIGITAL LOW-NOISE DC POWER-SUPPLY FOR TUNNELING SPECTROMETERS
    EIB, NK
    FURUKAWA, T
    MITTAL, KL
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1984, 55 (07): : 1161 - 1163
  • [35] BIOMAGNETIC MEASUREMENTS USING LOW-NOISE INTEGRATED SQUID MAGNETOMETERS OPERATING IN LIQUID-NITROGEN
    DILORIO, MS
    YANG, KY
    YOSHIZUMI, S
    APPLIED PHYSICS LETTERS, 1995, 67 (13) : 1926 - 1928
  • [36] Simplified inelastic electron tunneling spectroscopy based on low-noise derivatives
    Kesarwani, Shankar
    Misra, Shobhna
    Saha, Dipankar
    Della Rocca, Maria Luisa
    Roy, Indrajit
    Ganguly, Swaroop
    Mahajan, Ashutosh
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [37] Fast low-noise transimpedance amplifier for scanning tunneling microscopy and beyond
    Stubian, Martin
    Bobek, Juraj
    Setvin, Martin
    Diebold, Ulrike
    Schmid, Michael
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2020, 91 (07):
  • [38] Low-noise switched integration amplifier for low-photon flux radiometry
    Li, Yue
    Hu, Yadong
    Yuan, Yinlin
    Zou, Peng
    Zheng, Xiaobing
    APPLIED OPTICS, 2023, 62 (26) : 7058 - 7066
  • [39] Noise properties and epitaxial quality in low-noise integrated YBa2Cu3O7 magnetometers
    Scharnweber, R
    Dieckmann, N
    Schilling, M
    APPLIED PHYSICS LETTERS, 1997, 70 (16) : 2189 - 2191
  • [40] Analysis and Design of Inductorless Wideband Low-Noise Amplifier With Noise Cancellation Technique
    Yu, Yiming
    Kang, Kai
    Fan, Yiming
    Zhao, Chenxi
    Liu, Huihua
    Wu, Yunqiu
    Ban, Yong-Ling
    Yin, Wen-Yan
    IEEE ACCESS, 2017, 5 : 9389 - 9397