Comparison of residual shunt rate and complications across 6 different closure devices for patent foramen ovale

被引:26
|
作者
Fleming, Rubine Gevorgyan [1 ]
Kumar, Preetham [1 ]
West, Brian [1 ]
Noureddin, Nabil [2 ]
Rusheen, Joshua [3 ]
Aboulhosn, Jamil [1 ]
Tobis, Jonathan M. [1 ]
机构
[1] Univ Calif Los Angeles, Dept Med, Div Cardiol, Los Angeles, CA 90024 USA
[2] Univ Nevada, Dept Internal Med, Las Vegas, NV 89154 USA
[3] Univ Calif Los Angeles, David Geffen Sch Med, Los Angeles, CA 90095 USA
关键词
complications; patent foramen ovale; patent foramen ovale closure; PFO closure device; residual shunt rate; MEDICAL THERAPY; IMPACT; METAANALYSIS; DIAGNOSIS; DOPPLER; DISEASE; STROKE;
D O I
10.1002/ccd.28527
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives To compare residual shunt rate and complications associated with six different devices used for PFO closure. Background Transcutaneous PFO closure is an effective treatment for preventing recurrent stroke in patients with a history of cryptogenic stroke. The rate of residual shunt is one metric by which the technical success of PFO closure can be measured. Methods Patients who underwent PFO closure at a single center between February 2001 and July 2019 were retrospectively enrolled in the study. Right-to-left shunt at baseline and during follow-up was assessed using transcranial Doppler (TCD) or transthoracic echocardiography (TTE). Periprocedural and device-related complications, including atrial fibrillation, were also assessed. Results Of 467 PFO closures performed during this period, 320 patients received quantitative assessment of right-to-left shunting both before and after percutaneous closure. The highest effective closure was achieved with the Cardioform device (100%, n = 104), followed by the Amplatzer Cribriform (93%, n = 14), Helex (90%, n = 137), Amplatzer ASO (88%, n = 17), CardioSEAL (86%, n = 14), and Amplatzer PFO (85%, n = 33) devices. The most common significant adverse event was atrial fibrillation, which was more common with the Cardioform device (13%) than the Helex (4%) or the Amplatzer PFO (4%) devices. Conclusions The Gore Cardioform Septal Occluder provides more robust closure of a PFO when compared to other devices but its effectiveness is offset by the higher prevalence of transient atrial fibrillation.
引用
收藏
页码:365 / 372
页数:8
相关论文
共 50 条
  • [21] Implantation of a second closure device in patients with residual shunt after percutaneous closure of patent foramen ovale
    Schwerzmann, M
    Windecker, S
    Wahl, A
    Nedeltchev, K
    Mattle, HP
    Seiler, C
    Meier, B
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 63 (04) : 490 - 495
  • [22] Persistent High Residual Shunt Rate 2 Years After Patent Foramen Ovale Closure Using a Bioabsorbable Device
    Snijder, Roel J. R.
    Post, Martijn C.
    Mulder, Thijs B. J. M.
    Van den Branden, Ben J.
    Ten Berg, Jurien M.
    Suttorp, Maarten J.
    JACC-CARDIOVASCULAR INTERVENTIONS, 2014, 7 (01) : 106 - 107
  • [23] Residual Shunt After Patent Foramen Ovale Closure and Long-Term Stroke Recurrence
    Karagianni, Alexia
    Eriksson, Peter
    Dellborg, Mikael
    ANNALS OF INTERNAL MEDICINE, 2020, 173 (11) : 945 - 946
  • [24] EFFECTIVENESS OF OCCLUDER DEVICES FOR CLOSURE OF PATENT FORAMEN OVALE
    Oto, Ali
    Aytemir, Kudret
    Ozkutlu, Hilmi
    Ozkutlu, Suheyla
    Kaya, Baris
    Ates, Ahmet Hakan
    Yorgun, Hikmet
    Kabakci, Giray
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2010, 140 : S3 - S3
  • [25] Patent foramen ovale closure—not all devices are equal
    Bernhard Meier
    Nature Reviews Cardiology, 2013, 10 : 558 - 559
  • [26] Effect of Residual Interatrial Shunt on Migraine Burden After Transcatheter Closure of Patent Foramen Ovale
    Ben-Assa, Eyal
    Rengifo-Moreno, Pablo
    Al-Bawardy, Rasha
    Kolte, Dhaval
    Cigarroa, Ricardo
    Cruz-Gonzalez, Ignacio
    Sakhuja, Rahul
    Elmariah, Sammy
    Pomerantsev, Eugene
    Vaina, Lucia M.
    Ning, MingMing
    Buonanno, Ferdinando S.
    Hung, Judy W.
    Inglessis, Ignacio
    Palacios, Igor F.
    JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (03) : 293 - 302
  • [27] Transcranial Doppler as a Screening Test for Evaluation of Residual Shunt After Patent Foramen Ovale Closure
    Jang, Albert
    Lee, Joonpyo
    Won, Yoonsun
    Kang, Woong Chol
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 84 (18) : B280 - B281
  • [28] COMPARISON OF DEVICES UTILIZED FOR TRANSCATHETER CLOSURE OF PATENT FORAMEN OVALE: A META ANALYSIS
    Agarwal, Shikhar
    Bajaj, Navkaranbir
    Tuzcu, E. Murat
    Kapadia, Samir
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 59 (13) : E238 - E238
  • [29] Residual Shunt After Patent Foramen Ovale Closure and Long-Term Stroke Recurrence RESPONSE
    Deng, Wenjun
    Yin, Shanye
    McMullin, David
    Inglessis-Azuaje, Ignacio
    Elmariah, Sammy
    Hung, Judy
    Lo, Eng H.
    Palacios, Igor F.
    Buonanno, Ferdinando S.
    Ning, MingMing
    ANNALS OF INTERNAL MEDICINE, 2020, 173 (11) : 946 - 947
  • [30] Association of atrial cardiopathy and residual shunt after patent foramen ovale closure in patients experiencing migraine
    Gao, Xin
    Zhang, Xinxin
    Song, Wei
    Liu, Yan
    Guo, Ran
    Jiang, Yinong
    EUROPEAN JOURNAL OF MEDICAL RESEARCH, 2025, 30 (01)