Experimental investigation of the flow downstream of a dysfunctional bileaflet mechanical aortic valve

被引:9
|
作者
Darwish, Ahmed [1 ]
Di Labbio, Giuseppe [1 ]
Saleh, Wael [1 ]
Smadi, Othman [2 ]
Kadem, Lyes [1 ]
机构
[1] Concordia Univ, Lab Cardiovasc Fluid Dynam Mech Ind & Aerosp Engn, Montreal, PQ, Canada
[2] Hashemite Univ, Dept Biomed Engn, Zarqa, Jordan
基金
加拿大自然科学与工程研究理事会;
关键词
bileaflet mechanical heart valve; flow dynamics; shear stress accumulation; valve dysfunction; viscous energy dissipation; HEART-VALVES; FLUID-MECHANICS; SHEAR-STRESS; OBSTRUCTION; VELOCITY; PARAMETERS; PROSTHESES; ACCURACY; ANGLES; FORCE;
D O I
10.1111/aor.13483
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Mechanical heart valve replacement is the preferred alternative in younger patients with severe symptomatic aortic valve disease. However, thrombus and pannus formations are common complications associated with bileaflet mechanical heart valves. This leads to risks of valve leaflet dysfunction, a life-threatening event. In this experimental study, we investigate, using time-resolved planar particle image velocimetry, the flow characteristics in the ascending aorta in the presence of a dysfunctional bileaflet mechanical heart valve. Several configurations of leaflet dysfunction are investigated and the induced flow disturbances in terms of velocity fields, viscous energy dissipation, wall shear stress, and accumulation of viscous shear stresses are evaluated. We also explore the ability of a new set of parameters, solely based on the analysis of the normalized axial velocity profiles in the ascending aorta, to detect bileaflet mechanical heart valve dysfunction and differentiate between the different configurations tested in this study. Our results show that a bileaflet mechanical heart valve dysfunction leads to a complex spectrum of flow disturbances with each flow characteristic evaluated having its own worst case scenario in terms of dysfunction configuration. We also show that the suggested approach based on the analysis of the normalized axial velocity profiles in the ascending aorta has the potential to clearly discriminate not only between normal and dysfunctional bilealfet heart valves but also between the different leaflet dysfunction configurations. This approach could be easily implemented using phase-contrast MRI to follow up patients with bileaflet mechanical heart valves.
引用
收藏
页码:E249 / E263
页数:15
相关论文
共 50 条
  • [21] SIMULATIONS OF THE HINGE FLOW FIELDS OF A BILEAFLET MECHANICAL HEART VALVE UNDER PHYSIOLOGIC PULSATILE AORTIC CONDITIONS
    Simon, Helene A.
    Ge, Liang
    Borazjani, Iman
    Sotiropoulos, Fotis
    Yoganathan, Ajit P.
    PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE 2008, PTS A AND B, 2009, : 1093 - 1094
  • [22] Computational hemodynamic investigation of a new bileaflet mechanical heart valve
    Khellaf, Belkhiri
    Boussad, Boumeddane
    SIMULATION-TRANSACTIONS OF THE SOCIETY FOR MODELING AND SIMULATION INTERNATIONAL, 2020, 96 (05): : 459 - 469
  • [23] Experience with the Edwards MIRA mechanical bileaflet valve - In the aortic and mitral positions
    Kale, Arzum
    Yildiz, Ulku
    Can, Benhur
    Kandemir, Ozer
    Tokmakogiu, Hilmi
    Tezcaner, Tevfik
    Zorlutuna, Yaman
    TEXAS HEART INSTITUTE JOURNAL, 2006, 33 (03) : 328 - 332
  • [24] In vitro investigation of the blood flow downstream of a 3D-printed aortic valve
    Till Zeugin
    Fergal B. Coulter
    Utku Gülan
    André R. Studart
    Markus Holzner
    Scientific Reports, 14
  • [25] In vitro investigation of the blood flow downstream of a 3D-printed aortic valve
    Zeugin, Till
    Coulter, Fergal B.
    Gulan, Utku
    Studart, Andre R.
    Holzner, Markus
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [26] Simulation of the Three-Dimensional Hinge Flow Fields of a Bileaflet Mechanical Heart Valve Under Aortic Conditions
    Hélène A. Simon
    Liang Ge
    Fotis Sotiropoulos
    Ajit P. Yoganathan
    Annals of Biomedical Engineering, 2010, 38 : 841 - 853
  • [27] Particle image velocimetry investigation of intra-valvular flow fields of a bileaflet mechanical heart valve in a pulsatile flow
    Subramanian, A
    Mu, H
    Kadambi, JR
    Wernet, MP
    Brendzel, AM
    Harasaki, H
    JOURNAL OF HEART VALVE DISEASE, 2000, 9 (05): : 721 - 731
  • [28] ORIENTATION OF BILEAFLET MECHANICAL AORTIC VALVE PROSTHESES FOR OPTIMAL EVALUATION BY TRANSTHORACIC ECHOCARDIOGRAPHY
    Bouzas-Mosquera, Alberto
    Alvarez-Garcia, Nemesio
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2015, 150 (02): : 428 - 430
  • [29] Impact of bileaflet mechanical heart valve leaflet dysfunction on left ventricular blood flow: An experimental study
    Qiang, Yan
    Duan, Tianci
    Zhang, Minzu
    Qi, Liang
    Wei, Liejiang
    Zhong, Wenqi
    PHYSICS OF FLUIDS, 2023, 35 (09)
  • [30] Simulation of the Three-Dimensional Hinge Flow Fields of a Bileaflet Mechanical Heart Valve Under Aortic Conditions
    Simon, Helene A.
    Ge, Liang
    Sotiropoulos, Fotis
    Yoganathan, Ajit P.
    ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (03) : 841 - 853