Second-order systematic errors in Mueller matrix dual rotating compensator ellipsometry

被引:15
|
作者
Broch, Laurent [1 ]
Naciri, Aotmane En [1 ]
Johann, Luc [1 ]
机构
[1] Univ Paul Verlaine Metz, Lab Phys Milieux Denses, F-57078 Metz 3, France
关键词
POLARIMETER; DECOMPOSITION; OPTIMIZATION;
D O I
10.1364/AO.49.003250
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We investigate the systematic errors at the second order for a Mueller matrix ellipsometer in the dual rotating compensator configuration. Starting from a general formalism, we derive explicit second-order errors in the Mueller matrix coefficients of a given sample. We present the errors caused by the azimuthal inaccuracy of the optical components and their influences on the measurements. We demonstrate that the methods based on four-zone or two-zone averaging measurement are effective to vanish the errors due to the compensators. For the other elements, it is shown that the systematic errors at the second order can be canceled only for some coefficients of the Mueller matrix. The calibration step for the analyzer and the polarizer is developed. This important step is necessary to avoid the azimuthal inaccuracy in such elements. Numerical simulations and experimental measurements are presented and discussed. (C) 2010 Optical Society of America
引用
收藏
页码:3250 / 3258
页数:9
相关论文
共 50 条
  • [1] Systematic errors for a Mueller matrix dual rotating compensator ellipsometer
    Broch, Laurent
    Naciri, Aotmane En
    Johann, Luc
    OPTICS EXPRESS, 2008, 16 (12): : 8814 - 8824
  • [2] Depolarization artifacts in dual rotating-compensator Mueller matrix ellipsometry
    Li, Weiqi
    Zhang, Chuanwei
    Jiang, Hao
    Chen, Xiuguo
    Liu, Shiyuan
    JOURNAL OF OPTICS, 2016, 18 (05)
  • [3] Analysis of systematic errors in Mueller matrix ellipsometry as a function of the retardance of the dual rotating compensators
    Broch, Laurent
    Naciri, Aotmane En
    Johann, Luc
    THIN SOLID FILMS, 2011, 519 (09) : 2601 - 2603
  • [4] Experimental study of the systematic errors for a Mueller matrix double rotating compensator ellipsometer
    Piller, G.
    Broch, L.
    Johann, L.
    PHYSICA STATUS SOLIDI C - CURRENT TOPICS IN SOLID STATE PHYSICS, VOL 5, NO 5, 2008, 5 (05): : 1027 - 1030
  • [5] SYSTEMATIC-ERRORS IN ROTATING-COMPENSATOR ELLIPSOMETRY
    KLEIM, R
    KUNTZLER, L
    ELGHEMMAZ, A
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1994, 11 (09): : 2550 - 2559
  • [6] Measurement errors induced by axis tilt of biplates in dual-rotating compensator Mueller matrix ellipsometers
    Gu, Honggang
    Zhang, Chuanwei
    Jiang, Hao
    Chen, Xiuguo
    Li, Weiqi
    Liu, Shiyuan
    MODELING ASPECTS IN OPTICAL METROLOGY V, 2015, 9526
  • [7] Multichannel Mueller matrix ellipsometer based on the dual rotating compensator principle
    Chen, C
    An, I
    Ferreira, GM
    Podraza, NJ
    Zapien, JA
    Collins, RW
    THIN SOLID FILMS, 2004, 455 : 14 - 23
  • [8] Second-order error propagation in the Mueller matrix of a spectropolarimeter
    Ilyin, I.
    ASTRONOMISCHE NACHRICHTEN, 2012, 333 (03) : 213 - 229
  • [9] Eigenvalue calibration method for dual rotating-compensator Mueller matrix polarimetry
    Sheng, Sheng
    Chen, Xiuguo
    Chen, Chao
    Zhuang, Jinfeng
    Wang, Cai
    Gu, Honggang
    Liu, Shiyuan
    OPTICS LETTERS, 2021, 46 (18) : 4618 - 4621
  • [10] Mueller matrix microscope with a dual continuous rotating compensator setup and digital demodulation
    Arteaga, Oriol
    Baldris, Marta
    Anto, Joan
    Canillas, Adolf
    Pascual, Esther
    Bertran, Enric
    APPLIED OPTICS, 2014, 53 (10) : 2236 - 2245