Peritoneal Catheter Fixation to the Abdominal Wall in Surgical Catheter Implantation to Prevent Malfunction

被引:7
|
作者
Wang, Huiming [1 ]
Jia, Houjun [2 ]
Lv, Xifeng [1 ]
Ding, Guohua [1 ]
机构
[1] Wuhan Univ, Renmin Hosp, Renal Dept, Wuhan 430060, Peoples R China
[2] Chongqing Med Univ, Affiliated Hosp 1, Dept Gen Surg, Chongqing, Peoples R China
基金
中国国家自然科学基金;
关键词
Peritoneal dialysis; Catheter malposition; Chronic renal failure; DIALYSIS CATHETER; PLACEMENT; INSERTION; PD;
D O I
10.1159/000368214
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Catheter failure, especially catheter displacement and obstruction remain the major barriers for peritoneal dialysis. We have developed a new surgery technique of catheter fixation on the lower abdominal wall in the catheter planting to avert the catheter mechanical complications. Method: A retrospective study was performed on 93 patients; among them, 52 patients underwent the traditional method of surgery for catheter insertion and 41 patients received additional catheter fixation. Comparisons of complications including infection, leak, infusion pains, catheter displacement, and obstruction occurred during a follow-up period of 6 months, were made between the fixed and non-fixed groups. Results: Catheter fixation cost more time than the conventional operating procedure (94.2 +/- 14.6 min vs. 83 +/- 13.3 min, p = 0.043). Complications of infection, leak, and infusion pain that occurred in the fixed and unfixed groups are comparable. Catheter fixation reduced the complications of catheter displacement or obstruction to 0 episode in the fixed group, whereas those complications were encountered by 7 patients in the unfixed group (0/41 vs. 7/52, p = 0.0 22). All these 7 patients received re-exploration and catheter replacement with further catheter fixation. In the following time until now (ranging from 3 to 16 months), no catheter dysfunction was observed. Conclusions: These results suggest that catheter fixation is effective in preventing catheter displacement and obstruction in peritoneal dialysis. (C) 2014 S. Karger AG, Basel
引用
收藏
页码:109 / 114
页数:6
相关论文
共 50 条
  • [42] Simultaneous abdominal wall defect repair and Tenckhoff catheter placement in candidates for peritoneal dialysis
    Sodo, Maurizio
    Bracale, Umberto
    Argentino, Gennaro
    Merola, Giovanni
    Russo, Roberta
    Sannino, Giuseppina
    Strazzullo, Tania
    Russo, Domenico
    JOURNAL OF NEPHROLOGY, 2016, 29 (05) : 699 - 702
  • [43] Simultaneous abdominal wall defect repair and Tenckhoff catheter placement in candidates for peritoneal dialysis
    Maurizio Sodo
    Umberto Bracale
    Gennaro Argentino
    Giovanni Merola
    Roberta Russo
    Giuseppina Sannino
    Tania Strazzullo
    Domenico Russo
    Journal of Nephrology, 2016, 29 : 699 - 702
  • [44] Shunt malfunction due to proximal migration and subcutaneous coiling of a peritoneal catheter
    Agarwal, Amit
    Kakani, Anand
    JOURNAL OF NEUROSCIENCES IN RURAL PRACTICE, 2010, 1 (02) : 120 - +
  • [45] When it seemed impossible: Irreparable abdominal wall hernia and placement of a peritoneal dialysis catheter
    Menendez Garcia, Natalia
    Gijon, Maria Moreno
    Fernandez, Vanesa Alonso
    Suarez, Carmen Rodriguez
    Bustamante, Adriana Cavada
    Garcia, Clara Sanz
    Ramos, Mateo Garcia
    Miravalles, Jose Luis Rodicio
    Cortes, Elena Astudillo
    PERITONEAL DIALYSIS INTERNATIONAL, 2023, 43 (06): : 479 - 481
  • [46] Primary malfunction of a peritoneal dialysis catheter due to encasement in an encapsulating sheath
    Kazory, Amir
    Cendan, Juan C.
    Hollen, Tracy L.
    Ross, Edward A.
    PERITONEAL DIALYSIS INTERNATIONAL, 2007, 27 (06): : 707 - 709
  • [47] Peritoneal Dialysis Catheter Malfunction Because of Encasement by an Extraluminal Fibrin Sheath
    Singh, S. K. S.
    Common, A.
    Perl, J.
    PERITONEAL DIALYSIS INTERNATIONAL, 2012, 32 (02): : 218 - 220
  • [48] Ventriculoperitoneal Shunt Malfunction Due to a Fibrous Capsule Obstructing the Peritoneal Catheter
    Heaney, Jonathan
    Johnson, Reuben
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2020, 12 (01)
  • [49] Peritoneal catheter obstruction by sigmoid colon wall
    Shuta, Motonishi
    Yoshitaka, Ishibashi
    Hiroo, Kawarazaki
    Daisuke, Katagiri
    Yugo, Hirata
    Haruki, Kume
    Toshiro, Fujita
    KIDNEY INTERNATIONAL, 2011, 79 (09) : 1032 - 1032
  • [50] DUPLICATED PERITONEAL CATHETER AS A CAUSE OF SHUNT MALFUNCTION - CASE-REPORT
    FERRARESI, S
    GRIFFINI, C
    TORCELLO, L
    CASSINARI, V
    NEUROSURGICAL REVIEW, 1991, 14 (02) : 149 - 150