关键词
Anterior cruciate ligament; Anterior cruciate ligament reconstruction; Femoral fixation; Interference screw; Quadriceps tendon graft; Hamstring tendon graft; Biomechanics; ANTERIOR CRUCIATE LIGAMENT; PRESS-FIT FIXATION; GRAFT FIXATION; BONE TUNNEL; PATELLAR TENDON; BIOMECHANICAL EVALUATION; SITE MORBIDITY; PART II; STRENGTH; AUTOGRAFT;
D O I
10.1007/s00167-016-4001-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This cadaveric study compares the biomechanical properties of femoral graft fixation in ACL reconstruction of either quadriceps or hamstring tendon grafts with four different interference screws. The hypothesis was that quadriceps tendon grafts provide at least equal results concerning gap formation during cyclic loading and ultimate failure load compared to hamstring tendon grafts with four different interference screws. Eighty porcine femora underwent interference screw fixation of human tendon grafts for ACL reconstruction. Either quadriceps (Q) or hamstring (H) tendon grafts and four different bioabsorbable interference (Wolf (W), Storz (S), Mitek (M), Arthrex (A)) screws were used, resulting in 8 groups with 10 specimens per groups (WQ, WH, SQ, SH, MQ, MH, AQ, AH). Biomechanical analysis included pretensioning the constructs with 60 N for 30 s, then cyclic loading of 500 cycles between 60 and 250 N at 1 Hz in a servohydraulic testing machine, with measurement of elongation and stiffness including video measurements. After this, ultimate failure load and failure mode analysis were performed. No statistically significant difference could be noted between the groups regarding gap formation during cyclic loading [Cycles 21-500 (mm): WQ 3.6 +/- 0.8, WH 3.9 +/- 1.4, SQ 3.6 +/- 0.8, SH 3.3 +/- 1.5, MQ 4.3 +/- 0.8, MH 4.6 +/- 1.0, AQ 4.8 +/- 0.8, AH 4.3 +/- 1.5, n.s.], stiffness during cyclic loading [Cycles 21-500 (N/mm): WQ 72.9 +/- 16.9, WH 71.6 +/- 20.7, SQ 69.5 +/- 23.9, SH 77.4 +/- 25.1, MQ 59.6 +/- 11.2, MH 48.4 +/- 15.4, AQ 48.8 +/- 12.7, AH 51.9 +/- 22.2, n.s.], and ultimate failure load [(N): WQ 474.4 +/- 88.0, WH 579.3 +/- 124.2, SQ 493.9 +/- 105.2, SH 576.0 +/- 90.4, MQ 478.6 +/- 59.0, MH 543.9 +/- 119.7, AQ 480.2 +/- 93.8, AH 497.8 +/- 74.2, n.s.]. Quadriceps tendon grafts yield comparable biomechanical results for femoral interference screw fixation in ACL reconstruction compared to hamstring tendon grafts. From a clinical perspective, quadriceps tendon grafts should therefore be considered as a good option in ACL reconstruction in the future.
引用
收藏
页码:1241 / 1248
页数:8
相关论文
共 50 条
  • [21] A prospective randomized controlled study of hamstring and bone-free quadriceps tendons autografts in arthroscopic ACL reconstruction
    Tang, Neset
    Eren, Murat
    Gurpinar, Tahsin
    Ozturkmen, Yusuf
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2024, 34 (01): : 293 - 301
  • [22] A prospective randomized controlled study of hamstring and bone-free quadriceps tendons autografts in arthroscopic ACL reconstruction
    Neset Tang
    Murat Eren
    Tahsin Gurpinar
    Yusuf Ozturkmen
    European Journal of Orthopaedic Surgery & Traumatology, 2024, 34 : 293 - 301
  • [23] Pain level after ACL reconstruction: A comparative study between free quadriceps tendon and hamstring tendons autografts
    Buescu, Cristian Tudor
    Onutu, Adela Hilda
    Lucaciu, Dan Osvald
    Todor, Adrian
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2017, 51 (02) : 100 - 103
  • [24] Harvesting hamstring tendons for ACL reconstruction influences postoperative hamstring muscle performance
    Adachi, N
    Ochi, M
    Uchio, Y
    Sakai, Y
    Kuriwaka, M
    Fujihara, A
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2003, 123 (09) : 460 - 465
  • [25] Interference screw divergence in femoral tunnel fixation during endoscopic anterior cruciate ligament reconstruction using hamstring grafts
    Miller, CM
    Tibone, JE
    Hewitt, M
    Kharrazi, FD
    ElAttrache, NS
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2002, 18 (05): : 510 - 514
  • [26] Hamstring To Quadriceps Ratio After ACL Reconstruction In Males And Females
    Kaur, Mandeep
    Hart, Joseph
    MEDICINE & SCIENCE IN SPORTS & EXERCISE, 2020, 52 (07) : 796 - 797
  • [27] Anatomical hamstring tendons ACL reconstruction with the ?linking rings? technique
    Eriksson, Karl
    Cristiani, Riccardo
    JOURNAL OF ISAKOS JOINT DISORDERS & ORTHOPAEDIC SPORTS MEDICINE, 2023, 8 (01) : 57 - 58
  • [28] Clinical Results of Hamstring Autografts in Anterior Cruciate Ligament Reconstruction: A Comparison of Femoral Knot/Press-Fit Fixation and Interference Screw Fixation
    Ho, Wei-Pin
    Lee, Chian-Her
    Huang, Chang-Hung
    Chen, Chih-Hwa
    Chuang, Tai-Yuan
    ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2014, 30 (07): : 823 - 832
  • [29] Hamstring Strength Recovery is Similar Regardless of which Hamstring Tendons are Harvested for ACL Reconstruction
    Ardern, Clare L.
    Webster, Kate E.
    Taylor, Nicholas F.
    Feller, Julian A.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2010, 42 (05): : 149 - 150
  • [30] Dynamic loss of pressure on ACL hamstring grafts - an explanation for poor biomechanical performance with interference screw fixation
    Meyer, D. C.
    Koch, P. P.
    Snedeker, J. G.
    Farshad, M.
    SWISS MEDICAL WEEKLY, 2010, 140 (23-24) : 17S - 17S