The calibration of a turbine meter as a secondary flow-rate standard by a one-point relative method using laser Doppler velocimetry
被引:1
|作者:
March, JF
[1
]
机构:
[1] Phys Tech Bundesanstalt, Fachlab Messung Therm Energie, D-10587 Berlin, Germany
关键词:
D O I:
10.1088/0957-0233/9/1/017
中图分类号:
T [工业技术];
学科分类号:
08 ;
摘要:
Heat district suppliers demand carefully constructed and calibrated flow sensors of heat meters at temperatures above 100 degrees C for long-term use. The calibration at these temperatures cannot be performed by the common gravimetric standard method due to technical difficulties. A novel one-point relative method using a laser Doppler velocimeter (LDV) was therefore developed which is based on the gravimetric method below 100 degrees C and on the Reynolds law. This method allows a turbine meter to be calibrated as a secondary flow-rate standard with a relative uncertainty less than or equal to 0.2% for temperatures up to 180 degrees C with regard to the flow rate.