Dynamic diffusion model for tracing the real-tune potential response of polymeric membrane ion-selective electrodes

被引:48
|
作者
Radu, A [1 ]
Meir, AJ [1 ]
Bakker, E [1 ]
机构
[1] Auburn Univ, Dept Chem & Math, Auburn, AL 36849 USA
关键词
D O I
10.1021/ac049348t
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A numerical solution for the prediction of the time-dependent potential response of a polymeric-based ion-selective electrode (ISE) is presented. The model addresses short- and middle-term potential drifts that are dependent on changes in concentration gradients in the aqueous sample and organic membrane phase. This work has important implications for the understanding of the real-time response behavior of potentiometric sensors with low detection limits and with nonclassical super-Nernstian response slopes. As a model system, the initial exposure of membranes containing the well-examined silver ionophore O,O"-bis[2-(methylthio)ethyl]-tert-butylcalix[4]arene was monitored, and the large observed potential drifts were compared to theoretical predictions. The model is based on an approximate solution of the diffusion equation for both aqueous and organic diffusion layers using a numerical scheme (finite difference in time and finite elements in space). The model may be evaluated on the basis of experimentally available parameters and gives time-dependent information previously inaccessible with a simpler steady-state diffusion model. For the cases studied, the model gave a very good correlation with experimental data, albeit with lower than expected diffusion coefficients for the organic phase. This model may address numerous open questions regarding the response time and memory effects of low-detection-limit ion-selective electrodes and for other membrane electrodes where ion fluxes are relevant.
引用
收藏
页码:6402 / 6409
页数:8
相关论文
共 50 条
  • [41] Molecular recognition of amino acid esters in liquid polymeric membrane ion-selective electrodes
    Krondak, M
    Shishkanova, TV
    Holakovsky, R
    Volf, R
    Stibor, I
    Král, V
    ANALYTICA CHIMICA ACTA, 2001, 448 (1-2) : 19 - 25
  • [42] MERCURY ION-SELECTIVE POLYMERIC MEMBRANE ELECTRODES BASED ON SUBSTITUTED DIAZA CROWN ETHERS
    BRZOZKA, Z
    PIETRASZKIEWICZ, M
    ELECTROANALYSIS, 1991, 3 (08) : 855 - 858
  • [43] Thin polymeric membrane ion-selective electrodes for trace-level potentiometric detection
    Wang, Junhao
    Liang, Rongning
    Qin, Wei
    ANALYTICA CHIMICA ACTA, 2020, 1139 (1139) : 1 - 7
  • [44] Single-Piece Solid-Contact Polymeric Membrane Ion-Selective Electrodes for Silver Ion
    Yin, Tanji
    Liu, Yuanyuan
    Qin, Wei
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2013, 160 (08) : B91 - B94
  • [45] Basic response principles of voltammetric ion-selective solvent polymeric membrane sensors
    Bakker, E
    Jadhav, S
    SOLID-STATE IONIC DEVICES, 1999, 99 (13): : 390 - 397
  • [46] Application of the interface equilibria-triggered dynamic diffusion model of the boundary potential for the numerical simulation of neutral carrier-based ion-selective electrodes response
    Egorov, Vladimir V.
    Novakovskii, Andrei D.
    ANALYTICA CHIMICA ACTA, 2018, 1043 : 20 - 27
  • [47] Electrochemical response characteristics and analytical application of papaverine ion-selective membrane electrodes
    Kharitonov, S
    ANALYTICAL LETTERS, 2006, 39 (02) : 259 - 273
  • [48] STUDY OF RESPONSE OF ION-SELECTIVE LIQUID MEMBRANE ELECTRODES TO AROMATIC ACID ANIONS
    HIRSCH, RF
    OLDERMAN, GM
    WALDMAN, B
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1972, 164 (AUG-S): : 26 - &
  • [49] RESPONSE-TIME STUDIES ON NEUTRAL CARRIER ION-SELECTIVE MEMBRANE ELECTRODES
    LINDNER, E
    TOTH, K
    PUNGOR, E
    MORF, WE
    SIMON, W
    ANALYTICAL CHEMISTRY, 1978, 50 (12) : 1627 - 1631
  • [50] RESPONSE OF CYANIDE ION-SELECTIVE MEMBRANE ELECTRODES IN PRESENCE OF METAL-IONS
    MASCINI, M
    NAPOLI, A
    ANALYTICAL CHEMISTRY, 1974, 46 (03) : 447 - 449