Optimizing photon fluence measurements for the accurate determination of detective quantum efficiency

被引:1
|
作者
Wong, Molly [1 ,2 ]
Zhang, Da [1 ,2 ]
Rong, John [3 ]
Wu, Xizeng [4 ]
Liu, Hong [1 ,2 ]
机构
[1] Univ Oklahoma, Ctr Bioengn, Norman, OK 73019 USA
[2] Univ Oklahoma, Sch Elect & Comp Engn, Norman, OK 73019 USA
[3] Univ Texas MD Anderson Canc Ctr, Imaging Phys Dept, Houston, TX 77030 USA
[4] Univ Alabama, Dept Radiol, Birmingham, AL 35233 USA
基金
美国国家卫生研究院;
关键词
MODULATION TRANSFER-FUNCTION; COMPUTED RADIOGRAPHY SYSTEM; NOISE POWER SPECTRUM; X-RAY SPECTROSCOPY; DIGITAL MAMMOGRAPHY; IMAGING-SYSTEMS; ALIGNMENT; PERFORMANCE; RESOLUTION; DEVICES;
D O I
10.1117/1.3268363
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Our goal was to evaluate the error contributed by photon fluence measurements to the detective quantum efficiency (DQE) of an x-ray imaging system. The investigation consisted of separate error analyses for the exposure and spectrum measurements that determine the photon fluence. Methods were developed for each to determine the number of measurements required to achieve an acceptable error. A new method for calculating the magnification factor in the exposure measurements was presented and compared to the existing method. The new method not only produces much lower error at small source-to-image distances (SIDs) such as clinical systems, but is also independent of SID. The exposure and spectra results were combined to determine the photon fluence error contribution to the DQE of 4%. The error in this study is small because the measurements resulted from precisely controlled experimental procedures designed to minimize the error. However, these procedures are difficult to follow in clinical environments, and application of this method on clinical systems could therefore provide important insight into error reduction. This investigation was focused on the error in the photon fluence contribution to the DQE, but the error analysis method can easily be extended to a wide range of applications. (C) 2009 SPIE and IS&T. [DOI: 10.1117/1.3268363]
引用
收藏
页数:9
相关论文
共 50 条
  • [31] ACCURATE DETERMINATION OF NEUTRON FLUENCE IN FISSION TRACK DATING
    赵云龙
    Chinese Science Bulletin, 1990, (12) : 983 - 987
  • [32] Detective efficiency of photon counting detectors with spectral degradation and crosstalk
    Rajbhandary, Paurakh L.
    Persson, Mats
    Pelc, Norbert J.
    MEDICAL PHYSICS, 2020, 47 (01) : 27 - 36
  • [33] A NEW METHOD FOR THE EXPERIMENTAL-DETERMINATION OF THE DETECTIVE QUANTUM EFFICIENCY OF X-RAY SCREENS
    DICK, CE
    MOTZ, JW
    ROEHRIG, H
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1980, 134 (04) : 853 - 853
  • [34] Optimizing the Efficiency of Solar Photon Upconversion
    Frazer, Laszlo
    Gallaher, Joseph K.
    Schmidt, T. W.
    ACS ENERGY LETTERS, 2017, 2 (06): : 1346 - 1354
  • [35] Modeling the frequency-dependent detective quantum efficiency of photon-counting x-ray detectors
    Stierstorfer, Karl
    MEDICAL PHYSICS, 2018, 45 (01) : 156 - 166
  • [36] Analysis of the detective quantum efficiency of a dual-energy photon-counting X-ray detector
    Lee, Junho
    Yoo, Seungjun
    Oh, Seokwon
    Park, Seongbon
    Lim, Chang Hwy
    Park, Jong Won
    Tanguay, Jesse
    Kim, Ho Kyung
    NDT & E INTERNATIONAL, 2025, 155
  • [37] MEASUREMENT OF DETECTIVE QUANTUM EFFICIENCY OF PHOTOGRAPHIC-EMULSIONS
    JENKINS, RL
    JOURNAL OF PHOTOGRAPHIC SCIENCE, 1974, 22 (05): : 250 - 251
  • [38] Detective Quantum Efficiency as a Quality Parameter of Imaging Equipment
    O. N. Morgun
    K. E. Nemchenko
    Yu. V. Rogov
    Biomedical Engineering, 2003, 37 (5) : 258 - 261
  • [39] Simulation of the detective quantum efficiency for a hybrid pixel detector
    Norrlid, LD
    Edling, F
    Fransson, K
    Brenner, R
    Bingefors, N
    Gustafsson, L
    Rönnqvist, C
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 543 (2-3): : 528 - 536
  • [40] THE DETECTIVE QUANTUM EFFICIENCY OF SCREEN-FILM SYSTEMS
    BUNCH, PC
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1988, 135 (08) : C389 - C389