Determination of phosphorus in small amounts of protein samples by ICP-MS

被引:56
|
作者
Becker, JS [1 ]
Boulyga, SF
Pickhardt, C
Becker, J [1 ]
Buddrus, S
Przybylski, M
机构
[1] Res Ctr Julich, Ctr Dept Analyt Chem, D-52425 Julich, Germany
[2] Univ Konstanz, Dept Analyt Chem, D-78457 Constance, Germany
关键词
ICP-SFMS; ICP-QMS; proteins; phosphorus;
D O I
10.1007/s00216-002-1737-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Inductively coupled plasma mass spectrometry (ICP-MS) is used for phosphorus determination in protein samples. A small amount of solid protein sample (down to 1 mug) or digest (1-10 muL) protein solution was denatured in nitric acid and hydrogen peroxide by closed-microvessel microwave digestion. Phosphorus determination was performed with an optimized analytical method using a double-focusing sector field inductively coupled plasma mass spectrometer (ICP-SFMS) and quadrupole-based ICP-MS (ICP-QMS). For quality control of phosphorus determination a certified reference material (CRM), single cell proteins (BCR 273) with a high phosphorus content of 26.8+/-0.4 mg g(-1), was analyzed. For studies on phosphorus determination in proteins while reducing the sample amount as low as possible the homogeneity of CRM BCR 273 was investigated. Relative standard deviation and measurement accuracy in ICP-QMS was within 2%, 3.5%, 11% and 12% when using CRM BCR 273 sample weights of 40 mg, 5 mg, 1 mg and 0.3 mg, respectively. The lowest possible sample weight for an accurate phosphorus analysis in protein samples by ICP-MS is discussed. The analytical method developed was applied for the analysis of homogeneous protein samples in very low amounts [1-100 mug of solid protein sample, e.g. beta-casein or down to 1 muL of protein or digest in solution (e.g., tau protein)]. A further reduction of the diluted protein solution volume was achieved by the application of flow injection in ICP-SFMS, which is discussed with reference to real protein digests after protein separation using 2D gel electrophoresis. The detection limits for phosphorus in biological samples were determined by ICP-SFMS down to the ng g(-1) level. The present work discusses the figure of merit for the determination of phosphorus in a small amount of protein sample with ICP-SFMS in comparison to ICP-QMS.
引用
收藏
页码:561 / 566
页数:6
相关论文
共 50 条
  • [41] Determination of neptunium in soil by ICP-MS
    Ayranov, M
    Krähenbühl, U
    Sahli, H
    Röllin, S
    Burger, M
    RADIOCHIMICA ACTA, 2005, 93 (9-10) : 631 - 635
  • [42] Isotopic analyses by ICP-MS in clinical samples
    Rodushkin, Ilia
    Engstrom, Emma
    Baxter, Douglas C.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2013, 405 (09) : 2785 - 2797
  • [43] ICP-MS determination of Re at ultra trace levels in rock and soil samples
    Tagami, K
    Uchida, S
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (06) : 669 - 671
  • [44] Isotopic analyses by ICP-MS in clinical samples
    Ilia Rodushkin
    Emma Engström
    Douglas C. Baxter
    Analytical and Bioanalytical Chemistry, 2013, 405 : 2785 - 2797
  • [45] EDXRF and ICP-MS analysis of environmental samples
    Nguyen, TH
    Boman, J
    Leermakers, M
    X-RAY SPECTROMETRY, 1998, 27 (04) : 265 - 276
  • [46] EDXRF and ICP-MS Analysis of Environmental Samples
    Physics and Engineering Physics, Chalmers University of Technology, S-412 96 Gothenburg, Sweden
    不详
    不详
    X-Ray Spectrom., 4 (265-276):
  • [47] Determination of thorium by ICP-MS and ICP-OES
    Holmes, L
    RADIATION PROTECTION DOSIMETRY, 2001, 97 (02) : 117 - 122
  • [48] Rapid Determination of Uranium in Ore Samples by Online Extraction Coupled ICP-MS
    Li, Faliang
    Guo, Dongfa
    Song, Lili
    Li, Boping
    Xu, Jiaquan
    Zhang, Xinglei
    ATOMIC SPECTROSCOPY, 2023, 44 (05) : 336 - 342
  • [49] Speciation and ICP-MS determination of vanadium, nickel and beryllium in soil samples.
    Ovári, M
    Papp, V
    Záray, G
    MAGYAR KEMIAI FOLYOIRAT, 2000, 106 (5-6): : 208 - 212
  • [50] Determination of Sr-90 at ultratrace levels in environmental samples by ICP-MS
    Abdel-Lateef, A. M.
    ARAB JOURNAL OF NUCLEAR SCIENCES AND APPLICATIONS, 2012, 45 (03): : 51 - 63