Commercial Slit Lamp Anterior Segment Photography versus Digital Compact Camera Mounted on a Standard Slit Lamp with an Adapter

被引:7
|
作者
Oliphant, Huw [1 ,2 ]
Kennedy, Alasdair [1 ,2 ]
Comyn, Oliver [1 ]
Spalton, David J. [3 ,4 ]
Nanavaty, Mayank A. [1 ,2 ]
机构
[1] Brighton & Sussex Univ Hosp NHS Trust, Sussex Eye Hosp, Eastern Rd, Brighton BN2 5BF, E Sussex, England
[2] Univ Sussex, Brighton & Sussex Med Sch, Brighton, E Sussex, England
[3] London Clin, London, England
[4] Guys & St Thomas NHS Fdn Trust, St Thomas Hosp, London, England
关键词
Anterior segment photography; compact digital camera; photography; SYSTEM;
D O I
10.1080/02713683.2018.1490435
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare slit lamp mounted anterior segment cameras (SLCs) versus digital compact camera (DCC) with slit-lamp adaptor when used by an inexperienced technician.Methods: In this cross sectional study, where posterior capsule opacification (PCO) was used as a comparator, patients were consented for one photograph with SLC and two with DCC (DCC1 and DCC2), with a slit lamp adaptor. An inexperienced clinic technician, who took all the photographs and masked the images, recruited one eye of each patient. Images were graded for PCO using EPCO 2000 software by two independent masked graders. Repeatability between DCC1 and DCC2, and limits-of-agreement between SLC and DCC1 mounted on slit-lamp with an adaptor were assessed. Coefficient-of-repeatability and Bland-Altmann plots were analyzed.Results: Seventy-two patients (eyes) were recruited in the study. First 9 patients (eyes) were excluded due to unsatisfactory image quality from both the systems. Mean evaluation of posterior capsule opacification (EPCO) score for SLC was 2.28 (95% CI: 2.09-2.45), for DCC1 was 2.28 (95% CI: 2.11-2.45), and for the DCC2 was 2.11 (95% CI: 2.11-2.45). There was no significant difference in EPCO scores between SLC vs. DCC1 (p=0.98) and between DCC1 and DCC2 (p=0.97). Coefficient of repeatability between DCC images was 0.42, and the coefficient of repeatability between DCC and SLC was 0.58.Conclusions: DCC on slit lamp with an adaptor is comparable to a SLC. There is an initial learning curve, which is similar for both for an inexperienced person. This opens up the possibility for low cost anterior segment imaging in the clinical, research, and teaching settings.
引用
收藏
页码:1290 / 1294
页数:5
相关论文
共 50 条
  • [31] Replacing the Slit Lamp with a Mobile Multi-Output Projector Device for Anterior Segment Imaging
    Sinha, Shantanu
    Savidis, Nickolaos
    Lawson, Everett
    Raskar, Ramesh
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)
  • [33] Comparing Remote Assessment of Limbal Anterior Chamber Depth Using Novel Stereoscopic Digital Slit Lamp Images with Anterior Segment OCT
    Cronin, Thomas
    Gupta, Ankur
    Abbas, Kinza
    Gupta, Akash
    Armstrong, Grayson
    Olson, Joanna
    Ingraham, Herbert
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [34] Renal cell carcinoma choroidal metastasis recorded by smartphone with interface eyepiece adapter mounted on slit lamp A case report
    Xu, Qi-bin
    Hu, Zhi-yi
    Wu, Shuang-qing
    MEDICINE, 2021, 100 (10) : E24910
  • [35] Comparison of Visante and slit-lamp anterior segment optical coherence tomography in imaging the anterior chamber angle
    L M Sakata
    T T L Wong
    H-T Wong
    R S Kumar
    H M Htoon
    H T Aung
    M He
    T Aung
    Eye, 2010, 24 : 578 - 587
  • [36] Comparison of Visante and slit-lamp anterior segment optical coherence tomography in imaging the anterior chamber angle
    Sakata, L. M.
    Wong, T. T. L.
    Wong, H-T
    Kumar, R. S.
    Htoon, H. M.
    Aung, H. T.
    He, M.
    Aung, T.
    EYE, 2010, 24 (04) : 578 - 587
  • [37] Design and Performance Characterization of a Novel, Smartphone-Based, Portable Digital Slit Lamp for Anterior Segment Screening Using Telemedicine
    Dutt, Sreetama
    Nagarajan, Shanmuganathan
    Vadivel, Siva S.
    Baig, Adeeb Ulla
    Savoy, Florian M.
    Ganapathy, Vasudev M.
    Dominic, Mathew
    Sivaraman, Anand
    Rao, Divya Parthasarathy
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2021, 10 (08):
  • [38] Anterior Segment Variations with Age and Accommodation Demonstrated by Slit-Lamp-Adapted Optical Coherence Tomography
    Yan, Pi-Song
    Lin, Hao-Tian
    Wang, Qi-Lin
    Zhang, Zhen-Pin
    OPHTHALMOLOGY, 2010, 117 (12) : 2301 - 2307
  • [39] Remote Assessment of Anterior Chamber Depth Using Novel Stereoscopic Digital Slit Lamp Images
    Cronin, Thomas
    Gupta, Ankur
    Armstrong, Grayson Wilkes
    Olson, Joanna
    Ingraham, Herbert
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [40] SLIT-LAMP STUDIES OF THE RHESUS-MONKEY EYE .1. SURVEY OF THE ANTERIOR SEGMENT
    KORETZ, JF
    NEIDER, MW
    KAUFMAN, PL
    BERTASSO, AM
    DEROUSSEAU, CJ
    BITO, LZ