Bimodal electric tissue ablation (BETA) - in-vivo evaluation of the effect of applying direct current before and during radiofrequency ablation of porcine liver

被引:10
|
作者
Cockburn, J. F.
Maddern, G. J.
Wemyss-Holden, S. A.
机构
[1] Norfolk & Norwich Univ Hosp, Dept Radiol, Norwich NR4 7UY, Norfolk, England
[2] Norfolk & Norwich Univ Hosp, Dept Surg, Norwich NR4 7UY, Norfolk, England
[3] Univ Adelaide, Queen Elizabeth Hosp, Dept Surg, Woodville, SA 5011, Australia
关键词
D O I
10.1016/j.crad.2006.11.008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
AIM: To examine the effect of applying increasing amounts of direct current (DC) before and during alternating current radiofrequency ablation of porcine liver. MATERIALS AND METHODS: Using a Radiotherapeutics RF3000 generator, a 9 V AC/DC transformer and a 16 G plain aluminium tube as an electrode, a control group of 24 porcine hepatic radiofrequency ablation zones was compared with 24 zones created using a bimodal electric tissue ablation (BETA) technique in three pigs. All ablations were terminated when tissue impedance rose to greater than 999 Omega or radiofrequency energy input felt below 5 W on three successive measurements taken at 1 min intervals. BETA ablations were performed in two phases: an initial phase of variable duration DC followed by a second phase during which standard radiofrequency ablation was applied simultaneously with DC. During this second phase, radiofrequency power input was regulated by the feedback circuitry of the RF3000 generator according to changes in tissue impedance. The diameters (mm) of each ablation zone were measured by two observers in two planes perpendicular to the plane of needle insertion. The mean short axis diameter of each ablation zone was subjected to statistical analysis. RESULTS: With increased duration of prior application of DC, there was a progressive increase in the diameter of the ablation zone (p < 0.001). This effect increased sharply up to 300 s of pre-treatment after which a further increase in diameter occurred, but at a much lesser rate. A maximum ablation zone diameter of 32 mm was produced (control diameters 10-13 mm). CONCLUSION: Applying a 9 V DC to porcine liver in vivo, and continuing this DC application during subsequent radiofrequency ablation, results in larger ablation zone diameters compared with radiofrequency ablation atone. (c) 2006 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:213 / 220
页数:8
相关论文
共 23 条
  • [1] Bimodal Electric Tissue Ablation (BETA)-Effect of Reversing the Polarity of the Direct Current on the Size of Ablation
    Tiong, Leong U.
    Finnie, John W.
    Field, John B.
    Maddern, Guy J.
    JOURNAL OF SURGICAL RESEARCH, 2012, 174 (02) : 305 - 311
  • [2] Measurement of Intrahepatic Pressure During Radiofrequency Ablation In Vivo Porcine Liver
    Kawamoto, C.
    Yamauchi, A.
    Kaneko, K.
    Miyagi, N.
    Yakabi, K.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (05)
  • [3] Bimodal Electric Tissue Ablation (BETA): A Study on Ablation Size When the Anode is Placed on the Peritoneum and the Liver
    Tiong, Leong U.
    Finnie, John W.
    Field, John B.
    Maddern, Guy J.
    JOURNAL OF SURGICAL RESEARCH, 2012, 176 (01) : 114 - 120
  • [4] A Comparison of Direct Heating During Radiofrequency and Microwave Ablation in Ex Vivo Liver
    Anita Andreano
    Christopher L. Brace
    CardioVascular and Interventional Radiology, 2013, 36 : 505 - 511
  • [5] A Comparison of Direct Heating During Radiofrequency and Microwave Ablation in Ex Vivo Liver
    Andreano, Anita
    Brace, Christopher L.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 36 (02) : 505 - 511
  • [6] In-vivo evaluation of a simulation model for interstitial tumor ablation by radiofrequency thermotherapy in porcine liver with normal and interrupted hepatic perfusion
    Ritz, JP
    Isbert, C
    Roggan, A
    Lehmann, K
    Buhr, H
    GASTROENTEROLOGY, 2004, 126 (04) : A801 - A801
  • [7] In-vivo evaluation of a simulation model for interstitial tumor ablation by radiofrequency therapy in porcine liver with normal and interrupted hepatic perfusion
    Ritz, JP
    Isbert, C
    Lehmann, KS
    Roggan, A
    Knappe, V
    Buhr, HJ
    Germer, CT
    GASTROENTEROLOGY, 2003, 124 (04) : A806 - A806
  • [8] Optimal conditions for "heat-conduct effect" in liver tissue during radiofrequency ablation
    Jiang, K.
    Su, M.
    Liu, Y.
    Zhao, X.
    Liu, J.
    Zhang, W.
    Wang, J.
    Dong, J.
    Huang, Z.
    PANMINERVA MEDICA, 2013, 55 (03) : 297 - 302
  • [9] Influence of interapplicator distance on multibipolar radiofrequency ablation during physiological and interrupted liver perfusion in an in vivo porcine model
    Poch, F. G. M.
    Neizert, C. A.
    Geyer, B.
    Gemeinhardt, O.
    Bruder, L.
    Niehues, S. M.
    Vahldiek, J. L.
    Bressem, K. K.
    Kreis, M. E.
    Lehmann, K. S.
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [10] Influence of interapplicator distance on multibipolar radiofrequency ablation during physiological and interrupted liver perfusion in an in vivo porcine model
    F. G. M. Poch
    C. A. Neizert
    B. Geyer
    O. Gemeinhardt
    L. Bruder
    S. M. Niehues
    J. L. Vahldiek
    K. K. Bressem
    M. E. Kreis
    K. S. Lehmann
    Scientific Reports, 10