Endovascular procedures under near-real-time magnetic resonance imaging guidance:: An experimental feasibility study

被引:22
|
作者
Dion, YM
El Kadi, HB
Boudoux, C
Gourdon, J
Chakfé, N
Traoré, A
Moisan, C
机构
[1] CHU Quebec, St Francois Assise Hosp, IMRI Unit, Quebec City, PQ G1L 3L5, Canada
[2] Univ Laval, Dept Surg, Quebec City, PQ, Canada
[3] Univ Laval, Dept Engn, Quebec City, PQ, Canada
[4] Univ Laval, Lab Anim Div, Quebec City, PQ, Canada
[5] Hop Univ Strasbourg, Dept Surg, Strasbourg, France
关键词
D O I
10.1067/mva.2000.109208
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: The purpose of this study was to assess the feasibility of insertion of endovascular stents and the precision of an open-field interventional nlagnetic resonance imaging (iMRI) system in an in vivo model, Methods: A feasibility study was undertaken at a university-affiliated hospital. Three male piglets with an average age of 6 months and a weight between 70 and 77 kg and two 3-month-old male piglets that weighed 40 to 44 kg were anesthetized. The five piglets underwent placement of nitinol stents inserted through the right femoral artery under the guidance of a SIGNA-SP 0.5T open-configuration iMRI unit. With a dedicated high-resolution near-real-time MRI sequence, the stent was guided and deployed onto a predefined target. Results: The main outcome measures were the duration of the procedure from the beginning of positioning to the end of deployment of the stent, the final position of the stent in relation to the target on the iMRI screen, and comparison with autopsy findings. Three stents were deployed within the aorta at the level of the renal arteries, and two were deployed within the right iliac artery just below the aortic trifurcation, The average duration of the endovascular deployment was 13 minutes. There was an agreement of 0.6 mm in the position of the stent as observed on iMR images and found at autopsy. When the piglets were sacrificed, the average distance between the stents and the predefined target was 7.8 mm, mostly because of the migration of one stent. Axial views allowed for accurate determination of stent impaction on the vascular wall. Conclusions: This study confirms the feasibility of stent deployment under near-real-time MRI guidance. It also emphasizes some inherent characteristics that hold promise with regard to other conventional techniques: stents and vascular structures are visualized in near-real-time in any desired plane, and the technique is performed without the potential adverse effects of ionizing radiations and iodinated contrast agents.
引用
收藏
页码:1006 / 1014
页数:9
相关论文
共 50 条
  • [1] Real-Time Magnetic Resonance Imaging Guidance for Cardiovascular Procedures
    Horvath, Keith A.
    Li, Ming
    Mazilu, Dumitru
    Guttman, Michael A.
    McVeigh, Elliot R.
    SEMINARS IN THORACIC AND CARDIOVASCULAR SURGERY, 2007, 19 (04) : 330 - 335
  • [2] Near-real-time guidance using intraoperative magnetic resonance imaging for radical evacuation of hypertensive hematomas in the basal ganglia
    Bernays, RL
    Kollias, SS
    Romanowski, B
    Valavanis, A
    Yonekawa, Y
    NEUROSURGERY, 2000, 47 (05) : 1081 - 1089
  • [3] Real-Time Magnetic Resonance Imaging to Guide Pediatric Endovascular Procedures
    A.N. Raval
    R.J. Lederman
    Pediatric Cardiology, 2005, 26 : 251 - 259
  • [4] Real-time magnetic resonance imaging to guide pediatric endovascular procedures
    Raval, AN
    Lederman, RJ
    PEDIATRIC CARDIOLOGY, 2005, 26 (03) : 251 - 259
  • [6] Near-real-time guidance using intraoperative magnetic resonance imaging for radical evacuation of hypertensive hematomas in the basal ganglia - Comment
    Fahlbusch, R
    Nimsky, C
    NEUROSURGERY, 2000, 47 (05) : 1089 - 1089
  • [7] Near real time magnetic resonance image registration of the heart for guidance of interventional procedures
    Costa, A
    Petrie, D
    Yen, Y
    Drangova, M
    MEDICAL PHYSICS, 2002, 29 (06) : 1311 - 1311
  • [8] Dynamic near-real-time magnetic resonance imaging for analyzing the velopharyngeal closure in comparison with videofluoroscopy
    Beer, AJ
    Hellerhoff, P
    Zimmermann, A
    Mady, K
    Sader, R
    Rummeny, EJ
    Hannig, C
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (05) : 791 - 797
  • [9] Feasibility of real-time magnetic resonance imaging for catheter guidance in electrophysiology studies
    Nazarian, Saman
    Kolandaivelu, Aravindan
    Zviman, Menekhem M.
    Meininger, Glenn R.
    Kato, Ritsushi
    Susil, Robert C.
    Roguin, Ariel
    Dickfeld, Timm L.
    Ashikaga, Hiroshi
    Calkins, Hugh
    Berger, Ronald D.
    Bluemke, David A.
    Lardo, Albert C.
    Halperin, Henry R.
    CIRCULATION, 2008, 118 (03) : 223 - 229
  • [10] Feasibility of Near-Real-Time GOLD Data Products
    Codrescu, S. M.
    Rowland, W. F.
    Plummer, T. M.
    Vanier, B. A.
    Berger, T. E.
    Codrescu, M. V.
    JOURNAL OF GEOPHYSICAL RESEARCH-SPACE PHYSICS, 2020, 125 (06)