摘要
Distortion exists in the present capsule endoscope image resulting from the confined space and the wide-angle requirement [8]. Based on the previous two lens works, the optimal design had obtained that the field of view was about 86 degrees, and MTF was about 18% at 100 1p/mm, but distortion would go to -26%. It's difficult to add another lens on the 7mm optical path between the dome and imaging lenses for improving distortion. In order to overcome this problem, we intend to design the optical dome as another optical lens. The original dome is transparent and has an equal thickness, namely without refracting light almost. Our objective in this paper is to design the inner curvature of the dome and associate two aspheric imaging lenses in front of the CMOS sensors to advance the distortion with maintaining field of view and MTF under the same capsule volume. Furthermore, the paper proposes the real object plane of intestine is nearly a curved surface rather than an ideal flat surface. Taking those reasons under consideration, we design three imaging lenses with curved object plane and obtain the field of view is about 86 degrees, MTF is about 26% at 100 1p/mm, and the distortion improve to -7.5%. Adding the dome lens is not only to enhance the image quality, but also to maintain the tiny volume requirement.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Capsule endoscope aspiration: A common complication?
    Sanchez-Chavez, X.
    Martinez-Garcia, C. L.
    REVISTA DE GASTROENTEROLOGIA DE MEXICO, 2013, 78 (04): : 257 - 258
  • [42] Radiofrequency controlled movement of capsule endoscope
    Kochman, Michael L.
    Mosse, Sandi
    Swain, Paul
    AMERICAN JOURNAL OF GASTROENTEROLOGY, 2006, 101 (09): : S501 - S501
  • [43] Prototype design of wireless capsule endoscope
    Chan, Yawen
    Meng, Max Q. -H.
    Wang, Xiaona
    2005 IEEE INTERNATIONAL CONFERENCE ON MECHATRONICS AND AUTOMATIONS, VOLS 1-4, CONFERENCE PROCEEDINGS, 2005, : 400 - 403
  • [44] Delayed presentation of an aspirated capsule endoscope
    Singh, Rohit A.
    Pai, Krishnanand R.
    Satur, Chris M. R.
    JOURNAL OF SURGICAL CASE REPORTS, 2013, (09):
  • [45] High Speed Receiver for Capsule Endoscope
    Woo, S. H.
    Yoon, K. W.
    Moon, Y. K.
    Lee, J. H.
    Park, H. J.
    Kim, T. W.
    Choi, H. C.
    Won, C. H.
    Cho, J. H.
    JOURNAL OF MEDICAL SYSTEMS, 2010, 34 (05) : 843 - 847
  • [46] Transit times of the wireless capsule endoscope
    Appleyard, MN
    Glukhovsky, A
    Jacob, H
    Gat, D
    Lewkowicz, S
    Swain, P
    GASTROINTESTINAL ENDOSCOPY, 2001, 53 (05) : AB122 - AB122
  • [47] MIMO Antennas for Capsule Endoscope Systems
    Poshtgol, Parisa Lotfi
    Li Jichao
    Soltani, Saber
    Murch, Ross D.
    2016 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM, 2016, : 1175 - 1176
  • [48] The application of the inkjet technology in capsule endoscope
    Chiu, CL
    Huang, YF
    Sha, YW
    Tseng, ZF
    Chen, CT
    IS&T'S NIP21: INTERNATIONAL CONFERENCE ON DIGITAL PRINTING TECHNOLOGIES, FINAL PROGRAM AND PROCEEDINGS, 2005, : 299 - 302
  • [49] High Speed Receiver for Capsule Endoscope
    S. H. Woo
    K. W. Yoon
    Y. K. Moon
    J. H. Lee
    H. J. Park
    T. W. Kim
    H. C. Choi
    C. H. Won
    J. H. Cho
    Journal of Medical Systems, 2010, 34 : 843 - 847
  • [50] Endoscopic placement of the capsule endoscope in children
    Barth, BA
    Donovan, K
    Fox, VL
    GASTROINTESTINAL ENDOSCOPY, 2004, 60 (05) : 818 - 821