On the Detection of Propofol in Exhaled Air by MOS Gas Sensors

被引:1
|
作者
Bur, Christian [1 ]
Lensch, Henrik [1 ]
Maurer, Felix [2 ,3 ]
Kreuer, Sascha [2 ,3 ]
Schuetze, Andreas [1 ]
机构
[1] Saarland Univ, Lab Measurement Technol, Campus A5 1, D-66123 Saarbrucken, Germany
[2] Saarland Univ Med Ctr, CBR Ctr Breath Res, Dept Anaesthesiol Intens Care & Pain Therapy, D-66424 Homburg, Germany
[3] Saarland Univ, Fac Med, D-66424 Homburg, Germany
关键词
propofol; MOS sensor; temperature cycled operation; breath analysis; DISEASE DIAGNOSIS; BREATH ANALYSIS;
D O I
10.1515/teme-2022-0060
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Bedside and non-invasive measurement of drugs in exhaled air within short time frames or even online could have a high therapeutic relevance. In this study, a commercially available metal oxide semiconductor (MOS) gas sensor, i.e. ENS160 from ScioSense, in temperature cycled operation (TCO) is studied for quantification of the intravenous anesthetic propofol. Propofol was chosen because a correlation of the plasma concentration to the concentration in exhaled air was already proven. A partial least regression model (PLSR) was built for quantification of propofol while suppressing the influence of humidity and ethanol as a main interferent. The root mean squared error (RMSE) of validation was as low as 4 ppb for a concentration range of 0 ppb to 40 ppb. The results suggest the usage of MOS sensor for drug monitoring.
引用
收藏
页码:66 / 71
页数:6
相关论文
共 50 条
  • [1] Metal oxide-based gas sensors for the detection of exhaled breath markers
    Vajhadin, Fereshteh
    Mazloum-Ardakani, Mohammad
    Amini, Abbas
    Medical Devices and Sensors, 2021, 4 (01):
  • [2] EXHALED BREATH ANALYSIS BY RESISTIVE GAS SENSORS
    Chludzinski, Tomasz
    Kwiatkowski, Andrzej
    METROLOGY AND MEASUREMENT SYSTEMS, 2020, 27 (01) : 81 - 89
  • [3] DEVICE WITH SEMICONDUCTOR GAS SENSOR FOR ALCOHOL VAPOR DETECTION IN AN EXHALED AIR SAMPLE
    MAKSIMOVICH, NP
    YEREMINA, LE
    SENSORS AND ACTUATORS B-CHEMICAL, 1993, 13 (1-3) : 256 - 258
  • [4] Device with semiconductor gas sensor for alcohol vapour detection in an exhaled air sample
    Maksimovich, Nelly P.
    Yeremina, L.E.
    Sensors and Actuators, B: Chemical, 1993, B13 (1 -3 pt 1) : 256 - 258
  • [5] Zeolite A coated Zn1-XCuXO MOS sensors for NO gas detection
    Galioglu, Sezin
    Karaduman, Irmak
    Corlu, Tugba
    Akata, Burcu
    Yildirim, Memet Ali
    Ates, Aytunc
    Acar, Selim
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN ELECTRONICS, 2018, 29 (02) : 1356 - 1368
  • [6] PRINTED CANTILEVERS AND MOS CIAS SENSORS FOR HAZARDOUS GAS DETECTION AT ROOM TEMPERATURE
    Debeda, Helene
    Nguyen, Van Son
    Almazan, Fernando
    Pilar, Maria Pina
    Jubera, Veronique
    Lucat, Claude
    CERAMICS FOR ENVIRONMENTAL SYSTEMS, 2016, 257 : 139 - 145
  • [7] Zeolite A coated Zn1−XCuXO MOS sensors for NO gas detection
    Sezin Galioglu
    Irmak Karaduman
    Tuğba Çorlu
    Burcu Akata
    Memet Ali Yıldırım
    Aytunç Ateş
    Selim Acar
    Journal of Materials Science: Materials in Electronics, 2018, 29 : 1356 - 1368
  • [8] Pulse Heating Combined With Machine Learning for Enhanced Gas Identification and Concentration Detection With MOS Gas Sensors
    Zhuang, Yi
    Liu, Xiaojiang
    Wang, Xue
    Niu, Gaoqiang
    Cheng, Ran
    Wang, Fei
    IEEE SENSORS LETTERS, 2023, 7 (10)
  • [9] Detection of Asymptomatic Nigrostriatal Dopaminergic Lesion in Rats by Exhaled Air Analysis Using Carbon Nanotube Sensors
    Tisch, Ulrike
    Aluf, Yuval
    Ionescu, Radu
    Nakhleh, Morad
    Bassal, Rana
    Axelrod, Noa
    Robertman, Dorina
    Tessler, Yael
    Finberg, John P. M.
    Haick, Hossam
    ACS CHEMICAL NEUROSCIENCE, 2012, 3 (03): : 161 - 166
  • [10] NEUTRON DETECTION PROPERTIES OF MOS SENSORS
    DANZIGER, M
    BERNHARDT, A
    DORSCHEL, B
    SCHOOP, K
    STREUBEL, G
    BRAUNE, A
    KERNENERGIE, 1985, 28 (09): : 397 - 403