Preliminary biomechanical results of a novel pin configuration for external fixation of vertical shear pelvic fractures

被引:8
|
作者
Stewart, Rebecca [1 ,2 ]
Kieser, David C. [1 ,2 ]
Scholze, Mario [3 ]
Hammer, Niels [3 ]
Stone, Brad [1 ,2 ]
Hooper, Gary [1 ,2 ]
机构
[1] Christchurch Hosp, Dept Orthopaed Surg, 2 Riccarton Ave, Christchurch 8011, New Zealand
[2] Christchurch Hosp, MSM, 2 Riccarton Ave, Christchurch 8011, New Zealand
[3] Univ Otago, Dept Anat, Anat Biomech Lab, Dunedin, New Zealand
关键词
external fixator; fracture; pelvic injury; trauma; vertical shear; RING FRACTURES; INJURIES; MANAGEMENT; REDUCTION;
D O I
10.1111/ans.14822
中图分类号
R61 [外科手术学];
学科分类号
摘要
BackgroundVertical shear fractures are unstable and potentially life-threatening injuries that require urgent reduction and stabilization. The aim of this study was to compare the biomechanical efficacy of three different external fixation pin configurations for vertical shear pelvic fractures in a cadaveric model. We hypothesized that a modified external fixation pin configuration with a crestal (CR) pin in the stable hemipelvis and bilateral supra-acetabular (SA) pins provides the greatest overall stability to axial loading. MethodsThe force to failure within a standard standing axial load (maximum 650N) was tested on 10 human cadaveric pelvises with vertical shear fractures. Three pin configurations were compared including iliac crest (IC), SA and a modified SA frame with a third CR pin on the stable hemipelvis. Both displacement at the posterior pelvis at 650N and force to failure of >25mm displacement was recorded. ResultsThe mean force to failure was highest with CR (499N), then IC (350N) and then SA (265N) pin configurations, being statistically non-significant (P=0.165). The minimum force to failure followed a similar trend with 296, 68 and 43N for CR, IC and SA, respectively. About 1/4 CR, 1/4 IC and 2/9 SA pins sustained 650N or more without failure. ConclusionIt was shown that this new design may reliably withstand a seated physiological load of 250N. However, none of the three pin configurations tested can reliably withstand a standing load of 650N. Further experiments are needed to quantify these findings under physiological loading.
引用
收藏
页码:1051 / 1055
页数:5
相关论文
共 50 条
  • [41] METHODS AND TECHNIQUES OF PERCUTANEOUS EXTERNAL FIXATION IN PELVIC FRACTURES
    Apostolov, Pavlin
    Burnev, Martin
    Milkov, Petar
    JOURNAL OF IMAB, 2011, 17 (01): : 166 - 171
  • [42] Distraction external fixation in lateral compression pelvic fractures
    Bellabarba, C
    Ricci, WM
    Bolhofner, BR
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2000, 14 (07) : 475 - 482
  • [43] Surgical complications and implications of external fixation of pelvic fractures
    Palmer, S
    Fairbank, AC
    Bircher, M
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1997, 28 (9-10): : 649 - 653
  • [44] An intramedullary Echidna pin for fixation of comminuted clavicle fractures: a biomechanical study
    David Ackland
    Ian Griggs
    Patrick Hislop
    Wen Wu
    Minoo Patel
    Martin Richardson
    Journal of Orthopaedic Surgery and Research, 12
  • [45] An intramedullary Echidna pin for fixation of comminuted clavicle fractures: a biomechanical study
    Ackland, David
    Griggs, Ian
    Hislop, Patrick
    Wu, Wen
    Patel, Minoo
    Richardson, Martin
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2017, 12
  • [46] Augmentation by cerclage wire improves fixation of vertical shear femoral neck fractures-A biomechanical analysis
    Kuan, Fa-Chuan
    Yeh, Ming-Long
    Hong, Chih-Kai
    Chiang, Florence L.
    Jou, I-Ming
    Wang, Ping-Hui
    Su, Wei-Ren
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (10): : 2081 - 2086
  • [47] Vertical transtalar Steinmann pin fixation for unstable ankle fractures
    Morgan-Jones, RL
    Smith, KD
    Thomas, PBM
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2000, 82 (03) : 185 - 189
  • [48] VERTICAL TRANSARTICULAR-PIN FIXATION FOR UNSTABLE ANKLE FRACTURES
    CHILDRESS, HM
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1965, A 47 (07): : 1323 - +
  • [49] THE USE OF EXTERNAL PIN FIXATION IN LATE COMPOUND FRACTURES OF WAR
    JANES, EC
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1948, 30 (03): : 572 - 572
  • [50] A safer pin position for external fixation of distal radial fractures
    Emami, A
    Mjöberg, B
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 (09): : 749 - 750