An algorithm for source checking continuous air monitors using radon progeny

被引:0
|
作者
Hogue, MG
Smith, AB
Hadlock, DJ
机构
来源
HEALTH PHYSICS | 2000年 / 79卷 / 03期
关键词
air sampling; instrumentation; statistics; radon;
D O I
10.1097/00004032-200009000-00012
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Department of Energy requirements contained within 10CFR835 require that continuous air monitors be periodically checked for operability. The DOE air monitoring implementation guide for 10CFR835 allows the use of radon progeny to perform the recommended weekly source check. The Defense Waste Processing Facility located at the Savannah River Site has demonstrated that, through the use of the Hypotheses Concerning Two Means, diurnal change in the radon progeny detected by the monitors meets the requirements for weekly source checks. The use of the diurnal change in radon progeny has replaced the person-hours expended performing direct weekly source checking with an automated system requiring minimal person-hour expenditure.
引用
收藏
页码:299 / 304
页数:6
相关论文
共 50 条
  • [2] RADON CHAMBER IN THE CENTRAL MINING INSTITUTE-THE CALIBRATION FACILITY FOR RADON AND RADON PROGENY MONITORS
    Skubacz, K.
    Chalupnik, S.
    Urban, P.
    Wysocka, M.
    [J]. RADIATION PROTECTION DOSIMETRY, 2017, 177 (1-2) : 164 - 167
  • [3] MEASUREMENT OF RADON PROGENY CONCENTRATIONS IN AIR
    KERR, GD
    [J]. TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1973, 17 (NOV): : 542 - 543
  • [4] INTEGRATING RADON PROGENY AIR SAMPLER
    SCHIAGER, KJ
    [J]. AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1974, 35 (03): : 165 - 174
  • [5] INTERPRETATION OF THE RESPONSE OF CONTINUOUS RADON MONITORS TO TRANSIENT RADON CONCENTRATIONS
    BUSIGIN, A
    VANDERVOOREN, AW
    PHILLIPS, CR
    [J]. HEALTH PHYSICS, 1979, 37 (05): : 659 - 667
  • [6] Studies of radon and radon progeny in air conditioned rooms in hospitals
    Marley, F
    Denman, AR
    Phillips, PS
    [J]. RADIATION PROTECTION DOSIMETRY, 1998, 76 (04) : 273 - 276
  • [7] THE CONTROL OF RADON PROGENY BY AIR TREATMENT DEVICES
    RAJALA, M
    JANKA, K
    LEHTIMAKI, M
    KULMALA, V
    GRAEFFE, G
    KESKINEN, J
    [J]. SCIENCE OF THE TOTAL ENVIRONMENT, 1985, 45 (OCT) : 493 - 498
  • [8] Radon exhalation and ultrafine fraction of radon progeny in closed room air
    El-Hussein, A
    Mohammed, A
    Ahmed, AA
    [J]. ATMOSPHERIC ENVIRONMENT, 1999, 33 (02) : 183 - 190
  • [9] Continuous measurements of radon and radon progeny as a basis for management of radon as a hazard in a tourist cave
    W. Zahorowski
    S. Whittlestone
    J. M. James
    [J]. Journal of Radioanalytical and Nuclear Chemistry, 1998, 236 : 219 - 225
  • [10] Continuous measurements of radon and radon progeny as a basis for management of radon as a hazard in a tourist cave
    Zahorowski, W
    Whittlestone, S
    James, JM
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 1998, 236 (1-2) : 219 - 225