Comparison of quantitative measurements between two different intravascular ultrasound catheters and consoles: in vitro and in vivo studies

被引:2
|
作者
Okura, Hiroyuki [1 ,2 ]
Watanabe, Makoto [1 ]
Miura, Akihiro [3 ]
Kurokawa, Muneo [3 ]
Ueda, Tomoya [1 ]
Soeda, Tsunenari [1 ]
Saito, Yoshihiko [1 ]
机构
[1] Nara Med Univ, Dept Cardiovasc Med, Kashihara, Nara, Japan
[2] Gifu Univ, Dept Cardiol, Grad Sch Med, Gifu, Japan
[3] Nara Med Univ, Dept Med Engn, Kashihara, Nara, Japan
关键词
Coronary artery disease; Intravascular ultrasound; Atherosclerosis; Stent;
D O I
10.1007/s12928-021-00759-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Previous studies suggested possible discordant quantitative measurements between different IVUS catheters and/or systems. The purpose of this study was to assess compatibility of two different IVUS catheters and consoles for quantitative measurements of coronary arteries. (1). In vitro study: IVUS imaging was performed in a concentric cylindrical phantom with 6 sections of known, cross-sectional diameter ranging from 3.0 to 8.0 mm. The lumen diameter (LD) and lumen cross-sectional area (LA) were measured and compared. To compare between 2 different IVUS consoles, IVUS images were obtained using a single IVUS catheter (catheter 1) connected to 2 different IVUS consoles (console 1 and 2). To compare between 2 different IVUS catheters, IVUS imaging was obtained using 2 different IVUS catheters (catheter 1 and 2) connected to a single IVUS console (console 2). (2). In vivo study: IVUS imaging was performed in 40 stented coronary arterial segments from 40 patients. The maximal stent diameter (Max SD), minimal stent diameter (minSD), and stent area (SA) were measured at both distal and proximal stent edges and compared between the two IVUS consoles (console 1 and 2) connected to a single IVUS catheter (catheter 1) (n = 20). IVUS imaging was also performed to compare between catheter 1 and 2 connected to IVUS console 2 (n = 20). Both in vitro and in vivo studies showed good correlation between the two IVUS consoles as well as two IVUS catheters. In conclusion, two IVUS catheters and consoles provide comparable IVUS measures both in vitro and in vivo.
引用
收藏
页码:109 / 115
页数:7
相关论文
共 50 条
  • [1] Comparison of quantitative measurements between two different intravascular ultrasound catheters and consoles: in vitro and in vivo studies
    Hiroyuki Okura
    Makoto Watanabe
    Akihiro Miura
    Muneo Kurokawa
    Tomoya Ueda
    Tsunenari Soeda
    Yoshihiko Saito
    Cardiovascular Intervention and Therapeutics, 2022, 37 : 109 - 115
  • [2] Comparison of quantitative measurements between two different intravascular ultrasound systems: In vitro and in vivo studies
    Yamada, Ryotaro
    Okura, Hiroyuki
    Kume, Teruyoshi
    Hayashida, Akihiro
    Neishi, Yoji
    Kawamoto, Takahiro
    Yoshida, Kiyoshi
    JOURNAL OF CARDIOLOGY, 2013, 61 (3-4) : 201 - 205
  • [3] Variability in quantitative measurement of the same segment with two different intravascular ultrasound systems: In vivo and in vitro studies
    Li, YX
    Honye, J
    Saito, S
    Takayama, T
    Yokoyama, SI
    Saruya, T
    Kotani, M
    Harasawa, K
    Ando, H
    Endo, M
    Kanmatsuse, K
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 62 (02) : 175 - 180
  • [4] Intravascular ultrasound imaging of atherosclerotic renal arteries: comparison between in vitro and in vivo studies
    Yasuda, G
    Takizawa, T
    Takasaki, I
    Shionoiri, H
    Umemura, S
    Shimoyama, K
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 1998, 13 (07) : 1690 - 1695
  • [5] In vitro and in vivo comparison of three different intravascular ultrasound catheter designs
    Fort, S
    Freeman, NA
    Johnston, P
    Cohen, EA
    Foster, FS
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 52 (03) : 382 - 392
  • [6] Comparison of Lumen Measurements Between Quantitative Coronary Angiography and Intravascular Ultrasound: A PROSPECT substudy
    Goto, Kosaku
    Mintz, Gary S.
    Lansky, Alexandra
    Weisz, Giora
    McPherson, John
    De Bruyne, Bernard
    Serruys, Patrick W.
    Xu, Ke
    Stone, Gregg W.
    Maehara, Akiko
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (18) : B199 - B199
  • [7] Comparison of Bioresorbable Scaffold Measurements between Intravascular Ultrasound and Quantitative Coronary Angiography in the ABSORB Japan Trial
    Okada, Kozo
    Honda, Yasuhiro
    Kitahara, Hideki
    Hollak, M. Brooke
    Yock, Paul G.
    Popma, Jeffrey J.
    Kusano, Hajime
    Cheong, Wai-Fung
    Sudhir, Krishna
    Fitzgerald, Peter J.
    Kimura, Takeshi
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 70 (18) : B57 - B58
  • [8] Variability in quantitative measurement of the same segment with two different intravascular ultrasound
    Li, Y.
    Advances in Heart Disease, 2006, : 207 - 214
  • [9] Comparison of intravascular ultrasound and quantitative angiographic measurements for reference vessel lumen dimensions
    Mintz, GS
    Desai, K
    Negotia, M
    Kane, K
    Laird, JR
    Satler, LF
    Kent, KM
    CIRCULATION, 2000, 102 (18) : 635 - 635
  • [10] Head-to-head comparison of quantitative measurements between intravascular imaging systems: An in vitro phantom study
    Nishi, Takeshi
    Imura, Shinji
    Kitahara, Hideki
    Kobayashi, Yoshio
    Yock, Paul G.
    Fitzgerald, Peter J.
    Honda, Yasuhiro
    IJC HEART & VASCULATURE, 2021, 36