Liquid chromatography-electrospray ionisation mass spectrometry method for the determination of escitalopram in human plasma and its application in bioequivalence study

被引:46
|
作者
Singh, SS [1 ]
Shah, H [1 ]
Gupta, S [1 ]
Jain, M [1 ]
Sharma, K [1 ]
Thakkar, P [1 ]
Shah, R [1 ]
机构
[1] Zydus Res Ctr, Biomed DMPK Dept, Ahmedabad 382213, Gujarat, India
关键词
escitalopram; validation; bioequivalence;
D O I
10.1016/j.jchromb.2004.09.001
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel liquid chromatographic-electrospray ionisation mass spectrometric (LC-ESI-MS) method has been developed for the determination of escitalopram, an antidepressant in human plasma using paroxetine as internal standard. The method involved liquid-liquid extraction of the analyte from human plasma with a mixture of diethyl ether and dichloromethane (70:30, v/v). The chromatographic separation was achieved within 7.0 min by using 2.0 mM ammonium acetate (pH 5.0)-acetonitrile (54:46, v/v) as mobile phase and a ODS YMC(TM) AQ 150 mm x 4.6 mm analytical column; the flow-rate was 1.0 ml/min. Ion signals m/z 325.0 and 330.0 for escitalopram and internal standard, were measured in the positive mode. A detailed validation of the method was performed as per USFDA guidelines and the standard curves were found to be linear in the range of 1.0-200 ng/ml with a mean correlation coefficient more than 0.99. The absolute recovery was more than 75% for both escitalopram and internal standard. The method was simple, sensitive, precise, accurate and was successfully applied to the bioequivalence study of escitalopram in healthy, male, human subjects. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:209 / 215
页数:7
相关论文
共 50 条
  • [1] Determination of levocetirizine in human plasma by liquid chromatography-electrospray tandem mass spectrometry: Application to a bioequivalence study
    Morita, M. R.
    Berton, D.
    Boldin, R.
    Barros, F. A. P.
    Meurer, E. C.
    Amarante, A. R.
    Campos, D. R.
    Calafatti, S. A.
    Pereira, R.
    Abib, E., Jr.
    Pedrazolli, J., Jr.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 862 (1-2): : 132 - 139
  • [2] Determination of memantine in human plasma by liquid chromatography-electrospray tandem mass spectrometry: Application to a bioequivalence study
    Almeida, A. A.
    Campos, D. R.
    Bernasconi, G.
    Calafatti, S.
    Barros, F. A. P.
    Eberlin, M. N.
    Meurer, E. C.
    Paris, E. G.
    Pedrazzoli, J.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 848 (02): : 311 - 316
  • [3] Validated method for determination of bromopride in human plasma by liquid chromatography-electrospray tandem mass spectrometry: application to the bioequivalence study
    Nazare, P
    Massaroti, P
    Duarte, LF
    Campos, DR
    Marchioretto, MAM
    Bernasconi, G
    Calafatti, S
    Barros, FAP
    Meurer, EC
    Pedrazzoli, J
    Moraes, LAB
    JOURNAL OF MASS SPECTROMETRY, 2005, 40 (09): : 1197 - 1202
  • [5] Determination of bevantolol in human plasma using liquid chromatography-electrospray ionization tandem mass spectrometry and its application to a bioequivalence study
    Ren, Li
    Wang, Zheng
    Lou, Yiceng
    Zheng, Lu
    Zheng, Heng
    Yin, Chunping
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2014, 959 : 16 - 21
  • [6] Determination of flumazenil in human plasma by liquid chromatography-electrospray ionisation tandem mass spectrometry
    Lavén, M
    Appel, L
    Moulder, R
    Tyrefors, N
    Markides, K
    Långström, B
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 808 (02): : 221 - 227
  • [7] Determination of huperzine A in human plasma by liquid chromatography-electrospray tandem mass spectrometry: application to a bioequivalence study on Chinese volunteers
    Li, Wei
    Li, Jianhua
    Hu, Qin
    BIOMEDICAL CHROMATOGRAPHY, 2008, 22 (04) : 354 - 360
  • [8] Determination of formoterol in rat plasma by liquid chromatography-electrospray ionisation mass spectrometry
    Kakubari, I.
    Dejima, H.
    Miura, K.
    Koga, Y.
    Mizu, H.
    Takayasu, T.
    Yamauchi, H.
    Takayama, S.
    Takayama, K.
    PHARMAZIE, 2007, 62 (02): : 94 - 95
  • [9] Determination of fenofibric acid in human plasma by ultra performance liquid chromatography-electrospray ionization mass spectrometry: application to a bioequivalence study
    Bhavesh, Dasandi
    Shah, Sanjay
    Shivprakash
    BIOMEDICAL CHROMATOGRAPHY, 2009, 23 (09) : 922 - 928
  • [10] Determination of lafutidine in human plasma by high-performance liquid chromatography-electrospray ionization mass spectrometry: application to a bioequivalence study
    Wu, LL
    Zhang, ZJ
    Tian, Y
    Li, W
    Xu, FG
    Chen, Y
    Wei, HL
    JOURNAL OF MASS SPECTROMETRY, 2005, 40 (12): : 1637 - 1643