Determination of bevantolol in human plasma using liquid chromatography-electrospray ionization tandem mass spectrometry and its application to a bioequivalence study

被引:1
|
作者
Ren, Li [2 ]
Wang, Zheng [3 ]
Lou, Yiceng [3 ]
Zheng, Lu [4 ]
Zheng, Heng [1 ]
Yin, Chunping [2 ]
机构
[1] Huazhong Univ Sci & Technol, Tongji Med Coll, Toni Hosp, Dept Pharm, Wuhan 430030, Peoples R China
[2] Huazhong Univ Sci & Technol, Tongji Med Coll, Coll Pharm, Wuhan 430030, Peoples R China
[3] Wuhan Univ Technol, Sch Chem & Engn, Wuhan 430070, Peoples R China
[4] Shanghai Haini Pharmaceut Co Ltd, Yangtze River Pharmaceut Grp, Shanghai 201318, Peoples R China
关键词
Bevantolol; LC-MS/MS; Bioequivalence; ENANTIOMERS;
D O I
10.1016/j.jchromb.2014.03.040
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A liquid chromatography-electrospray ionization tandem mass spectrometry method was established and validated for the determination of bevantolol in human plasma using propranolol as the internal standard. The optimal chromatographic behavior of bevantolol and propranolol was achieved on a Welch Ultimate XB-C-18 column (5 mu m, 150 mm x 2.1 mm, Maryland, USA) with a mobile phase of acetonitrile-water (40:60, v/v) containing 10 mM ammonium acetate and 0.1% formic acid. The mass spectrometer was operated in selected reaction monitoring mode using the transition m/z 346.1 > 165.1 for bevantolol and m/z 260.3 > 116.1 for propranolol. Sample preparation was carried out through protein precipitation with acetonitrile. The calibration curves were linear over the range of 5.00-1000 ng/ml. The intra- and inter-day pretisions were less than 6.7% and 6.6%, respectively. This method was successfully applied to the bioequivalence study of two kinds of bevantolol hydrochloride tablets in 24 Chinese male volunteers in fasting and postprandial experiment. (c) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:16 / 21
页数:6
相关论文
共 50 条
  • [1] Determination of levocetirizine in human plasma by liquid chromatography-electrospray tandem mass spectrometry: Application to a bioequivalence study
    Morita, M. R.
    Berton, D.
    Boldin, R.
    Barros, F. A. P.
    Meurer, E. C.
    Amarante, A. R.
    Campos, D. R.
    Calafatti, S. A.
    Pereira, R.
    Abib, E., Jr.
    Pedrazolli, J., Jr.
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 862 (1-2): : 132 - 139
  • [2] Determination of memantine in human plasma by liquid chromatography-electrospray tandem mass spectrometry: Application to a bioequivalence study
    Almeida, A. A.
    Campos, D. R.
    Bernasconi, G.
    Calafatti, S.
    Barros, F. A. P.
    Eberlin, M. N.
    Meurer, E. C.
    Paris, E. G.
    Pedrazzoli, J.
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 848 (02): : 311 - 316
  • [3] Determination of cyclobenzaprine in human plasma by liquid chromatography-electrospray ionization tandem mass spectrometry and its application in a pharmacokinetic study
    Xiang, Yi
    Zhou, Liping
    Qian, Zhenyu
    Peng, Kelong
    Li, Dongyan
    Chen, Xiaomei
    Jiang, Hongliang
    Zheng, Heng
    [J]. BIOMEDICAL CHROMATOGRAPHY, 2012, 26 (09) : 1083 - 1088
  • [4] Validated method for determination of bromopride in human plasma by liquid chromatography-electrospray tandem mass spectrometry: application to the bioequivalence study
    Nazare, P
    Massaroti, P
    Duarte, LF
    Campos, DR
    Marchioretto, MAM
    Bernasconi, G
    Calafatti, S
    Barros, FAP
    Meurer, EC
    Pedrazzoli, J
    Moraes, LAB
    [J]. JOURNAL OF MASS SPECTROMETRY, 2005, 40 (09): : 1197 - 1202
  • [5] Determination of huperzine A in human plasma by liquid chromatography-electrospray tandem mass spectrometry: application to a bioequivalence study on Chinese volunteers
    Li, Wei
    Li, Jianhua
    Hu, Qin
    [J]. BIOMEDICAL CHROMATOGRAPHY, 2008, 22 (04) : 354 - 360
  • [6] Determination of clarithromycin in human plasma by liquid chromatography-electrospray ionization tandem mass spectrometry
    Jiang, Yao
    Wang, Jiang
    Li, Hao
    Wang, Yingwu
    Gu, Jingkai
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 43 (04) : 1460 - 1464
  • [7] Determination of glimepiride in human plasma by liquid chromatography-electrospray ionization tandem mass spectrometry
    Salem, II
    Idrees, J
    Al Tamimi, JI
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 799 (01): : 103 - 109
  • [8] Determination of bulleyaconitine A in human plasma by liquid chromatography-electrospray ionization tandem mass spectrometry
    Weng, WY
    Xu, HN
    Huang, JM
    Wang, GQ
    Shen, T
    Zhang, JF
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 816 (1-2): : 315 - 320
  • [9] Determination of glimepiride in human plasma by liquid chromatography-electrospray ionization tandem mass spectrometry
    Kim, HY
    Chang, KY
    Lee, HJ
    Han, SB
    [J]. BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2004, 25 (01) : 109 - 114
  • [10] Determination of penciclovir in human plasma by liquid chromatography-electrospray ionization tandem mass spectrometry: application to a clinical pharmacokinetic study
    Kanneti, Ravi
    Bhavesh, Dasandi
    Paramar, Dharmesh
    Shivaprakash, R.
    Bhatt, Parloop A.
    [J]. BIOMEDICAL CHROMATOGRAPHY, 2011, 25 (04) : 458 - 465