Determination of Slow-Neutron Fluence Rate Using Gamma-Ray Spectroscopy

被引:0
|
作者
Gawelczyk, B. [1 ]
Konefal, A. [1 ]
机构
[1] Univ Silesia Katowice, Inst Phys, Katowice, Poland
关键词
HPGe; Monte Carlo; Neutrons;
D O I
10.12693/APhysPolA.139.280
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
In this work, the method of determination of the slow-neutron fluence rate using a gamma-ray spectroscopy with a high-purity germanium detector (HPGe) and Monte Carlo calculations was presented. The prompt gamma rays with energies of 595.9, 867.9, and 1204.2 keV, originating from the nuclear reaction Ge-73(n, gamma)Ge-74 in the germanium crystal of the HPGe detector, were registered. A special system composed of two neutron sources (Cf-252, PuBe), a lead shield and a HPGe detector was made to find correlations between net areas of the peaks from these prompt gamma rays in the measured spectra and the fluence rate of slow neutrons getting to the germanium crystal. The net areas of the peaks were related to the slow-neutron fluence rate using the Monte Carlo calculations based on the GEANT4 code. The presented method using the gamma-ray spectroscopy can be applied in laboratories which do not have detectors for neutron measurements.
引用
收藏
页码:280 / 282
页数:3
相关论文
共 50 条
  • [31] Gamma-ray spectroscopy
    Knodlseder, Jurgen
    HIGHLIGHTS OF ASTRONOMY, VOL 13, 2005, 13 : 784 - 786
  • [32] Determination of plutonium isotopic composition by gamma-ray spectroscopy
    Tang, Peijia
    Wu, Xin
    Liu, Daming
    Wang, Siguang
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 1998, 32 (03): : 214 - 220
  • [33] URANIUM ISOTOPIC RATIO DETERMINATION BY GAMMA-RAY SPECTROSCOPY
    ROWSON, JW
    HONTZEAS, SA
    NUCLEAR INSTRUMENTS & METHODS, 1978, 154 (03): : 541 - 548
  • [34] THE DETERMINATION OF PLUTONIUM ISOTOPIC COMPOSITION BY GAMMA-RAY SPECTROSCOPY
    SAMPSON, TE
    HSUE, ST
    PARKER, JL
    JOHNSON, SS
    BOWERSOX, DF
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 193 (1-2): : 177 - 183
  • [35] A new active method for the measurement of slow-neutron fluence in modern radiotherapy treatment rooms
    Gomez, F.
    Iglesias, A.
    Sanchez Doblado, F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (04): : 1025 - 1039
  • [36] Gamma-ray spectroscopy and neutron activation analysis using GAMANAL-PC and QUACANAL
    Nucl Instrum Methods Phys Res Sect B, 1-4 (509):
  • [38] Nuclear Forensics using Gamma-ray Spectroscopy
    Norman, E. B.
    HEAVY ION ACCELERATOR SYMPOSIUM 2015: INTERNATIONAL NUCLEAR STRUCTURE CONFERENCE IN REMEMBRANCE OF GEORGE DRACOULIS, 2016, 123
  • [39] Gamma spectroscopy and gamma-ray intensity determination for 48V
    Tramontano, R
    Helene, O
    Pascholati, PR
    Maidana, NL
    Vanin, VR
    INTERNATIONAL CONFERENCE ON NUCLEAR DATA FOR SCIENCE AND TECHNOLOGY, PTS 1 AND 2, 2005, 769 : 261 - 264
  • [40] THE USE OF GAMMA-RAY SPECTROSCOPY IN THE DETERMINATION OF THE GAMMA-RAY DOSE DUE TO RADIOACTIVE FALLOUT PRESENT ON THE GROUND
    GUSTAFSON, PF
    MARINELLI, LD
    BRAR, SS
    RADIATION RESEARCH, 1958, 9 (01) : 125 - 125