Influence of human induced floor vibrations on the measurement precision of force plates in a biomechanics laboratory

被引:0
|
作者
Werkle, H. [1 ]
Firus, A. [2 ]
Francke, W. [1 ]
机构
[1] Hsch Konstanz HTWG, Fak Bauingn, Brauneggerstr 55, D-78462 Constance, Germany
[2] Tech Univ Darmstadt, Inst Stat & Konstrukt, Franziska Braun Str 3, D-64287 Darmstadt, Germany
来源
BAUINGENIEUR | 2016年 / 91卷
关键词
D O I
暂无
中图分类号
TU [建筑科学];
学科分类号
0813 ;
摘要
In gait analysis laboratories the ground reaction force time histories of the footfall of persons are usually measured using a force plate. The accelerations of the floor, in which the force plate is embedded, have to be limited, as they may influence the accuracy of the force measurements. For the numerical simulation of vibrations induced by humans in gait analysis laboratories, loading scenarios are defined. They include continuous motions of persons (walking, running) as well as jumps, typical for biomechanical investigations on athletes. Criteria for the assessment of the measuring error provoked by floor vibrations are given. As example a floor designed to accommodate a force platform in a biomechanical laboratory of the University Hospital in Tubingen, Germany, has been investigated for footfall induced vibrations. The numerical simulation by a finite element analysis has been validated by field measurements. As a result, the measuring error of the force plate installed in the laboratory is obtained for diverse scenarios.
引用
收藏
页码:110 / 119
页数:10
相关论文
共 35 条
  • [21] Robust saturated control of human-induced floor vibrations via a proof-mass actuator
    Diaz, I. M.
    Reynolds, P.
    SMART MATERIALS AND STRUCTURES, 2009, 18 (12)
  • [22] Continuous wave Doppler ultrasound measurement of micro-vibrations induced by a focused acoustic radiation force
    Pavan, T. Z.
    Baggio, A. L.
    Carneiro, A. A. O.
    4TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, 2009, 22 (1-3): : 852 - 855
  • [23] Simulation and Measurement of Human-Induced Vibrations of the Beijing Olympic Watchtower with Tuned Mass Dampers
    Zhang, Zhiqiang
    Guo, Tong
    Yang, Kang
    Ma, Fei
    Li, Aiqun
    JOURNAL OF PERFORMANCE OF CONSTRUCTED FACILITIES, 2017, 31 (06)
  • [24] Motion and force measurement of human fingertips during manual operation to achieve high-precision assembly by articulated robots
    Yamashita T.
    Suzuki H.
    Tasaki R.
    Measurement: Sensors, 2022, 24
  • [25] Influence of the kinematic constraints on dynamic residuals in inverse dynamic analysis during human gait without using force plates
    Juana Mayo
    Joaquin Ojeda
    Multibody System Dynamics, 2020, 50 : 305 - 321
  • [26] Influence of the kinematic constraints on dynamic residuals in inverse dynamic analysis during human gait without using force plates
    Mayo, Juana
    Ojeda, Joaquin
    MULTIBODY SYSTEM DYNAMICS, 2020, 50 (03) : 305 - 321
  • [27] Measurement of human body surface vibrations induced by complex low-frequency noise composed of two pure tones
    Takahashi, Y
    Maeda, S
    JOURNAL OF LOW FREQUENCY NOISE VIBRATION AND ACTIVE CONTROL, 2003, 22 (04): : 209 - 223
  • [28] Numerical investigation of the influence of span-wise force variation in circular cylinders undergoing vortex induced vibrations at high Reynolds number
    Murrin, David
    Militzer, Julio
    Bose, Neil
    Campregher, Rubens
    2007 OCEANS, VOLS 1-5, 2007, : 827 - +
  • [29] Precision of INR (International Normalized Ratio) measurement using laboratory and point-of-care assays: Understanding the impact of system influence on patient management.
    LaDuca, FM
    Gunneman, T
    Ruybalid, R
    Jacobson, AK
    Zucker, ML
    CLINICAL CHEMISTRY, 2003, 49 (06) : A148 - A148
  • [30] Modeling individual-specific human optic nerve head biomechanics. Part I: IOP-induced deformations and influence of geometry
    Ian A. Sigal
    John G. Flanagan
    Inka Tertinegg
    C. Ross Ethier
    Biomechanics and Modeling in Mechanobiology, 2009, 8 : 85 - 98