Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplasty

被引:108
|
作者
Brodner, W
Grübl, A
Jankovsky, R
Meisinger, V
Lehr, S
Gottsauner-Wolf, F
机构
[1] Univ Vienna, Vienna Gen Hosp, Dept Orthopaed, A-1090 Vienna, Austria
[2] Gen Hosp Salzburg, Dept Orthopaed, Salzburg, Austria
[3] Vienna Gen Hosp, Dept Internal Med 4, Clin Dept Occupat Med, Vienna, Austria
[4] Dept Med Comp Sci, Vienna, Austria
[5] Gen Hosp Krems, Dept Orthopaed, Krems Donau, Austria
来源
JOURNAL OF ARTHROPLASTY | 2004年 / 19卷 / 08期
关键词
total hip arthroplasty; metal-on-metal; serum cobalt; serum chromium; cup inclination;
D O I
10.1016/j.arth.2004.09.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We investigated whether cup inclination influences serum levels of cobalt and chromium after metal-on-metal total hip arthroplasty. Out of a consecutive series, 3 groups of 20 patients each were established according to their cup inclination: greatest inclination (55degrees-63degrees, mean 58degrees), intermediate inclination (44degrees-46degrees, mean 45degrees), and smallest inclination (23degrees-37degrees, mean 33degrees). We did not detect any statistically significant difference of serum cobalt (P =.23) or serum chromium (P =.13) levels between the 3 groups applying the Kruskal Wallis test. However , 3 patients with cup inclinations of 58degrees, 63degrees, and 61degrees exhibited 9.8-53.6-fold elevated cobalt and 9.5-30.5-fold elevated chromium levels when compared with the median concentrations of this trial. We therefore recommend accurate CLIP placement also for metal-on-metal articulations. Key words: total hip arthroplasty, metal-on-metal, serum cobalt, serum chromium, cup inclination. (C) 2004 Elsevier Inc. All rights reserved.
引用
收藏
页码:66 / 70
页数:5
相关论文
共 50 条
  • [21] How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing?
    Vendittoli, Pascal-A
    Lavigne, Martin
    Roy, Alain-Guy
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (451) : 292 - 293
  • [22] How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing?
    Back, DL
    Young, AA
    Shimmin, AJ
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (438) : 177 - 181
  • [23] Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing
    Hart, A. J.
    Buddhdev, P.
    Winship, P.
    Faria, N.
    Powell, J. J.
    Skinner, J. A.
    [J]. HIP INTERNATIONAL, 2008, 18 (03) : 212 - 219
  • [24] Serum Metal Ion Concentration: Comparison Between Small and Large Head Metal-on-Metal Total Hip Arthroplasty
    Hallows, Rhett K.
    Pelt, Christopher E.
    Erickson, Jill A.
    Peters, Christopher L.
    [J]. JOURNAL OF ARTHROPLASTY, 2011, 26 (08): : 1176 - 1181
  • [25] Metal-on-metal articulation in total hip arthroplasty
    Marchetti, P.G.
    Binazzi, R.
    Vaccari, V.
    Manca, A.
    Rossi, R.
    Campagna, A.
    Magnani, M.
    Baldarelli, M.
    Pasini, E.
    [J]. 1600, Trans Tech Publ, Uetikon-Zuerich, Switzerland (192-195): : 975 - 977
  • [26] Metal-on-Metal Total Hip Arthroplasty: The Concerns
    MacDonald, S. J.
    [J]. BIOCERAMICS AND ALTERNATIVE BEARINGS IN JOINT ARTHROPLASTY, 2005, : 47 - 47
  • [27] Metal-on-metal articulation in total hip arthroplasty
    Marchetti, PG
    Binazzi, R
    Vaccari, V
    Manca, A
    Rossi, R
    Campagna, A
    Magnani, M
    Baldarelli, M
    Pasini, E
    [J]. BIOCERAMICS, 2000, 192-1 : 975 - 977
  • [28] Serum metal ion levels after second-generation metal-on-metal total hip arthroplasty
    Takao Imanishi
    Masahiro Hasegawa
    Akihiro Sudo
    [J]. Archives of Orthopaedic and Trauma Surgery, 2010, 130 : 1447 - 1450
  • [29] Metal-on-metal total hip arthroplasty - The concerns
    MacDonald, SJ
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) : 86 - 93
  • [30] Serum metal ion levels after second-generation metal-on-metal total hip arthroplasty
    Imanishi, Takao
    Hasegawa, Masahiro
    Sudo, Akihiro
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2010, 130 (12) : 1447 - 1450