Model-based detection of partially obstructed endotracheal tube

被引:14
|
作者
Visaria, RK [1 ]
Westenskow, DR
机构
[1] Univ Utah, Dept Biomed Engn, Salt Lake City, UT 84112 USA
[2] Univ Utah, Dept Anesthesiol, Salt Lake City, UT 84112 USA
关键词
obstructed endotracheal tube; airway obstruction; bronchospasm; stiff chest wall; chest compression; positive end-expiratory pressure; pulmonary complications; lumped pulmonary model; real time identification; methacholine; respiratory impedance;
D O I
10.1097/01.CCM.0000150656.01906.9F
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives. A five-element lumped pulmonary model was developed to estimate respiratory mechanics automatically and noninvasively. The model was applied to diagnose obstructed endotracheal tube. Events like bronchospasm and stiff chest wall were also tested to determine the specificity of the diagnosis. Cases with positive end-expiratory pressure were also included in the analysis to see the effects of positive end-expiratory pressure on the model. Design: Randomized controlled animal study. Setting. University department of anesthesiology. Subjects: Ten anesthetized, paralyzed, and mechanically ventilated mongrel dogs (19-45 kg) of either gender. Interventions. Two levels of upper airway obstruction were induced in ten dogs by partially constricting the endotracheal tube. Acute bronchial constriction was produced in five dogs by injecting methacholine through a central venous catheter. In the same five dogs, the chest wall was stiffened by wrapping a pressure cuff around the chest. Positive end-expiratory pressure was also applied as a separate event in these five animals. Measurements and Main Results: Airway pressure and flow were continuously recorded at the mouth. Model parameters were iteratively identified until the root mean square error between respiratory impedance (obtained from airway pressure and flow) and model-predicted impedance (calculated using Ohm's law) was minimum. The peak inspiratory pressure increased and the peak expiratory flow rate decreased with increasing levels of partial obstruction. The value of the model parameters R-1 and C-2 increased and C-2 decreased with partial obstructed endotracheal tube, whereas R-1 increased and L and C-2 decreased with bronchospasm. With stiff chest wall, R-2 increased and C-2 decreased. With positive end-expiratory pressure, the L parameter decreased and no significant change in other model parameters was observed. Obstructed endotracheal tube is indicated if R, increased greater than or equal to3ft C, decreased greater than or equal to10% and C-2 increased greater than or equal to10% from baseline. The test results using 45 events, including control, three complications, and positive end-expiratory pressure, show that when the model is used to diagnose obstructed endotracheal tube, the method has a sensitivity of 90% and specificity of 97%. Conclusions: During an obstructed endotracheal tube, model parameters change such that the event can be diagnosed noninvasively, automatically, and accurately. The model differentiates between upper airway obstruction and complications like bronchospasm and stiff chest wall.
引用
收藏
页码:149 / 154
页数:6
相关论文
共 50 条
  • [1] Endotracheal Tube Management and Obstructed Airway
    Sancheti, Manu
    Force, Seth
    [J]. THORACIC SURGERY CLINICS, 2015, 25 (03) : 279 - +
  • [2] Shock Due to an Obstructed Endotracheal Tube
    Chua, Stefanie Foong Ling
    Chan, Chi Ho
    Thangavelautham, Suhitharan
    [J]. JOURNAL OF CRITICAL CARE MEDICINE, 2021, 7 (04): : 308 - 311
  • [3] OBSTRUCTED ENDOTRACHEAL TUBE DEMONSTRATED BY ROENTGENOGRAM
    THOMPSON, RC
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1956, 162 (03): : 194 - 196
  • [4] Acute Management of the Obstructed Endotracheal Tube
    Johnson, Kevin M.
    Lehman, Richard E.
    [J]. RESPIRATORY CARE, 2012, 57 (08) : 1342 - 1344
  • [5] Diagnosing partially obstructed endotracheal tubes: The advent of intelligent ventilator monitors
    Simpson, SQ
    [J]. CRITICAL CARE MEDICINE, 2005, 33 (01) : 249 - 250
  • [6] Clearance of an obstructed endotracheal tube with an arterial embolectomy catheter with the patient in the prone position
    Grimmett, WG
    Poh, J
    [J]. ANAESTHESIA AND INTENSIVE CARE, 1998, 26 (05) : 579 - 581
  • [7] Model-Based Detection of CSRF
    Rocchetto, Marco
    Ochoa, Martin
    Dashti, Mohammad Torabi
    [J]. ICT SYSTEMS SECURITY AND PRIVACY PROTECTION, IFIP TC 11 INTERNATIONAL CONFERENCE, SEC 2014, 2014, 428 : 30 - 43
  • [8] Anesthesia case of the month - The endotracheal tube had become obstructed by the contents of the lung abscess
    不详
    [J]. JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2006, 228 (06): : 850 - 853
  • [9] The muscular compliance of the auditory tube: A model-based survey
    Leuwer, R
    Schubert, R
    Kucinski, T
    Liebig, T
    Maier, H
    [J]. LARYNGOSCOPE, 2002, 112 (10): : 1791 - 1795
  • [10] SKELETONISATION AS MODEL-BASED FEATURE DETECTION
    WRIGHT, MW
    FALLSIDE, F
    [J]. IEE PROCEEDINGS-I COMMUNICATIONS SPEECH AND VISION, 1993, 140 (01): : 7 - 11