Delivery of Inhaled Nitric Oxide During MRI to Ventilated Neonates and Infants

被引:1
|
作者
Carter, Bradley G. [1 ]
Swain, Rachel [1 ]
Hislop, Jaime [1 ]
Escudie, Mathilde [1 ]
Williams, Rachel H. [1 ]
机构
[1] Royal Childrens Hosp, Clin Technol Serv, Neonatal & Paediat Intens Care Units, Parkville, Vic, Australia
关键词
nitric oxide; magnetic resonance imaging; mechanical ventilators; critical care; infant; newborn child; BLOOD-FLOW;
D O I
10.4187/respcare.08408
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Many pediatric and neonatal ICU patients receive nitric oxide (NO), with some also requiring magnetic resonance imaging (MRI) scans. MRI-compatible NO delivery devices are not always available. We describe and bench test a method of delivering NO during MRI using standard equipment in which a NO delivery device was positioned in the MRI control room with the NO blender component connected to oxygen and set to 80 ppm and delivering flow via 12 m of tubing to a MRI-compatible ventilator, set up inside the MRI scanner magnet room. METHODS: For our bench test, the ventilator was set up normally and connected to an infant test lung to simulate several patients of differing weight (ie, 4 kg, 10 kg, 20 kg). The NO blender delivered flows of 2-10 L/min to the ventilator to achieve a range of NO and oxygen concentrations monitored via extended tubing. The measured values were compared to calculated values. RESULTS: A range of NO concentrations (12-41 ppm) and F-IO2 values (0.67-0.97) were achieved during the bench testing. The additional flow increased delivered peak inspiratory pressure and PEEP by 1-5 cm H2O. Calculated values were within acceptable ranges and were used to create a lookup table. CONCLUSIONS: In clinical use, this system can safely generate a range of NO flows of 15-42 ppm with an accompanying F-IO2 range of 0.34-0.98. (C) 2021 Daedalus Enterprises.
引用
收藏
页码:1254 / 1262
页数:9
相关论文
共 50 条
  • [41] Inhaled nitric oxide delivery by anesthesia machines
    Ceccarelli, P
    Bigatello, LM
    Hess, D
    Kwo, J
    Melendez, L
    ANESTHESIA AND ANALGESIA, 2000, 90 (02): : 482 - 488
  • [42] Inhaled nitric oxide delivery systems - Reply
    Hess, D
    Imanaka, H
    ANESTHESIOLOGY, 1997, 87 (06) : 1592 - 1593
  • [43] Exhaled nitric oxide in chronically ventilated preterm infants
    Williams, O
    Bhat, RY
    Cheeseman, P
    Rafferty, GF
    Hannam, S
    Greenough, A
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2004, 89 (01): : F88 - F89
  • [44] Value of inhaled nitric oxide in neonates with right ventricular dysfunction
    Saygili, A
    Ledieu, C
    Casterman, P
    Leke, A
    Maingourd, Y
    Krim, G
    ARCHIVES DE PEDIATRIE, 1998, 5 (01): : 93 - 94
  • [45] SAFETY REQUIREMENT FOR USE OF INHALED NITRIC-OXIDE IN NEONATES
    BOUCHET, M
    RENAUDIN, MH
    RAVEAU, C
    MERCIER, JC
    DEHAN, M
    ZUPAN, V
    LANCET, 1993, 341 (8850): : 968 - 969
  • [46] Inhaled nitric oxide (NO) in neonates: Responders and nonresponders. A review
    Krause, M
    Hohn, T
    MONATSSCHRIFT KINDERHEILKUNDE, 1996, 144 (12) : 1352 - 1356
  • [47] Plasma arginine levels and the response to inhaled nitric oxide in neonates
    Kavvadia, V
    Greenough, A
    Lilley, J
    Laubscher, B
    Dimitriou, G
    Boa, F
    Poyser, K
    BIOLOGY OF THE NEONATE, 1999, 76 (06): : 340 - 347
  • [48] Inhaled nitric oxide to prevent bronchopulmonary dysplasia in preterm neonates
    Mercier, Jean-Christophe
    Olivier, Paul
    Loron, Gauthier
    Fontaine, Romain
    Maury, Laure
    Baud, Olivier
    SEMINARS IN FETAL & NEONATAL MEDICINE, 2009, 14 (01): : 28 - 34
  • [49] Inhaled Nitric Oxide Use in Neonates With Congenital Diaphragmatic Hernia
    Campbell, Brendan T.
    Herbst, Katherine W.
    Briden, Kelleigh E.
    Neff, Stephen
    Ruscher, Kimberly A.
    Hagadorn, James I.
    PEDIATRICS, 2014, 134 (02) : E420 - E426
  • [50] Inhaled Nitric Oxide for Respiratory Failure in Preterm Infants
    Soll, Roger F.
    NEONATOLOGY, 2012, 102 (04) : 251 - 253