In vivo measurement of time-resolved autofluorescence at the human fundus

被引:114
|
作者
Schweitzer, D
Hammer, M
Schweitzer, F [1 ]
Anders, R
Doebbecke, T
Schenke, S
Gaillard, ER
Gaillard, ER
机构
[1] Univ Jena, Dept Expt Ophthalmol, Eye Clin, Bachstr 18, D-07740 Jena, Germany
[2] No Illinois Univ, Dept Biochem & Chem, De Kalb, IL 60115 USA
关键词
time-resolved autofluorescence; fluorescence lifetime imaging; human fundus; time-correlated single photon counting; scanning laser ophthalmoscopy; lipofuscin; coenzymes;
D O I
10.1117/1.1806833
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An experimental setup for measurement of time-resolved autofluorescence of the human eye funclus is demonstrated. The method combines laser scanning technique and time-correlated single photon counting. The light source is a laser diode, delivering pulses of about 100 ps duration at a repetition rate of 40 MHz. The excitation wavelength is 446 nm and the cutoff wavelength of fluorescence detection is at 475 nm. The autofluorescence can be determined with a spatial resolution of 80X80 mum(2) and 25 ps time resolution. The fluorescence decay is optimally approximated by a biexponential model. The dominating lifetime lambda1 is shortest in the macula (320 to 380 ps) and reaches 1500 ps in the optic disk. The lifetime lambda2 varies between 2 ns and 5 ns, but the spatial distribution is more homogeneous. Respiration of 100% oxygen for 6 min leads to changes in the fluorescence lifetime pointing to detection of coenzymes. Diagrams of lifetime lambda2 versus -lambda1 are well suited for comparison of substances. Such lifetime clusters of a 20 deg macular field of a young healthy subject and of a patient suffering from dry age-related macular degeneration overlap only partially with lambda2-lambda1 clusters of lipofuscin. (C) 2004 Society of Photo-Optical Instrumentation Engineers.
引用
收藏
页码:1214 / 1222
页数:9
相关论文
共 50 条
  • [41] Time-resolved measurement technology for Z pinch
    Institute of Fluid Physics, China Acad. of Eng. Phys., Mianyang 621900, China
    Guangzi Xuebao, 2007, 5 (922-925): : 922 - 925
  • [42] A new measurement method of time-resolved spectrum
    Shi, Zhi-Gang
    Huang, Shi-Hua
    Liang, Chun-Jun
    Lei, Quan-Sheng
    Guang Pu Xue Yu Guang Pu Fen Xi/Spectroscopy and Spectral Analysis, 2007, 27 (02): : 217 - 219
  • [43] Quantum coherence in the time-resolved Auger measurement
    Smirnova, O
    Yakovlev, VS
    Scrinzi, A
    PHYSICAL REVIEW LETTERS, 2003, 91 (25)
  • [44] Time-resolved measurement system of electrofluorescence spectra
    Tsushima, M
    Ushizaka, T
    Ohta, N
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (02): : 479 - 485
  • [45] MEASUREMENT MODELS FOR TIME-RESOLVED SPECTROSCOPY - A COMMENT
    HOFMANN, HF
    MAHLER, G
    QUANTUM AND SEMICLASSICAL OPTICS, 1995, 7 (04): : 489 - 497
  • [46] Time-resolved ammonia measurement in vehicle exhaust
    Mohn, J
    Forss, AM
    Brühlmann, S
    Zeyer, K
    Lüscher, R
    Emmenegger, L
    Novak, P
    Heeb, N
    INTERNATIONAL JOURNAL OF ENVIRONMENT AND POLLUTION, 2004, 22 (03) : 342 - 356
  • [47] Time-Resolved Spectral Measurement for Fluorescence Analysis
    Saito, Mitsunori
    Koketsu, Takahiro
    Itai, Yusuke
    2019 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2019, : 450 - 454
  • [48] Time-resolved measurement of stimulated polariton relaxation
    Bloch, J
    Sermage, B
    Jacquot, C
    Senellart, P
    Thierry-Mieg, V
    PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE, 2002, 190 (03): : 827 - 831
  • [49] TIME-RESOLVED SPECTROSCOPY OF THE HUMAN FOREARM
    FERRARI, M
    WEI, QN
    CARRARESI, L
    DEBLASI, RA
    ZACCANTI, G
    JOURNAL OF PHOTOCHEMISTRY AND PHOTOBIOLOGY B-BIOLOGY, 1992, 16 (02) : 141 - 153
  • [50] Detection of cell metabolism via wavelength- and time-resolved intracellular autofluorescence
    Wu, Yicong
    Zheng, Wei
    Qu, Jianan Y.
    ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS V, 2007, 6430