Wireless esophageal catheter dedicated to respiratory diseases diagnostic

被引:0
|
作者
Desilets, Tommy [1 ]
Sawan, Mohamad
Bellemare, Francois
机构
[1] Ecole Polytech, Polystim Neurotechnologies Lab, Montreal, PQ, Canada
[2] CHUM Hotel Dieu, Montreal, PQ, Canada
关键词
D O I
暂无
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
We present the design and test of a portable wireless catheter system combining the simultaneous assessment of the transdiaphragmatic pressure and EMG of the diaphragm. The esophageal catheter includes two micro fabricated pressure sensors and five platinum ring electrodes. The low noise analog front-end features an electrode DC mismatch correction circuitry, a selectable gain (58 dB - 97 dB) as well as a high CMRR (80.3 dB). Pressure signals are. sampled at 1 kHz while EMG at 4 kHz. The data is transmitted over a wireless Bluetooth (R) connection towards a computing host to display the received signals in real time. Testing of the device demonstrates its reliability to process pressure and EMC signals and to be used as a stand-alone respiratory parameters assessment device.
引用
收藏
页码:2581 / 2584
页数:4
相关论文
共 50 条
  • [41] The Safety of Dedicated-Team Catheter-Based Diagnostic Cerebral Angiography in the Era of Advanced Noninvasive Imaging
    Thiex, R.
    Norbash, A. M.
    Frerichs, K. U.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (02) : 230 - 234
  • [42] Diagnostic of morphological and functional esophageal disfunction in patients exoposed to radiofrequency catheter ablation of atrial fibrilation
    Sikora, Krzysztof
    Stec, Sebastian
    Tarnowski, Wieslaw
    Sikora, Marek
    Kulakowski, Piotr
    KARDIOLOGIA POLSKA, 2015, 73 (07) : 571 - 571
  • [43] A dedicated protocol and environment for central venous catheter removal in pediatric patients affected by onco-hematological diseases
    Crocoli, Alessandro
    Martucci, Cristina
    Leopardi, Eleonora
    Padua, Mariangela
    Serra, Annalisa
    Cacchione, Antonella
    Coletti, Valentina
    Palumbo, Giuseppe
    Atti, Marta Luisa Ciofi Degli
    Rava, Lucilla
    Inserra, Alessandro
    JOURNAL OF VASCULAR ACCESS, 2014, 15 (06): : 486 - 491
  • [44] Esophageal diseases
    AsherWolf, W.
    Dellon, Evan S.
    Shaheen, Nicholas J.
    ENDOSCOPY, 2014, 46 (09) : 767 - 771
  • [45] Esophageal diseases
    Wolf, W. Asher
    Dellon, Evan S.
    Shaheen, Nicholas J.
    GASTROINTESTINAL ENDOSCOPY, 2014, 80 (03) : 374 - 379
  • [46] Dedicated catheter lumens for parenteral nutrition - Reply
    O'Grady, NP
    CLINICAL INFECTIOUS DISEASES, 2003, 36 (12) : 1626 - 1626
  • [47] Wearable bioelectronic masks for wireless detection of respiratory infectious diseases by gaseous media
    Wang, Bingfang
    Yang, Deqi
    Chang, Zhiqiang
    Zhang, Ru
    Dai, Jing
    Fang, Yin
    MATTER, 2022, 5 (12) : 4347 - 4362
  • [48] Esophageal Manometry in Pediatric Acute Respiratory Distress Syndrome: Necessity or Simply a Neighbor of the Pulmonary Artery Catheter?
    Napolitano, Natalie
    Yehya, Nadir
    Keim, Garrett
    PEDIATRIC CRITICAL CARE MEDICINE, 2025, 26 (03) : e393 - e395
  • [49] Neural respiratory drive during CO2 rebreathing assessed by a combined esophageal electrode catheter
    Tan, Kangxin
    Lu, DongXiao
    Luo, Yuanming
    MEDICAL MATERIALS AND ENGINEERING, 2012, 140 : 191 - 194
  • [50] CLASSIC VERSUS SEQUENTIAL DIAGNOSTIC SUPPORT FOR CHRONIC NONSPECIFIC RESPIRATORY-DISEASES
    LIEBHART, J
    KRUSINSKA, E
    MEDICAL DECISION MAKING, 1993, 13 (02) : 103 - 113