Finite element analysis of polyethylene wear in total hip replacement: A literature review

被引:21
|
作者
Wang, Lin [1 ,2 ]
Isaac, Graham [2 ]
Wilcox, Ruth [2 ]
Jones, Alison [2 ]
Thompson, Jonathan [1 ,2 ]
机构
[1] DePuy Synthes Joint Reconstruct, Hip Dev, Worldwide Res & Dev, St Anthonys Rd, Leeds LS11 8DT, W Yorkshire, England
[2] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
Artificial hip joint; biotribology; contact modelling; wear modelling; wear mechanics; MOLECULAR-WEIGHT POLYETHYLENE; CROSS-LINKED POLYETHYLENE; FEMORAL-HEAD PENETRATION; JOINT SIMULATOR; ACETABULAR CUPS; CONTACT STRESS; RELATIVE MOTION; STAINLESS-STEEL; UHMWPE WEAR; KNEE-JOINT;
D O I
10.1177/0954411919872630
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Evaluation and prediction of wear play a key role in product design and material selection of total hip replacements, because wear debris is one of the main causes of loosening and failure. Multifactorial clinical or laboratory studies are high cost and require unfeasible timeframes for implant development. Simulation using finite element methods is an efficient and inexpensive alternative to predict wear and pre-screen various parameters. This article presents a comprehensive literature review of the state-of-the-art finite element modelling techniques that have been applied to evaluate wear in polyethylene hip replacement components. A number of knowledge gaps are identified including the need to develop appropriate wear coefficients and the analysis of daily living activities.
引用
收藏
页码:1067 / 1088
页数:22
相关论文
共 50 条
  • [21] Finite Element Analysis of Stress Distribution of Cemented Stem After Total Hip Replacement
    Zheng, Xiao-Wen
    2009 3RD INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING, VOLS 1-11, 2009, : 2765 - 2768
  • [22] Computational modeling of polyethylene wear in total hip arthroplasty using patient-specific kinematics-coupled finite element analysis
    Peng, Yun
    Arauz, Paul
    An, Shuai
    Kwon, Young-Min
    TRIBOLOGY INTERNATIONAL, 2019, 129 : 162 - 166
  • [23] Finite element simulation of early creep and wear in total hip arthroplasty
    Bevill, SL
    Bevill, GR
    Penmetsa, JR
    Petrella, AJ
    Rullkoetter, PJ
    JOURNAL OF BIOMECHANICS, 2005, 38 (12) : 2365 - 2374
  • [24] How does bicycling affect the longevity of Total Hip Arthroplasty? A finite element wear analysis
    Toh, Shawn Ming Song
    Ashkanfar, Ariyan
    English, Russell
    Rothwell, Glynn
    Langton, David J.
    Joyce, Thomas J.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2023, 139
  • [25] Local head roughening as a factor contributing to variability of total hip wear: A finite element analysis
    Brown, TD
    Stewart, KJ
    Nieman, JC
    Pedersen, DR
    Callaghan, JJ
    JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2002, 124 (06): : 691 - 698
  • [26] The effect of femoral offset on polyethylene wear in total hip arthroplasty: a CT-based finite element analysis of 15 patients with simulated gait
    Rudiger, Hannes
    Latypova, Adeliya
    Terrier, Alexandre
    SWISS MEDICAL WEEKLY, 2018, 148 : 38S - 38S
  • [27] Experimental assessment of precision and accuracy of radio stereometric analysis for the determination of polyethylene wear in a total hip replacement model
    Bragdon, CR
    Malchau, H
    Yuan, XH
    Perinchief, R
    Kärrholm, J
    Börlin, N
    Estok, DM
    Harris, WH
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (04) : 688 - 695
  • [28] Hypertrophy Training Following A Total Hip Replacement: A Literature Review
    Bull, Tyson
    Erzen, Andrew
    O'Donnell, John
    Rafla, Michael
    Georgy, David
    Bailey, Mia
    Takla, Amir
    INTERNATIONAL JOURNAL OF SPORTS PHYSICAL THERAPY, 2024, 19 (03): : 337 - 350
  • [29] Polyethylene wear in total hip and knee arthroplasties
    Lewis, G
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1997, 38 (01): : 55 - 75
  • [30] Polyethylene wear in total hip and knee arthroplasties
    Department of Mechanical Engineering, University of Memphis, Memphis, TN 38152, United States
    J. BIOMED. MATER. RES., 1 (55-75):