The Influence of Free Hemoglobin and Bilirubin on Heparin Monitoring by Activated Partial Thromboplastin Time and Anti-Xa Assay

被引:56
|
作者
Kostousov, Vadim [1 ,3 ]
Nguyen, Kim [3 ]
Hundalani, Shilpa G. [2 ,4 ]
Teruya, Jun [1 ,2 ,3 ]
机构
[1] Baylor Coll Med, Dept Pathol & Immunol, Houston, TX 77030 USA
[2] Baylor Coll Med, Houston, TX 77030 USA
[3] Texas Childrens Hosp, Div Transfus Med & Coagulat, Houston, TX 77030 USA
[4] Texas Childrens Hosp, Dept Pediat, Sect Neonatol, Houston, TX 77030 USA
关键词
EXTRACORPOREAL MEMBRANE-OXYGENATION; COAGULATION MANAGEMENT; CENTRIFUGAL PUMPS; HEMOLYSIS; ANTICOAGULATION; INTERFERENCE;
D O I
10.5858/arpa.2013-0572-OA
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Context.-Elevated free hemoglobin (Hb) and bilirubinemia complicate extracorporeal membrane oxygenation and could affect unfractionated heparin (UH) therapy monitoring by anti-Xa assay and activated partial thromboplastin time (aPTT). Objectives.-To compare in vitro response of anti-Xa and aPTT assays to UH in samples with artificial hyperbilirubinemia and hyperhemoglobinemia and to estimate if this interference is also observed in vivo in pediatric extracorporeal membrane oxygenation. Design.-Measurement of aPTT and anti-Xa activity in plasma spiked with UH and increased concentration of free Hb and/or conjugated bilirubin. All samples with antiXa activity, antithrombin, free Hb, and bilirubin determination and infused dose of UH from inpatients on extracorporeal membrane oxygenation were extracted from the clinical patient database and analyzed. Results.-Each increment of free Hb by 100 mg/dL significantly shortened aPTT, whereas an increment of bilirubin by 6 mg/dL caused significant prolongation of aPTT and stepwise increase of free Hb and/or bilirubin in plasma decreased anti-Xa activity by 0.03 to 0.05 IU/mL. Extracorporeal membrane oxygenation samples with free Hb 50 mg/dL or greater had significantly lower anti-Xa activity compared with normal ones: 0.33 (0.25-0.42) versus 0.4 (0.31-0.48) IU/mL (P = .01), despite the identical UH infusion and similar antithrombin activity. Moderate increase of free Hb by 59 mg/dL was associated with absolute decrease of anti-Xa activity by 0.07 IU/mL. Conclusions.-Activated partial thromboplastin time and anti-Xa assay are affected by elevated level of free Hb and/or bilirubin in the presence of UH, and lower antiXa activity is noted in extracorporeal membrane oxygenation patients with elevated free Hb. Severe hemolysis and/or hyperbilirubinemia could compromise UH monitoring based on these assays.
引用
收藏
页码:1503 / 1506
页数:4
相关论文
共 50 条
  • [31] Variation in coagulation factor activity levels cause discrepancies between activated partial thromboplastin time and anti-Xa activity for heparin monitoring: a retrospective observational study
    Saito, Tomoyo
    Hayakawa, Mineji
    Kumano, Osamu
    Honma, Yoshinori
    Murashita, Mone
    Kato, Jun
    Fukui, Syouki
    Takahashi, Masaki
    Takahashi, Yuki
    Tsuchida, Takumi
    Mizugaki, Asumi
    Takauji, Shuhei
    Hayamizu, Mariko
    Yoshida, Tomonao
    Katabami, Kenichi
    Wada, Takeshi
    Maekawa, Kunihiko
    JOURNAL OF INTENSIVE CARE, 2023, 11 (01)
  • [32] Performance of Anti-Factor Xa Versus Activated Partial Thromboplastin Time for Heparin Monitoring Using Multiple Nomograms
    Whitman-Purves, Emily
    Coons, James C.
    Miller, Taylor
    DiNella, Jeannine V.
    Althouse, Andrew
    Schmidhofer, Mark
    Smith, Roy E.
    CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 2018, 24 (02) : 310 - 316
  • [33] Use of the activated partial thromboplastin time for heparin monitoring
    Smythe, MA
    Koerber, JM
    Westley, SJ
    Nowak, SN
    Begle, RL
    Balasubramaniam, M
    Mattson, JC
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2001, 115 (01) : 148 - 155
  • [34] Heparin monitoring in sheep by activated partial thromboplastin time
    Park, SG
    Kim, SC
    Choi, MJ
    Lee, HS
    Min, BG
    Cheong, J
    Lee, K
    ARTIFICIAL ORGANS, 2003, 27 (06) : 576 - 580
  • [35] MONITORING HEPARIN THERAPY WITH ACTIVATED PARTIAL THROMBOPLASTIN TIME
    STUART, RK
    MICHEL, A
    CANADIAN MEDICAL ASSOCIATION JOURNAL, 1971, 104 (05) : 385 - &
  • [36] Evaluation of a new anti-Xa heparin assay
    Davidson, S. J.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2009, 7 : 996 - 996
  • [37] Comparing anti-factor Xa and activated partial thromboplastin levels for monitoring unfractionated heparin.
    Wahking, Rebekah A.
    Hargreaves, Rachel H.
    Lockwood, Sean M.
    Haskell, Sabrina K.
    Davis, Kelly
    PHARMACOTHERAPY, 2018, 38 (07): : E54 - E54
  • [38] Interlaboratory Precision in the Monitoring of Unfractionated Heparin Using the Anti-Factor Xa-Correlated Activated Partial Thromboplastin Time
    Cuker, Adam
    Ptashkin, Beverly
    Konkle, Barbara A.
    Pipe, Steven W.
    Whinna, Herbert C.
    Zheng, X. Long
    Cines, Douglas B.
    Pollak, Eleanor
    BLOOD, 2008, 112 (11) : 166 - 166
  • [39] Interlaboratory agreement in the monitoring of unfractionated heparin using the anti-factor Xa-correlated activated partial thromboplastin time
    Cuker, A.
    Ptashkin, B.
    Konkle, B. A.
    Pipe, S. W.
    Whinna, H. C.
    Zheng, X. L.
    Cines, D. B.
    Pollak, E. S.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2009, 7 (01) : 80 - 86
  • [40] Discordant Partial Thromboplastin Time (PTT) vs Anti-Xa Heparin Activity: Using the Baseline PTT to Improve Concordance