Multi-site performance evaluation and Sigma metrics of 20 assays on the Atellica chemistry and immunoassay analyzers

被引:15
|
作者
Fasano, Tommaso [2 ]
Luis Bedini, Jose [3 ]
Fle, Pierre-Antoine [4 ]
Jlaiel, Malik [4 ]
Hubbert, Karl [5 ]
Datta, Harish [5 ]
Chicha-Cattoir, Valerie [6 ]
Mansour, Hayfa [7 ]
Mira, Aurea [3 ]
Gonzalez de la Presa, Bernardino [3 ]
Rico, Nayra [3 ]
Sanz de Pedro, Maria [1 ]
Diaz-Garzon, Jorge [1 ]
Buno Soto, Antonio [1 ]
Vecchia, Luigi [2 ]
Peoc'h, Katell [6 ]
机构
[1] Hosp Univ La Paz, Paseo Castellana 261, Madrid 28046, Spain
[2] Azienda Unita Sanit Locale IRCCS, Reggio Emilia, Italy
[3] Hosp Clin Barcelona, Barcelona, Spain
[4] Lab Biol Med Bioesterel, Mandelieu La Napoule, France
[5] James Cook Univ Hosp, Middlesbrough, Cleveland, England
[6] Hop Beaujon, APHP, HUPNVS, Clichy, France
[7] Hop Bichat Claude Bernard, APHP, HUPNVS, Paris, France
关键词
Atellica (R); chemistry; immunoassays; method comparison; performance; precision; BIOLOGICAL VARIATION;
D O I
10.1515/cclm-2019-0699
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The Atellica Solution comprises chemistry (CH) and immunoassay (IM) analyzers. Recently, six early adopter clinical laboratories across Europe evaluated the analytical performance of 20 CH and IM assays. To measure analytical performance quality, Sigma metrics were calculated for individual-site and pooled-site results. Methods: Precision, detection capability, linearity, and method comparison studies were performed according to Clinical Laboratory Standards Institute protocols. Global Sigma metrics across sites were calculated from pooled data at the medical decision level using total allowable error (TEa) goals from CLIA for CH assays, and TEa goals from Ri1iBAK for IM assays; and, the equation: Sigma metrics = %TEa - %bias/%CV. A pooled %CV was calculated by combining the imprecision obtained from individual sites. Bias calculations were performed against the ADVIA Chemistry system or ADVIA Centaur system using Deming regression analysis (Passing-Bablok regression for electrolytes) on the pooled-site data. The 103 individual-site Sigma metric calculations used individual-site imprecision and pooled-bias. Results: The limits of blank and detection results agreed with the manufacturer's claims. Most assays were linear across the assay range tested. Pooled Sigma metrics were good or better (>4 Sigma) for 18 of 20 assays; and, acceptable for urea nitrogen (3.1) and sodium (3.9), the latter values attributable to higher imprecision at one of five sites. Conclusions: Sigma metrics for data generated across multiple real-world sites evaluating the Atellica Solution demonstrated good or better performance of greater than 4 Sigma for 18 of 20 assays tested. Overall, results verified the manufacturer's claims that methods were fit for use in clinical laboratories.
引用
收藏
页码:59 / 68
页数:10
相关论文
共 38 条
  • [31] Brief educational interventions to improve performance on novel quality metrics in ambulatory settings in Kenya: A multi-site pre-post effectiveness trial
    Korom, Robert Ryan
    Onguka, Stephanie
    Halestrap, Peter
    McAlhaney, Maureen
    Adam, Mary
    PLOS ONE, 2017, 12 (04):
  • [32] Multi-site performance evaluation of the Alinity m Molecular assay for quantifying Epstein-Barr virus DNA in plasma samples
    Wessels, Els
    Albert, Eliseo
    Vreeswijk, Tom
    Claas, Eric C. J.
    Gimenez, Estela
    Reinhardt, Birgit
    Sasaki, Mark M.
    Navarro, David
    JOURNAL OF CLINICAL MICROBIOLOGY, 2023, 61 (10)
  • [33] Performance evaluation of ICA-based spike sorting methods to multi-unit data recorded by a multi-site electrode in the rat hippocampus in vivo
    Shiraishi, Yasushi
    Katayama, Norihiro
    Takasho, Koichi
    Karashima, Akihiro
    Nakao, Mitsuyuki
    NEUROSCIENCE RESEARCH, 2010, 68 : E440 - E440
  • [34] Multi-site verification and performance of body fluids in-between Horiba Medical Yumizen h2500 with Sysmex XN-10 automated hematology analyzers and manual microscopy for morphology assessment
    Hege, M. L.
    CLINICAL CHEMISTRY, 2024, 70 : I59 - I59
  • [35] Performance evaluation of a multi-site weather generator coupling maximum entropy resampling for estimating the probability distribution of annual maximum daily rainfall in the Loess Plateau
    Dengrui Mu
    Hongbo Zhang
    Shiliang Xie
    Chiheng Dang
    Shuqi Zhang
    Congcong Yao
    Yu Zhang
    Fengguang Lyu
    Stochastic Environmental Research and Risk Assessment, 2024, 38 : 1251 - 1269
  • [36] Performance evaluation of a multi-site weather generator coupling maximum entropy resampling for estimating the probability distribution of annual maximum daily rainfall in the Loess Plateau
    Mu, Dengrui
    Zhang, Hongbo
    Xie, Shiliang
    Dang, Chiheng
    Zhang, Shuqi
    Yao, Congcong
    Zhang, Yu
    Lyu, Fengguang
    STOCHASTIC ENVIRONMENTAL RESEARCH AND RISK ASSESSMENT, 2024, 38 (04) : 1251 - 1269
  • [37] Multi-site evaluation of the new IFCC-traceable ortho-clinical diagnostic vitros ALTv and LDHi slides: Can traceability improve performance and allow reference interval harmonization between vitros and Beckman platforms?
    Raizman, J. E.
    Goudreau, B.
    Fuzery, A. K.
    Tsui, A. K.
    CLINICA CHIMICA ACTA, 2022, 530 : S277 - S278
  • [38] A Commitment to HIV Diagnostic Accuracy a comment on "Towards more accurate HIV testing in sub-Saharan Africa: a multi-site evaluation of HIV RDTs and risk factors for false positives 'and' HIV misdiagnosis in sub-Saharan Africa: a performance of diagnostic algorithms at six testing sites"
    Del Solar, Amy S. Kravitz
    Parekh, Bharat
    Douglas, Meaghan O'Keefe
    Edgil, Dianna
    Kuritsky, Joel
    Nkengasong, John
    JOURNAL OF THE INTERNATIONAL AIDS SOCIETY, 2018, 21 (08)