Multi-site performance evaluation and Sigma metrics of 20 assays on the Atellica chemistry and immunoassay analyzers

被引:15
|
作者
Fasano, Tommaso [2 ]
Luis Bedini, Jose [3 ]
Fle, Pierre-Antoine [4 ]
Jlaiel, Malik [4 ]
Hubbert, Karl [5 ]
Datta, Harish [5 ]
Chicha-Cattoir, Valerie [6 ]
Mansour, Hayfa [7 ]
Mira, Aurea [3 ]
Gonzalez de la Presa, Bernardino [3 ]
Rico, Nayra [3 ]
Sanz de Pedro, Maria [1 ]
Diaz-Garzon, Jorge [1 ]
Buno Soto, Antonio [1 ]
Vecchia, Luigi [2 ]
Peoc'h, Katell [6 ]
机构
[1] Hosp Univ La Paz, Paseo Castellana 261, Madrid 28046, Spain
[2] Azienda Unita Sanit Locale IRCCS, Reggio Emilia, Italy
[3] Hosp Clin Barcelona, Barcelona, Spain
[4] Lab Biol Med Bioesterel, Mandelieu La Napoule, France
[5] James Cook Univ Hosp, Middlesbrough, Cleveland, England
[6] Hop Beaujon, APHP, HUPNVS, Clichy, France
[7] Hop Bichat Claude Bernard, APHP, HUPNVS, Paris, France
关键词
Atellica (R); chemistry; immunoassays; method comparison; performance; precision; BIOLOGICAL VARIATION;
D O I
10.1515/cclm-2019-0699
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The Atellica Solution comprises chemistry (CH) and immunoassay (IM) analyzers. Recently, six early adopter clinical laboratories across Europe evaluated the analytical performance of 20 CH and IM assays. To measure analytical performance quality, Sigma metrics were calculated for individual-site and pooled-site results. Methods: Precision, detection capability, linearity, and method comparison studies were performed according to Clinical Laboratory Standards Institute protocols. Global Sigma metrics across sites were calculated from pooled data at the medical decision level using total allowable error (TEa) goals from CLIA for CH assays, and TEa goals from Ri1iBAK for IM assays; and, the equation: Sigma metrics = %TEa - %bias/%CV. A pooled %CV was calculated by combining the imprecision obtained from individual sites. Bias calculations were performed against the ADVIA Chemistry system or ADVIA Centaur system using Deming regression analysis (Passing-Bablok regression for electrolytes) on the pooled-site data. The 103 individual-site Sigma metric calculations used individual-site imprecision and pooled-bias. Results: The limits of blank and detection results agreed with the manufacturer's claims. Most assays were linear across the assay range tested. Pooled Sigma metrics were good or better (>4 Sigma) for 18 of 20 assays; and, acceptable for urea nitrogen (3.1) and sodium (3.9), the latter values attributable to higher imprecision at one of five sites. Conclusions: Sigma metrics for data generated across multiple real-world sites evaluating the Atellica Solution demonstrated good or better performance of greater than 4 Sigma for 18 of 20 assays tested. Overall, results verified the manufacturer's claims that methods were fit for use in clinical laboratories.
引用
收藏
页码:59 / 68
页数:10
相关论文
共 38 条
  • [1] Evaluation of the Analytical Performance of Five Chemistry Assays on the Atellica CI Analyzer
    Lewisch, S.
    Camacho, J.
    Wilkins, G.
    Gupta, P.
    Shalhoub, V.
    Snyder, J.
    CLINICAL CHEMISTRY, 2024, 70
  • [2] Analytical performance evaluation of five chemistry assays on the Atellica CI analyzer
    Gupta, P.
    Rhea-Mcmanus, J.
    Camacho, J.
    Wilkins, G.
    Shalhoub, V.
    Joy, A.
    Snyder, J.
    CLINICA CHIMICA ACTA, 2024, 558 : 70 - 71
  • [3] Evaluation of the Analytical Performance of Atellica CH Prealbumin and β2-Microglobulin Chemistry Assays on the Atellica CI Analyzer
    Huff, J.
    Varhue, W.
    Gupta, P.
    Shalhoub, V.
    Camacho, J.
    Snyder, J.
    CLINICAL CHEMISTRY, 2024, 70 : I19 - I19
  • [4] A novel Sigma metric encompasses global multi-site performance of 18 assays on the Abbott Alinity system
    Taher, Jennifer
    Cosme, Jake
    Renley, Brian A.
    Daghfal, David J.
    Yip, Paul M.
    CLINICAL BIOCHEMISTRY, 2019, 63 : 106 - 112
  • [5] Analytical performance evaluation of magnesium and calcium general chemistry assays on the ATELLICA CI Analyzer
    Rhea-Mcmanus, J.
    Gupta, P.
    Varhue, W.
    Joy, A.
    Shalhoub, V.
    Camacho, J.
    Snyder, J.
    CLINICA CHIMICA ACTA, 2024, 558
  • [6] Evaluation of the Analytical Performance of Magnesium and Calcium General Chemistry Assays on the Atellica CI Analyzer
    Carter, R. H., Jr.
    Varhue, W.
    Gupta, P.
    Shalhoub, V.
    Camacho, J.
    Snyder, J.
    CLINICAL CHEMISTRY, 2024, 70
  • [7] Analytical performance evaluation of Prealbumin and ®2-microglobulin chemistry assays on the Atellica CI Analyzer
    Rhea-Mcmanus, J.
    Varhue, W.
    Gupta, P.
    Joy, A.
    Shalhoub, V.
    Camacho, J.
    Snyder, J.
    CLINICA CHIMICA ACTA, 2024, 558
  • [8] A multi-site evaluation of the new sIgE assays on the ADVIA Centaur.
    Plebani, M
    Faggian, D
    Westerhuis, L
    Dillon, P
    Robinson, PS
    CLINICAL CHEMISTRY, 2001, 47 (06) : A48 - A49
  • [9] Performance Evaluation of a Live Multi-Site LTE Network
    Sezgin, Gediz
    Coskun, Yagmur
    Basar, Ertugrul
    Kurt, Gunes Karabulut
    IEEE ACCESS, 2018, 6 : 49690 - 49704
  • [10] EVALUATION OF STUDENT PERFORMANCE IN A MULTI-SITE CLINICAL CLERKSHIP
    IRBY, DM
    VONTVER, LA
    STENCHEVER, MA
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1980, 136 (08) : 1020 - 1023