Comparison of thin-strut cobalt-chromium stents and stainless steel stents in a porcine model of neointimal hyperplasia

被引:0
|
作者
Milewski, Krzysztof [1 ]
Zurakowski, Aleksander [1 ]
Pajak, Jacek [2 ]
Zielinska-Pajak, Ewa [2 ]
Liszka, Lukasz [2 ]
Buszman, Piotr P. [1 ,3 ]
Bis, Jaroslaw [1 ]
Debinski, Marcin [1 ]
Buszman, Pawel [1 ,3 ]
机构
[1] Med Univ Silesia, Upper Silesian Ctr Cardiol, PL-40635 Katowice, Poland
[2] Med Univ Silesia, Dept Histopathol, PL-40635 Katowice, Poland
[3] Anim Cath Lab, Kostkowice, Poland
来源
MEDICAL SCIENCE MONITOR | 2010年 / 16卷 / 01期
关键词
stent; cobalt-chromium; intimal hyperplasia; porcine; BALLOON ANGIOPLASTY; ELUTING STENT; IMPLANTATION; RESTENOSIS;
D O I
暂无
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: The high radial force and durability of cobalt-chromium alloy enable the construction of low-profile stents with thin struts, which improves their elasticity and may play a key role in reducing the incidence of neointimal hyperplasia as well as allow faster endothelialization. The aim of this study was to compare cobalt-chromium (CoCr) and stainless steel (SS) coronary stents in a pig model of neointimal hyperplasia. Material/Methods: Eighteen stents were implanted into the coronary arteries of nine pigs. Control coronarography was performed 28 days after stent implantation. The animals were then sacrificed, their hearts explanted, and the coronary arteries isolated for further histopathological analysis. Results: Quantitative coronary angiography and histomorphometric analysis revealed no statistically significant difference between the two groups of stents. However, there was a trend toward greater late lumen loss (p=0.09) and neointimal area in the CoCr stents (p=0.08). The qualitative histopathology of the CoCr stents revealed findings typical of bare metal stents reviewed in the literature and approved for use in clinical practice. No signs of stent thrombosis, necrosis, or fibrin deposits were observed nor signs of excessive inflammatory reaction. Endothelialization was complete within 28 days. Conclusions: Despite thinner struts, CoCr stents have no advantage over stainless steel stents in terms of neointimal hyperplasia inhibition. However, their positive safety results together with their high radial strength, low profile, and excellent elasticity can ensure their usage, especially in lesions of complex morphology.
引用
收藏
页码:BR40 / BR44
页数:5
相关论文
共 50 条
  • [1] A novel thin-strut cobalt-chromium stent for inhibition of neointimal formation in a porcine coronary artery model
    Hou, D.
    Shinke, T.
    Jabra, R.
    Li, J.
    Chen, J.
    Venegoni, A.
    Colley, K.
    Klein, R.
    Chronos, N.
    Robinson, K.
    [J]. EUROPEAN HEART JOURNAL, 2008, 29 : 333 - 333
  • [2] Neointimal hyperplasia in a thin-strut cobalt-chromium stent: Insights from detailed 3-D intravascular ultrasound analysis
    Otake, Hiromasa
    Ako, Junya
    Waseda, Katsuhisa
    Sakurai, Ryota
    Hirohata, Atsushi
    Kaneda, Hideaki
    Hasegawa, Takao
    Honda, Yasuhiro
    Fitzgerald, Peter J.
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 2010, 145 (01) : 125 - 126
  • [3] Effect on Clinical Restenosis of an Ultra-Thin-Strut Bare Metal Cobalt-Chromium Stent Versus a Thin-Strut Stainless Steel Stent
    Silenzi, Simona
    Grossi, Pierfrancesco
    Mariani, Luca
    Fraccaro, Chiara
    Vagnarelli, Fabio
    Moretti, Luciano
    [J]. JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2016, 29 (03) : 300 - 310
  • [4] Multi-scale mechanical investigation of stainless steel and cobalt-chromium stents
    Kapnisis, Konstantinos
    Constantinides, Georgios
    Georgiou, Harry
    Cristea, Daniel
    Gabor, Camelia
    Munteanu, Daniel
    Brott, Brigitta
    Anderson, Peter
    Lemons, Jack
    Anayiotos, Andreas
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 40 : 240 - 251
  • [5] Long-Term Clinical Outcome of Thin-Strut Cobalt-Chromium Stents in the Drug-Eluting Stent Era: Results of the COBALT (Comparison of Bare-Metal Stents in All-Comers' Lesion Treatment) Registry
    Abdel-Wahab, Mohamed
    Toelg, Ralph
    Kassner, Guido
    Klatt, Lothar
    Sherif, Mohammad A.
    Geist, Volker
    Richardt, Gert
    [J]. JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2011, 24 (06) : 496 - 504
  • [6] Cobalt-chromium stents in porcine coronary arteries: angiographic and ultrasonographic outcomes
    Buszman, P.
    Zurakowski, A.
    Milewski, K.
    [J]. EUROPEAN HEART JOURNAL, 2005, 26 : 417 - 417
  • [7] BiOSS LIM C: thin-strut cobalt-chromium version of the dedicated bifurcation stent
    Gil, Robert J.
    Bil, Jacek
    Kaczynski, Mateusz
    Milewski, Krzysztof P.
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2017, 14 (04) : 279 - 284
  • [8] Methylprednisolone coated stents decrease neointimal hyperplasia in a porcine coronary model
    Ping, QB
    Yanming, H
    Wang, L
    Vermeire, I
    Verbeken, E
    Schacht, E
    De Scheerder, IK
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) : 74A - 74A
  • [9] Methylprednisolone coated stents decrease neointimal hyperplasia in a porcine coronary model
    De Scheerder, I
    Huan, Y
    Ping, QB
    Wang, L
    Vermeire, I
    Verbeken, E
    Schacht, E
    [J]. CIRCULATION, 2000, 102 (18) : 813 - 814
  • [10] Methylprednisolone coated stents decrease neointimal hyperplasia in a porcine coronary model
    Haung, Y
    Qiang, B
    Wang, L
    Vermeire, I
    Schacht, E
    De Scheerder, I
    Verbeken, E
    [J]. EUROPEAN HEART JOURNAL, 2001, 22 : 483 - +