Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients

被引:98
|
作者
Bur, A [1 ]
Herkner, H
Vlcek, M
Woisetschläger, C
Derhaschnig, U
Delle Karth, G
Laggner, AN
Hirschl, MM
机构
[1] Univ Clin, Vienna Gen Hosp, Dept Emergency Med, Vienna, Austria
[2] Univ Clin, Vienna Gen Hosp, Dept Cardiol, Vienna, Austria
关键词
oscillometric blood pressure measurement; cuff size; critically ill patients; emergency medicine; intensive care medicine; noninvasive; monitoring; upper arm circumference; mean arterial blood pressure; catecholamine treatment; mechanical ventilation;
D O I
10.1097/01.CCM.0000053650.12025.1A
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: Comparison of oscillometric blood pressure measurement with two different devices (M3000A using a new algorithm and M1008A using an established algorithm, both Hewlett Packard) and evaluation of current recommendations concerning the relation between cuff size and upper arm circumference in critically ill patients. Design: Prospective data collection. Setting: Emergency department in a 2000-bed inner-city hospital. Patients: A total of 30 patients categorized into three groups according to their upper arm circumference (I, 18-25 cm; II, 25.1-33 cm; III, 33.1-47.5 cm) were enrolled in the study protocol. Interventions: In each patient, two noninvasive blood pressure devices with three different cuff sizes were used to perform oscillometric blood pressure measurement. Invasive mean arterial blood pressure measurement was done by cannulation of the radial artery. Measurement and Main Results: Overall, 1,011 pairs of simultaneous oscillometric and invasive blood pressure measurements were collected in 30 patients (group I, n = 10; group II, n = 10; group III, n = 10). The overall discrepancy between both methods with the M3000A was -2.4 +/- 11.8 mm Hg (p < .0001) and, with the M1008A, -5.3 +/- 11.6 mm Hg (p < .0001) if the recommended cuff size according to the upper arm circumference was used (352 measurements). If smaller cuff sizes than recommended were used (308 measurements performed in group 11 and 111), the overall discrepancy between both methods with the M3000A was 1.3 +/- 13.4 mm Hg (p < .024) and, with the M1008A, -2.3 +/- 11.5 mm Hg (p < .0001). Conclusion: The new algorithm reduced the overall bias of the oscillometric method but still showed a significant discrepancy between both methods of blood pressure measurement, primarily due to the mismatch between upper arm circumference and cuff size. The improvement of the algorithm alone could not result in a sufficient improvement of oscillometric blood pressure measurement. A reevaluation of the recommendations concerning the relation between upper arm circumference and cuff size are urgently required if oscillometric blood pressure measurement should become a reasonable alternative to intra-arterial blood pressure measurement in critically ill patients.
引用
收藏
页码:793 / 799
页数:7
相关论文
共 50 条
  • [1] ACCURACY OF OSCILLOMETRIC BLOOD-PRESSURE MEASUREMENT IN CRITICALLY ILL NEONATES
    GEVERS, M
    VANGENDERINGEN, HR
    LAFEBER, HN
    HACK, WW
    [J]. PEDIATRIC RESEARCH, 1994, 36 (01) : A15 - A15
  • [2] Accuracy of oscillometric blood pressure measurement in critically ill neonates with reference to the arterial pressure wave shape
    Gevers, M
    vanGenderingen, HR
    Lafeber, HN
    Hack, WWM
    [J]. INTENSIVE CARE MEDICINE, 1996, 22 (03) : 242 - 248
  • [3] Accuracy Of Oscillometric Blood Pressure Monitoring In Obese Critically Ill Patients Receiving Vasopressors
    Okoye, O.
    Long, C.
    Delk, M.
    Eissa, K.
    Simeone, F.
    Kheir, F.
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2016, 193
  • [4] Factors influencing accuracy of blood glucose measurements in critically ill patients
    J Peng
    Y Liu
    Y Meng
    X Song
    L Weng
    B Du
    [J]. Critical Care, 12 (Suppl 2):
  • [5] Accuracy and precision of oscillometric noninvasive blood pressure measurement in critically ill patients: systematic review and meta-analysis
    Nedel, Wagner Luis
    Vasconcellos, Alessandro Tonin
    Gunsch, Karoline Ariane
    Soares, Pedro Henrique Rigotti
    [J]. ANAESTHESIOLOGY INTENSIVE THERAPY, 2022, 54 (05) : 425 - 431
  • [6] Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients
    Bur, A
    Hirschl, MM
    Herkner, H
    Oschatz, E
    Kofler, J
    Woisetschläger, C
    Laggner, AN
    [J]. CRITICAL CARE MEDICINE, 2000, 28 (02) : 371 - 376
  • [7] Accuracy of waste blood measurement in critically ill patients
    Philippe R. Bauer
    Rahul Kashyap
    Roshini S. Abraham
    Tobias Peikert
    Ognjen Gajic
    [J]. Intensive Care Medicine, 2011, 37 : 721 - 722
  • [8] Accuracy of waste blood measurement in critically ill patients
    Bauer, Philippe R.
    Kashyap, Rahul
    Abraham, Roshini S.
    Peikert, Tobias
    Gajic, Ognjen
    [J]. INTENSIVE CARE MEDICINE, 2011, 37 (04) : 721 - 722
  • [9] Accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients
    Hirschl, MM
    Binder, M
    Herkner, H
    Bur, A
    Brunner, M
    Seidler, D
    Stuhlinger, HG
    Laggner, AN
    [J]. CRITICAL CARE MEDICINE, 1996, 24 (10) : 1684 - 1689
  • [10] Increasing the Accuracy of Oscillometric Blood Pressure Measurement
    Sutyagina, Anastasiya D.
    Shlyakhotka, Anna V.
    Lyakhova, Ekaterina K.
    [J]. PROCEEDINGS OF THE 2018 IEEE CONFERENCE OF RUSSIAN YOUNG RESEARCHERS IN ELECTRICAL AND ELECTRONIC ENGINEERING (EICONRUS), 2018, : 1231 - 1234