Determination of sub-nanomolar levels of iron in seawater using flow injection with chemiluminescence detection

被引:147
|
作者
Bowie, AR
Achterberg, EP
Mantoura, RFC
Worsfold, PJ
机构
[1] Univ Plymouth, Plymouth Environm Res Ctr, Dept Environm Sci, Plymouth PL4 8AA, Devon, England
[2] Plymouth Marine Lab, Ctr Coastal & Marine Sci, Plymouth PL1 3DH, Devon, England
关键词
flow injection; luminol; chemiluminescence; iron; preconcentration; seawater; Atlantic meridional transect;
D O I
10.1016/S0003-2670(98)00015-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The development of a highly sensitive system for the shipboard determination of dissolved iron at the sub-nM level is presented. The technique is based on a flow injection method coupled with luminol chemiluminescence detection. Dissolved Fe(II+III) levels are determined after Fe(III) reduction using sulphite and in-line matrix elimination/preconcentration on an 8-hydroxyguinoline (8-quinolinol) chelating resin column. The detection limit (3s) is 40 when 1.5 ml of sample is loaded onto the column, and the relative standard deviation is 3.2% (n=5) for a 1.0 nhl Fe sample. One analytical cycle can be completed in 3 min. The automated method proved reliable when employed on-board the RRS James Clark Ross during Autumn 1996. mapping dissolvable Fe(II+III) levels along the Atlantic Meridional Transect from 50 degrees N to 50 degrees S. Data from vertical profiles through the upper water column are presented. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:189 / 200
页数:12
相关论文
共 50 条
  • [21] Sub-nanomolar detection of cesium with water-gated transistor
    Alghamdi, Newel
    Alqahtani, Zahrah
    Grell, Martin
    [J]. JOURNAL OF APPLIED PHYSICS, 2019, 126 (06)
  • [22] Determination of total iron in fresh waters using flow injection with potassium permanganate chemiluminescence detection
    M. Yaqoob
    A. Waseem
    A. Nabi
    [J]. Journal of Analytical Chemistry, 2006, 61 : 917 - 921
  • [23] Determination of total iron in fresh waters using flow injection with potassium permanganate chemiluminescence detection
    Yaqoob, M.
    Waseem, A.
    Nabi, A.
    [J]. Journal of Analytical Chemistry, 2006, 61 (09): : 917 - 921
  • [24] DIRECT DETERMINATION OF SUB-NANOMOLAR LEVELS OF ZINC IN SEA-WATER BY CATHODIC STRIPPING VOLTAMMETRY
    VAN DEN BERG, CMG
    [J]. TALANTA, 1984, 31 (12) : 1069 - 1073
  • [25] Exclusive Detection of Sub-Nanomolar Levels of Palladium(II) in Water: An Excellent Probe for Multiple Applications
    Kumari, Namita
    Dey, Nilanjan
    Kumar, Krishan
    Bhattacharya, Santanu
    [J]. CHEMISTRY-AN ASIAN JOURNAL, 2014, 9 (11) : 3174 - 3181
  • [26] Sensitive determination of ketoprofen using flow injection with chemiluminescence detection
    Zhuang, Yafeng
    Song, Hengli
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 44 (03) : 824 - 828
  • [27] Determination of Mancozeb and Maneb using flow injection chemiluminescence detection
    Asghar, Muhammad
    Yaqoob, Mohammad
    Munawar, Nusrat
    Nabi, Abdul
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2022, 102 (11) : 2586 - 2606
  • [28] FLOW-INJECTION DETERMINATION OF IRON(II),IRON(III) AND TOTAL IRON WITH CHEMILUMINESCENCE DETECTION
    SARANTONIS, EG
    TOWNSHEND, A
    [J]. ANALYTICA CHIMICA ACTA, 1986, 184 : 311 - 315
  • [29] Determination of iron-porphyrin-like complexes at nanomolar levels in seawater
    Vong, Lilita
    Laes, Agathe
    Blain, Stephane
    [J]. ANALYTICA CHIMICA ACTA, 2007, 588 (02) : 237 - 244
  • [30] Comparison of backing materials of screen printed electrochemical sensors for direct determination of the sub-nanomolar concentration of lead in seawater
    Molinero-Abad, B.
    Izquierdo, D.
    Perez, L.
    Escudero, I.
    Arcos-Martinez, M. J.
    [J]. TALANTA, 2018, 182 : 549 - 557