A new intracoronary measurement catheter, MetriCath, compared to intravascular ultrasound and quantitative coronary angiography in a stented porcine coronary model

被引:5
|
作者
van der Giessen, WJ
Carlier, SG
Regar, E
van Beusekom, HMM
Foley, DP
de Feyter, PJ
Verdouw, PD
Boersma, E
Wolthuis, R
Serruys, PW
机构
[1] Erasmus MC, Thoraxctr, Dept Cardiol, NL-3000 CA Rotterdam, Netherlands
[2] Interuniv Cardiol Inst Netherlands, Utrecht, Netherlands
关键词
pressure volume; coronary dimensions; intravascular ultrasound; quantitative coronary angiography; methodology;
D O I
10.1002/ccd.10242
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to compare measurements by MetriCath to intravascular ultrasound (IVUS) and quantitative coronary angiography (QCA). The MetriCath system consists of a low-pressure (200 mm Hg) balloon catheter connected to a pressure transducer and infusion pump linked to a computer that records pressure-volume curves. Cross-sectional area of blood vessels is obtained directly from the unrestrained and in-stent pressure-volume measurements. We compared stent cross-sectional area measurements by MetriCath, IVUS, and QCA in a porcine stented coronary artery model. Comparison of area measurements in 14 stents showed no significant differences between the three methods (P = 0.66). On average, values differed 0.37 0.60 mm2 between MetriCath and QCA, 0.13 0.55 mm2 between MetriCath and IVUS, and 0.22 0.80 mm2 between IVUS and QCA. This corresponds to 6.2% 10%, 3.0% 9.0%, and 3.1% 12.9% relative difference from the average of two corresponding measurements. Linear regression analysis showed excellent correlation between measurements (r = 0.99 for all comparisons). The differences in in-stent area measurements between MetriCath and both QCA and IVUS were small. Considering the ease and rapidity of obtaining MetriCath results, this technique may form an alternative to the others in evaluating stent expansion. Based on these findings, clinical evaluation seems warranted. 2002 Wiley-Liss, Inc.
引用
收藏
页码:2 / 9
页数:8
相关论文
共 50 条
  • [1] Intravascular ultrasound and quantitative coronary angiography
    Takahashi, T
    Honda, Y
    Russo, RJ
    Fitzgerald, PJ
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2002, 55 (01) : 118 - 128
  • [2] RELATIONSHIP OF INTRACORONARY ULTRASOUND TO QUANTITATIVE CORONARY ANGIOGRAPHY IN CORONARY INTERVENTIONS
    LIEBERMAN, SM
    MOSES, JW
    MARINO, N
    MCCROSSAN, J
    COHEN, N
    GLEIM, GW
    RYKACESKI, W
    [J]. CLINICAL RESEARCH, 1993, 41 (02): : A344 - A344
  • [3] Overestimation of acute lumen gain and late lumen loss by quantitative coronary angiography (compared with intravascular ultrasound) in stented lesions
    Hoffmann, R
    Mintz, GS
    Popma, JJ
    Satler, LF
    Kent, KM
    Pichard, AD
    Leon, MB
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1997, 80 (10): : 1277 - 1281
  • [4] Limitations of coronary angiography compared with intravascular ultrasound: Implications for coronary interventions
    Yamashita, T
    Colombo, A
    Tobis, JM
    [J]. PROGRESS IN CARDIOVASCULAR DISEASES, 1999, 42 (02) : 91 - 138
  • [5] Comparison of quantitative coronary angiography, intravascular ultrasound, and coronary pressure measurement to assess optimum stent deployment
    Hanekamp, CEE
    Koolen, JJ
    Pijls, NHJ
    Michels, HR
    Bonnier, HJRM
    [J]. CIRCULATION, 1999, 99 (08) : 1015 - 1021
  • [6] Coronary Computer Tomography Angiography Quantitative Lesion Assessment Compared with Intravascular Ultrasound in Patients With Coronary Artery Disease
    Boccalandro, Fernando
    Dris, Lamia
    Angirekula, Manohar
    Amaram, Sudhir
    Lopez, Janie
    Romero, Araceli
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 56 (13) : B15 - B15
  • [7] CORONARY PLAQUE BURDEN BY QUANTITATIVE CORONARY ANGIOGRAPHY CORRELATES WITH PLAQUE BURDEN BY CT CORONARY ANGIOGRAPHY AND INTRAVASCULAR ULTRASOUND
    Kalynych, Anna M.
    Vazquez, Gustavo
    Karmpaliotis, Dimitri
    Qian, Zhen
    Krivitsky, Eric
    Marvasty, Idean B.
    Rinehart, Sarah
    Voros, Szilard
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (10)
  • [8] Videodensitometric quantitative angiography after coronary balloon angioplasty, compared to edge-detection quantitative angiography and intracoronary ultrasound imaging
    Peters, RJG
    Kok, WEM
    Pasterkamp, G
    von Birgelen, C
    Prins, M
    Serruys, PW
    [J]. EUROPEAN HEART JOURNAL, 2000, 21 (08) : 654 - 661
  • [9] Comparison of Intravascular Ultrasound and Two Dimensional Quantitative Coronary Angiography for the Measurement of Left Main Coronary Artery Diameter
    Bouzghaia, Mohamed Saad-Eddine
    Lamraoui, Chakib
    Lebcira, Khalil
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2023, 81 (16) : S38 - S39
  • [10] Quantification of coronary lesions by 64-slice computed tomography compared with quantitative coronary angiography and intravascular ultrasound - Reply
    Achenbach, S
    Daniel, WG
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (04) : 891 - 892