Characteristics of False-Positive Alarms in the BacT/Alert 3D System

被引:6
|
作者
Amano, Misato [1 ]
Matsumoto, Mami [2 ]
Sano, Shigeru [1 ]
Oyama, Mayumi [1 ]
Nagumo, Hideto [2 ]
Watanabe-Okochi, Naoko [1 ]
Tsuno, Nelson H. [1 ]
Nakajima, Kazunori [1 ]
Muroi, Kazuo [1 ]
机构
[1] Japanese Red Cross Soc, Kanto Koshinetsu Block Blood Ctr, Tokyo, Japan
[2] Japanese Red Cross Soc, Cent Blood Inst, Tokyo, Japan
来源
MICROBIOLOGY SPECTRUM | 2022年 / 10卷 / 03期
关键词
BacT; Alert; blood culture; false positive; BLOOD-CELL STORAGE; PLATELET COMPONENTS; BACTERIAL-CONTAMINATION; CULTURE;
D O I
10.1128/spectrum.00055-22
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
In clinical settings, false-positive results are treated as positive until bacterial identification. It may result in the discarding of blood products in blood centers or affect clinical decisions in hospitals or testing facilities. Moreover, the management of these samples is usually time- and labor-consuming. The results of our study may help clinicians and laboratory staff in making a more precise evaluation of positive reactions in BacT/Alert. The BacT/Alert system has been used for detecting the presence of bacteria in various clinical settings as well as in blood services, but it is associated with a relatively high incidence of false-positive results. We analyzed the results of our quality control sterility testing of blood products by BacT/Alert 3D to understand the mechanism of false-positive results. Anaerobic and aerobic bottles were inoculated with 10 mL of samples and cultured in BacT/Alert 3D for 10 days. Positive-reaction cases were classified as true positive if any bacterium was identified or false positive if the identification test had a negative result. The detection algorithm and the bottle graph pattern of the positive reaction cases were investigated. Among the 43,374 samples, 25 true positives (0.06%) and 29 false positives (0.07%) were observed. Although the detection algorithm of all true positives and 25 of 29 false positives was accelerating production of CO2, a steep rise in the bottle graph was observed only in the true positives, and it was not observed in either of the false positives. Four of 29 false positives were dependent on high baseline scatter reflections. Furthermore, evaluating the bottle graph pattern of Streptococcus pneumoniae, a bacterium known to autolyze, we confirmed that no viable bacterium was detected even if a steep rise was observed. In conclusion, the bottle graph pattern of positive reactions allows the differentiation between true positives and false positives. In case of a steep rise without bacterium detection, the bacterium might have autolyzed. Moreover, positive reactions with high baseline scatter reflections, despite immediate loading of bottles after sampling, are potentially false positive. IMPORTANCE In clinical settings, false-positive results are treated as positive until bacterial identification. It may result in the discarding of blood products in blood centers or affect clinical decisions in hospitals or testing facilities. Moreover, the management of these samples is usually time- and labor-consuming. The results of our study may help clinicians and laboratory staff in making a more precise evaluation of positive reactions in BacT/Alert.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Effect of delayed incubation of blood culture bottles in detecting bacterial growth with the Bact/Alert 3D system
    Ceballos, E
    Martínez-Martínez, L
    de Cueto, M
    Clavijo, MJ
    Pascual, A
    ENFERMEDADES INFECCIOSAS Y MICROBIOLOGIA CLINICA, 2001, 19 (07): : 350 - 351
  • [22] 2D CNN versus 3D CNN for false-positive reduction in lung cancer screening
    Yu, Juezhao
    Yang, Bohan
    Wang, Jing
    Leader, Joseph
    Wilson, David
    Pu, Jiantao
    JOURNAL OF MEDICAL IMAGING, 2020, 7 (05)
  • [23] Effectiveness of the lysis centrifugation method isolator 10 system compared to BacT/Alert 3D in detecting bacteraemia
    Kharazmi, K.
    Kratzer, C.
    Tucek, G.
    Klimpfinger, M.
    Georgopoulos, A.
    INTERNATIONAL JOURNAL OF ANTIMICROBIAL AGENTS, 2007, 29 : S246 - S246
  • [24] Evaluation of a novel kit for use with the BacT/ALERT 3D system for drug susceptibility testing of Mycobacterium tuberculosis
    Werngren, J
    Klintz, L
    Hoffner, SE
    JOURNAL OF CLINICAL MICROBIOLOGY, 2006, 44 (06) : 2130 - 2132
  • [26] Bact/Alert 3D全自动血培养仪使用评价
    陈日荣
    李惠冰
    实用医技杂志, 2005, (12) : 1605 - 1606
  • [27] Bact/Alert 3D系统与常用检测方法的比较
    周辉
    张治国
    杜春英
    吴进凤
    临床肺科杂志, 2009, 14 (07) : 902 - 903
  • [28] BacT/Alert 3D血培养仪阳性报警时间分析
    刘莹
    中国现代药物应用, 2016, 10 (05) : 274 - 275
  • [29] Remote Monitoring of Implantable Loop Recorders: False-Positive Alert Episode Burden
    O'Shea, Catherine J.
    Middeldorp, Melissa E.
    Hendriks, Jeroen M.
    Brooks, Anthony G.
    Harper, Curtis
    Thomas, Gijo
    Emami, Mehrdad
    Thiyagarajah, Anand
    Feigofsky, Suzanne
    Gopinathannair, Rakesh
    Varma, Niraj
    Campbell, Kevin
    Lau, Dennis H.
    Sanders, Prashanthan
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2021, 14 (11): : 1048 - 1050
  • [30] Evaluation of a new generation culture bottle with the BacT/ALERT® 3D culture system with 8 contaminating organisms in platelets
    Brecher, ME
    Heath, D
    Hay, S
    Rothenberg, S
    Stutzman, LC
    TRANSFUSION, 2001, 41 (09) : 33S - 33S