Assessment of anterior chamber angle using Visante OCT, slit-lamp OCT, and Pentacam

被引:18
|
作者
Dinc, Umut A. [1 ]
Oncel, Banu [1 ]
Gorgun, Ebru [1 ]
Yalvac, Ilgaz S. [1 ]
机构
[1] Yeditepe Univ Eye Hosp, Dept Ophthalmol, Istanbul, Turkey
关键词
Anterior chamber angle; Pentacam; Slit-lamp OCT; Visante OCT; OPTICAL COHERENCE TOMOGRAPHY; SEGMENT; REPRODUCIBILITY; INTRAOBSERVER;
D O I
10.1177/112067211002000327
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To assess and compare the anterior chamber angle (ACA) by different anterior segment imaging techniques. Methods. Forty healthy eyes of 40 normal subjects were recruited and 3 consecutive measurements of ACA were determined prospectively utilizing Visante optical coherence tomography (OCT), slit-lamp OCT (SL-OCT), and Pentacam. Statistical significance of interdevice differences between measurements was evaluated by Wilcoxon signed rank test and Bland-Altman analysis. The repeatability of 3 consecutive measurements was analyzed by repeated measured analysis of variance. Results. Mean nasal and temporal ACA were detected as 31.2+/-8.7 and 32.1+/-8.5 degrees by Visante OCT; 41.6+/-5.9 and 41.8+/-5.2 degrees by Pentacam; and 42.9+/-10.0 and 43.3+/-10.1 degrees by SLOCT. All devices displayed a high intrasession repeatability (Scheffe multiple comparison, p>0.05). ACA detected by Visante OCT were found significantly different from ACA calculated by Pentacam and SL-OCT (p<0.05). On the other hand, ACA data obtained by Pentacam and SL-OCT were found to be statistically similar (p>0.05). Conclusions. Although noncontact ACA measurements using all modalities were easy to handle and demonstrated good repeatability in healthy participants, the tested devices were not regarded as comparable. Hence, the clinician should take the different modalities into consideration during ACA assessment using various devices. (Eur J Ophthalmol 2010; 20: 531-7)
引用
收藏
页码:531 / 537
页数:7
相关论文
共 50 条
  • [1] Quantitative assessment of anterior chamber volume using slit-lamp OCT and Pentacam
    Dinc, Umut Asli
    Oncel, Banu
    Gorgun, Ebru
    Alimgil, Levent
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2009, 19 (03) : 411 - 415
  • [2] Anterior chamber angle measurement with anterior segment optical coherence tomography: A comparison between slit lamp OCT and Visante OCT
    Leung, Christopher Kai-Shun
    Li, Haitao
    Weinreb, Robert Neal
    Liu, Jing
    Cheung, Carol Yim Lui
    Lai, Ricky Yiu Kwong
    Pang, Chi Pui
    Lam, Dennis Shun Chiu
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (08) : 3469 - 3474
  • [3] Comparison of Visante and slit-lamp anterior segment optical coherence tomography in imaging the anterior chamber angle
    Sakata, L. M.
    Wong, T. T. L.
    Wong, H-T
    Kumar, R. S.
    Htoon, H. M.
    Aung, H. T.
    He, M.
    Aung, T.
    [J]. EYE, 2010, 24 (04) : 578 - 587
  • [4] Comparison of Visante and slit-lamp anterior segment optical coherence tomography in imaging the anterior chamber angle
    L M Sakata
    T T L Wong
    H-T Wong
    R S Kumar
    H M Htoon
    H T Aung
    M He
    T Aung
    [J]. Eye, 2010, 24 : 578 - 587
  • [5] Central Corneal Thickness, Anterior Chamber Depth, and Pupil Diameter Measurements Using Visante OCT, Orbscan, and Pentacam
    Yazici, Ahmet Taylan
    Bozkurt, Ercument
    Alagoz, Cengiz
    Alagoz, Nese
    Pekel, Gokhan
    Kaya, Vedat
    Yilmaz, Omer Faruk
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2010, 26 (02) : 127 - 133
  • [6] Comparison of Gonioscopy with Cirrus and Visante Optical Coherence Tomography (OCT) for Anterior Chamber Angle Assessment in Glaucoma Patients
    Hu, Cindy
    Zangalli, Camila
    Mantravadi, Anand
    Ali, Mohsin
    Faria, Bruno
    Richman, Jesse
    Wizov, Sheryl
    Razeghinejad, Reza
    Moster, Marlene
    Katz, L. Jay
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [7] Visante® OCT and the Ahmed valve:: from the anterior chamber to the posterior chamber
    Hamard, P.
    El Maftouhi, A.
    Baudouin, C.
    [J]. JOURNAL FRANCAIS D OPHTALMOLOGIE, 2007, 30 (10): : 1089 - 1091
  • [8] SLIT-LAMP MEASUREMENT OF THE ANTERIOR-CHAMBER DEPTH
    DOUTHWAITE, WA
    SPENCE, D
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 1986, 70 (03) : 205 - 208
  • [9] ANTERIOR-CHAMBER DEPTH MEASUREMENT USING THE SLIT-LAMP MICROSCOPE
    JACOBS, IH
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1979, 88 (02) : 236 - 238
  • [10] Comparing Remote Assessment of Limbal Anterior Chamber Depth Using Novel Stereoscopic Digital Slit Lamp Images with Anterior Segment OCT
    Cronin, Thomas
    Gupta, Ankur
    Abbas, Kinza
    Gupta, Akash
    Armstrong, Grayson
    Olson, Joanna
    Ingraham, Herbert
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)