Anterior chamber angle measurement with anterior segment optical coherence tomography: A comparison between slit lamp OCT and Visante OCT

被引:91
|
作者
Leung, Christopher Kai-Shun [1 ,2 ]
Li, Haitao [1 ]
Weinreb, Robert Neal [2 ]
Liu, Jing [1 ]
Cheung, Carol Yim Lui [1 ]
Lai, Ricky Yiu Kwong [1 ]
Pang, Chi Pui [1 ]
Lam, Dennis Shun Chiu [1 ]
机构
[1] Chinese Univ Hong Kong, Dept Ophthalmol & Visual Sci, Hong Kong, Hong Kong, Peoples R China
[2] Univ Calif San Diego, Dept Ophthalmol, Hamilton Glaucoma Ctr, San Diego, CA 92103 USA
关键词
D O I
10.1167/iovs.07-1477
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To compare anterior chamber angle measurements obtained from two anterior segment optical coherence tomography (OCT) instruments and to evaluate their agreements and interobserver reproducibility. METHODS. Forty-nine eyes from 49 healthy normal subjects were studied. The anterior chamber angle was imaged with the Visante anterior segment OCT (Carl Zeiss Meditec, Dublin, CA) and the slit lamp OCT (SLOCT, Heidelberg Engineering, GmbH, Dossenheim, Germany) on one randomly selected eye in each subject and measured by two independent observers. The angle-opening distance (AOD 500), the trabecular-iris angle (TIA 500), and the trabecular-iris space area (TISA 500) at the nasal and temporal angles were measured. The agreements between SLOCT and Visante OCT measurements and the interobserver reproducibility were evaluated. RESULTS. The mean nasal/temporal anterior chamber angles measured by Visante OCT and SLOCT were 527 +/- 249/572 +/- 275 mu m (AOD), 0.180 +/- 0.091/0.193 +/- 0.102 mm(2) (TISA), and 38.1 +/- 12.3/39.6 +/- 13.2 degrees (TIA); and 534 +/- 234/628 +/- 254 mu m (AOD), 0.191 +/- 0.089/0.217 +/- 0.093 mm(2) (TISA), and 37.8 +/- 10.1/40.6 +/- 10.7 degrees (TIA), respectively. No significant difference was found between Visante OCT and SLOCT measurements except the temporal TISA (P = 0.034). The interobserver coefficient of variation ranged between 4.4% and 7.8% for Visante OCT and 4.9% and 7.0% for SLOCT. The spans of 95% limits of agreement of the nasal/temporal angle measurements between Visante OCT and SLOCT were 437/531 mm(2), 0.174/0.186 mm(2), and 25.3/28.0 degrees for AOD, TISA, and TIA, respectively. CONCLUSIONS. Although Visante OCT and SLOCT demonstrate high interobserver reproducibility for anterior chamber angle measurements, their agreement was poor.
引用
收藏
页码:3469 / 3474
页数:6
相关论文
共 50 条
  • [1] Comparison of Visante and slit-lamp anterior segment optical coherence tomography in imaging the anterior chamber angle
    Sakata, L. M.
    Wong, T. T. L.
    Wong, H-T
    Kumar, R. S.
    Htoon, H. M.
    Aung, H. T.
    He, M.
    Aung, T.
    [J]. EYE, 2010, 24 (04) : 578 - 587
  • [2] Comparison of Visante and slit-lamp anterior segment optical coherence tomography in imaging the anterior chamber angle
    L M Sakata
    T T L Wong
    H-T Wong
    R S Kumar
    H M Htoon
    H T Aung
    M He
    T Aung
    [J]. Eye, 2010, 24 : 578 - 587
  • [3] Assessment of anterior chamber angle using Visante OCT, slit-lamp OCT, and Pentacam
    Dinc, Umut A.
    Oncel, Banu
    Gorgun, Ebru
    Yalvac, Ilgaz S.
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2010, 20 (03) : 531 - 537
  • [4] Comparison of Gonioscopy with Cirrus and Visante Optical Coherence Tomography (OCT) for Anterior Chamber Angle Assessment in Glaucoma Patients
    Hu, Cindy
    Zangalli, Camila
    Mantravadi, Anand
    Ali, Mohsin
    Faria, Bruno
    Richman, Jesse
    Wizov, Sheryl
    Razeghinejad, Reza
    Moster, Marlene
    Katz, L. Jay
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [5] Evaluation of the Anterior Segment Angle-to-Angle Scan of Cirrus High-Definition Optical Coherence Tomography and Comparison With Gonioscopy and With the Visante OCT
    Tun, Tin A.
    Baskaran, Mani
    Tan, Shayne S.
    Perera, Shamira A.
    Aung, Tin
    Husain, Rahat
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (01) : 59 - 64
  • [6] Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomography
    Li, Haitao
    Leung, Christopher Kai Shun
    Cheung, Carol Yim Lui
    Wong, Lee
    Pang, Chi Pui
    Weinreb, Robert Neal
    Lam, Dennis Shun Chiu
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (11) : 1490 - 1492
  • [7] Fourier Domain Optical Coherence Tomography integrated into a slit lamp; a novel technique combining anterior and posterior segment OCT
    Stehouwer, M.
    Verbraak, F. D.
    de Vries, H.
    Kok, P. H. B.
    van Leeuwen, T. G.
    [J]. EYE, 2010, 24 (06) : 980 - 984
  • [8] Fourier Domain Optical Coherence Tomography integrated into a slit lamp; a novel technique combining anterior and posterior segment OCT
    M Stehouwer
    F D Verbraak
    H de Vries
    P H B Kok
    T G van Leeuwen
    [J]. Eye, 2010, 24 : 980 - 984
  • [9] Anterior Segment Optical Coherence Tomography (AS-OCT) Visualization of Anterior Vitritis
    Zicarelli, Federico
    Staurenghi, Giovanni
    Invernizzi, Alessandro
    [J]. OCULAR IMMUNOLOGY AND INFLAMMATION, 2023, 31 (05) : 1101 - 1102
  • [10] Slit-lamp-adapted optical coherence tomography of the anterior segment
    H. Hoerauf
    C. Wirbelauer
    C. Scholz
    R. Engelhardt
    P. Koch
    H. Laqua
    R. Birngruber
    [J]. Graefe's Archive for Clinical and Experimental Ophthalmology, 2000, 238 : 8 - 18