Anthropomorphic phantom measurements for the validation of a treatment planning system

被引:16
|
作者
Dunscombe, P
McGhee, P
Lederer, E
机构
[1] LAURENTIAN UNIV,DEPT PHYS,SUDBURY,ON P3E 2C6,CANADA
[2] UNIV OTTAWA,FAC MED,OTTAWA,ON K1H 8L6,CANADA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1996年 / 41卷 / 03期
关键词
D O I
10.1088/0031-9155/41/3/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A series of eight irradiations of an anthropomorphic phantom has been performed as part of a programme of treatment planning system (TPS) validation. The 'treatment' configurations used represent a realistic cross section of those carried our routinely in a radiotherapy centre. The dose distributions within the phantom were determined using 75 LiF TLD chips per irradiation and the measured dose distributions were compared with calculations performed on a TPS for three ranges of dose gradient. The accuracy of the technique per measurement point is estimated to be 3% in dose (low dose gradient) and 3 mm in position (high dose gradient). The measurement and data analysis techniques used permit evaluation of a TPS against recently recommended criteria of acceptability although it is noted that is is nor possible to isolate TPS performance from factors depending on, for example, open beam data fitting and treatment set up accuracy.
引用
收藏
页码:399 / 411
页数:13
相关论文
共 50 条
  • [21] Dose Measurements Outside the Treatment Field From a VMAT Technique Using an Anthropomorphic Phantom-Preliminary Data
    Roa, D. E.
    Lin, Y.
    Kuo, J.
    Al-Ghazi, M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S799 - S799
  • [22] Evaluation of the TomoTherapy planning station heterogeneity correction algorithm using an anthropomorphic phantom
    Followill, D
    Molineu, A
    McGary, J
    Hernandez, N
    Ibbott, G
    MEDICAL PHYSICS, 2005, 32 (06) : 2167 - 2167
  • [23] Validation of a Novel Vector Method for Blood Peak Velocity Detection in an Anthropomorphic Phantom
    Matera, R.
    Ricci, S.
    Yu, A. C. H.
    Yiu, B. Y. S.
    Tortoli, P.
    2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [24] Validation of spine SBRT using a 3D printed Anthropomorphic phantom
    Dekker, K. H.
    Alexander, K. M.
    de Moraes, F. Ynoe
    Olding, T.
    11TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, IC3DDOSE, 2022, 2167
  • [25] Phantom developments and measurements for detecting the influence of the heterogeneous boron distribution on the treatment planning
    Daquino, GG
    Salvadori, P
    Moss, R
    Menichetti, L
    Pisani, P
    Morrissey, J
    RESEARCH AND DEVELOPMENT IN NEUTRON CAPTURE THERAPY, 2002, : 623 - 627
  • [27] AN ANTHROPOMORPHIC PHANTOM FOR PERIODONTAL RADIOGRAPHY
    VISSER, H
    HERMANN, KP
    PFEIFFER, T
    UMLAUF, J
    KRUGER, W
    HARDER, D
    DENTOMAXILLOFACIAL RADIOLOGY, 1995, 24 (03) : 195 - 197
  • [28] USES OF AN ANTHROPOMORPHIC CHEST PHANTOM
    GILLAN, GD
    PEARCE, JG
    ROECK, WW
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1980, 134 (01) : 204 - 204
  • [29] IMPROVED CARDIAC ANTHROPOMORPHIC PHANTOM
    SMITH, SW
    COMBS, MP
    ADAMS, DB
    KISSLO, JA
    JOURNAL OF ULTRASOUND IN MEDICINE, 1994, 13 (08) : 601 - 625
  • [30] Prostate dosimetry in an anthropomorphic phantom
    Yu P.K.N.
    Cheung T.
    Butson M.J.
    Australasian Physics & Engineering Sciences in Medicine, 2004, 27 (2): : 60 - 62