Long-term in vivo wear of different bearing types used for the Oxford Unicompartmental Knee Replacement

被引:11
|
作者
Mohammad, H. R. [1 ]
Campi, S. [1 ]
Kennedy, J. A. [1 ]
Judge, A. [1 ]
Murray, D. W. [1 ]
Mellon, S. J. [1 ]
机构
[1] Univ Oxford, Nuffield Dept Orthopaed Rheumatol & Musculoskelet, Oxford, England
来源
BONE & JOINT RESEARCH | 2019年 / 8卷 / 11期
关键词
Mobile bearing unicompartmental knee replacement; Polyethylene; Long-term wear; CROSS-LINKED POLYETHYLENE; MODEL-BASED RSA; ARTHROPLASTY; RETRIEVAL; CREEP; HIP;
D O I
10.1302/2046-3758.811.BJR-2019-0163.R1
中图分类号
Q813 [细胞工程];
学科分类号
摘要
Objectives The aim of this study was to determine the polyethylene wear rate of Phase 3 Oxford Unicompartmental Knee Replacement bearings and to investigate the effects of resin type and manufacturing process. Methods A total of 63 patients with at least ten years' follow-up with three bearing types (1900 resin machined, 1050 resin machined, and 1050 resin moulded) were recruited. Patients underwent full weight-bearing model-based radiostereometric analysis to determine the bearing thickness. The linear wear rate was estimated from the change in thickness divided by the duration of implantation. Results The wear rate for 1900 resin machined (n = 19), 1050 machined (n = 21), and 1050 moulded bearings (n = 23) were 60 mu m/year (sd 42), 76 mu m/year (sd 32), and 57 mu m/year (sd 30), respectively. There was no significant difference between 1900 machined and 1050 machined (p = 0.20), but 1050 moulded had significantly less wear than the 1050 machined (p = 0.05). Increasing femoral (p < 0.001) and tibial (p < 0.001) component size were associated with increasing wear. Conclusion Wear rate is similar with 1050 and 1900 resin, but lower with moulded bearings than machined bearings. The currently used Phase 3 bearings wear rate is low (1050 moulded, 57 mu m/year), but higher than the previously reported Phase 2 bearings (1900 moulded, 20 mu m/year). This is unlikely to be due to the change in polyethylene but may relate to the minimally invasive approach used with the Phase 3. This approach, as well as improving function and thus increasing activity levels, may increase the risk of surgical errors, such as impingement or bearing overhang, which can increase wear. Surgeons should aim to use 4 mm thick bearings rather than 3 mm thick bearings in young patients, unless they are small and need conservative bone resections.
引用
收藏
页码:535 / 543
页数:9
相关论文
共 50 条
  • [31] Long-term Results of an Unloader Brace in Patients With Unicompartmental Knee Osteoarthritis
    Wilson, Becky
    Rankin, Heather
    Barnes, C. Lowry
    ORTHOPEDICS, 2011, 34 (08) : 575 - 575
  • [32] Extent of in vivo sagittal bearing movement and its relationship with tibial posterior slopes in Oxford mobile-bearing unicompartmental knee arthroplasty
    Kamenaga, Tomoyuki
    Hida, Yuichi
    Hiranaka, Takafumi
    Tanaka, Toshikazu
    Okimura, Kenjiro
    Tsubosaka, Masanori
    Kuroda, Yuichi
    Nakano, Naoki
    Hayashi, Shinya
    Niikura, Takahiro
    Kuroda, Ryosuke
    Matsumoto, Tomoyuki
    CLINICAL BIOMECHANICS, 2020, 80 (80)
  • [33] A LONG-TERM FOLLOW-UP-STUDY OF UNICOMPARTMENTAL REPLACEMENT WITH A MARMOR KNEE PROSTHESIS AND THE INTRODUCTION OF A NEW CERAMIC PROSTHESIS
    TATEISHI, H
    MARUOKA, T
    YOH, K
    IWATA, Y
    FUTANI, H
    YAMADA, H
    MARUO, S
    BULLETIN OF THE HOSPITAL FOR JOINT DISEASES ORTHOPAEDIC INSTITUTE, 1991, 51 (02): : 132 - 139
  • [34] The Mid- to Long-Term Outcomes of the Lateral Domed Oxford Unicompartmental Knee Replacement: An Analysis From the National Joint Registry for England, Wales, Northern Ireland, and the Isle of Man
    Mohammad, Hasan R.
    Matharu, Gulraj S.
    Judge, Andrew
    Murray, David W.
    JOURNAL OF ARTHROPLASTY, 2021, 36 (01): : 107 - 111
  • [35] LONG-TERM RESULTS OF GEOMEDIC TOTAL KNEE REPLACEMENT
    RILEY, D
    WOODYARD, JE
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1985, 67 (04): : 548 - 550
  • [36] The influence of surgical factors on dislocation of the meniscal bearing after Oxford medial unicompartmental knee replacement A CASE-CONTROL STUDY
    Lee, S. Y.
    Bae, J. H.
    Kim, J. G.
    Jang, K. M.
    Shon, W. Y.
    Kim, K. W.
    Lim, H. C.
    BONE & JOINT JOURNAL, 2014, 96B (07): : 914 - 922
  • [37] Good mid- to long-term results of the cemented oxford phase 3 unicompartmental knee arthroplasty in a non-designer centre
    D. M. Moore
    G. A. Sheridan
    A. Welch-Phillips
    J. M. O’Byrne
    P. Kenny
    Knee Surgery, Sports Traumatology, Arthroscopy, 2022, 30 : 3215 - 3219
  • [38] Comparison of long-term numerical and experimental total knee replacement wear during simulated gait loading
    Knight, Lucy A.
    Pal, Saikat
    Coleman, John C.
    Bronson, Fred
    Haider, Hani
    Levine, Danny L.
    Taylor, Mark
    Rullkoetter, Paul J.
    JOURNAL OF BIOMECHANICS, 2007, 40 (07) : 1550 - 1558
  • [39] Good mid- to long-term results of the cemented oxford phase 3 unicompartmental knee arthroplasty in a non-designer centre
    Moore, D. M.
    Sheridan, G. A.
    Welch-Phillips, A.
    O'Byrne, J. M.
    Kenny, P.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2022, 30 (09) : 3215 - 3219
  • [40] Long- term results of lateral unicompartmental knee arthroplasty with a mobile- bearing device
    Walker, T.
    Freericks, J.
    Mick, P.
    Trefzer, R.
    Lunz, A.
    Koch, K. -A
    Renkawitz, T.
    Hariri, M.
    BONE & JOINT JOURNAL, 2025, 107B (03): : 322 - 328