Survey of subrange inconsistency of long-stem standard platinum resistance thermometers

被引:6
|
作者
Peruzzi, A. [1 ]
Rusby, R. L. [2 ]
Pearce, J., V [2 ]
Eusebio, L. [3 ]
Bojkovski, J. [4 ]
Zuzek, V [4 ]
机构
[1] CNR, NRC, Ottawa, ON, Canada
[2] Natl Phys Lab, NPL, Teddington, Middx, England
[3] Portuguese Inst Qual, IPQ, Caparica, Portugal
[4] Univ Ljubljana, UL, Ljubljana, Slovenia
基金
欧盟地平线“2020”;
关键词
subrange inconsistency of the ITS-90; non-uniqueness of the ITS-90; long-stem standard platinum resistance thermometers;
D O I
10.1088/1681-7575/abe8c1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The subrange inconsistency (SRI) of a large ensemble of long-stem standard platinum resistance thermometers (SPRTs), representative of worldwide production, has been investigated for all pairs of ITS-90 overlapping subranges between 83.8058 K and 933.473 K. The results are reported in terms of various statistical parameters to facilitate their comparison with the results of different authors, although a statistical test, applied to one specific pair of subranges, supported a Gaussian distribution of the results and justified the subsequent use of the mean and standard deviation as statistical parameters. Depending on the pair of overlapping subranges, the mean SRI as calculated varied from -1.23 mK to +0.21 mK and the SRI standard deviation varied from 0.04 mK to 0.62 mK. These numbers generally increased with the upper temperature limit of the pair of subranges, especially where the lower subrange requires a point which is not included in the upper subrange. The contribution to SRI from the fixed-point uncertainty propagation (PoU) was evaluated. The results showed that, although the effect of PoU on SRI largely cancels out for points common to both subranges, PoU still amounts to 59% to 130% of the differences between overlapping pairs of subranges. This means that the differences as calculated are probably a substantial overestimate of the true SRI. It is suggested that this effect is taken into account in making recommendations for typical uncertainties due to non-uniqueness.
引用
收藏
页数:23
相关论文
共 49 条
  • [41] Type 3 Non-uniqueness in Interpolations Using Standard Platinum Resistance Thermometers Between − 196 °C and 100 °C
    R. L. Rusby
    H. Stemp
    J. V. Pearce
    R. I. Veltcheva
    International Journal of Thermophysics, 2019, 40
  • [42] Type 3 Non-uniqueness in Interpolations Using Standard Platinum Resistance Thermometers Between-196 °C and 100 °C
    Rusby, R. L.
    Stemp, H.
    Pearce, J., V
    Veltcheva, R., I
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2019, 40 (11)
  • [43] Comparison of long-stem and standard tube-stock performance 5 years after planting in a rainforest edge on the Central Coast of NSW
    Chalmers, Anita
    Bakewell, Geoff
    Taggart, Audrey
    ECOLOGICAL MANAGEMENT & RESTORATION, 2015, 16 (03) : 236 - 238
  • [44] SIM.T-K9.1: Canada/Peru bilateral comparison of standard platinum resistance thermometers from 0.01 °C to 419.527 °C
    Hill, K. D.
    Barrientos, A.
    Guillen, E.
    METROLOGIA, 2015, 52
  • [45] CCT-K2: key comparison of capsule-type standard platinum resistance thermometers from 13.8 K to 273.16K
    Steele, AG
    Fellmuth, B
    Head, DL
    Hermier, Y
    Kang, KH
    Steur, PPM
    Tew, WL
    METROLOGIA, 2002, 39 (06) : 551 - 571
  • [46] Final report on CCT-K2.5: NRC/NMIJ/INRIM comparison of capsule-type standard platinum resistance thermometers from 13.8 K to 273.16 K
    Hill, K. D.
    Nakano, T.
    Steur, P.
    METROLOGIA, 2015, 52
  • [47] CCT-K2.3: NRC/NMi-VSL bilateral comparison of capsule-type standard platinum resistance thermometers from 13.8 K to 273.16 K
    Hill, K. D.
    Peruzzi, A.
    Bosma, R.
    METROLOGIA, 2012, 49
  • [48] CCT-K2.4: NRC/INTiBS/LNE-Cnam trilateral comparison of capsule-type standard platinum resistance thermometers from 13.8 K to 273.16 K
    Hill, K. D.
    Szmyrka-Grzebyk, A.
    Lipinski, L.
    Hermier, Y.
    Pitre, L.
    Sparasci, F.
    METROLOGIA, 2012, 49
  • [49] Subsequent Bilateral Comparison to CCT-K3 CIPM Key comparison CCT-K3.1 Comparison of Standard Platinum Resistance Thermometers at the Triple Point of Water (T=273.16 K) and at the Melting Point of Gallium (T=302.9146 K)
    Picard, S.
    Nonis, M.
    Solve, S.
    Allisy-Roberts, P. J.
    Renaot, E.
    Martin, C.
    METROLOGIA, 2011, 48