Design and initial performance of PlanTIS: a high-resolution positron emission tomograph for plants

被引:39
|
作者
Beer, S. [1 ]
Streun, M. [1 ]
Hombach, T. [2 ]
Buehler, J. [2 ]
Jahnke, S. [2 ]
Khodaverdi, M. [1 ]
Larue, H. [1 ]
Minwuyelet, S. [2 ]
Parl, C. [1 ]
Roeb, G. [2 ]
Schurr, U. [2 ]
Ziemons, K. [1 ]
机构
[1] Forschungszentrum Juelich, Cent Inst Elect, D-52425 Julich, Germany
[2] Forschungszentrum Juelich, ICG Phytosphere 3, D-52425 Julich, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2010年 / 55卷 / 03期
关键词
SMALL ANIMAL PET; TRANSPORT; PHLOEM; LUYAP;
D O I
10.1088/0031-9155/55/3/006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Positron emitters such as C-11, N-13 and F-18 and their labelled compounds are widely used in clinical diagnosis and animal studies, but can also be used to study metabolic and physiological functions in plants dynamically and in vivo. A very particular tracer molecule is (CO2)-C-11 since it can be applied to a leaf as a gas. We have developed a Plant Tomographic Imaging System (PlanTIS), a high-resolution PET scanner for plant studies. Detectors, front-end electronics and data acquisition architecture of the scanner are based on the ClearPET (TM) system. The detectors consist of LSO and LuYAP crystals in phoswich configuration which are coupled to position-sensitive photomultiplier tubes. Signals are continuously sampled by free running ADCs, and data are stored in a list mode format. The detectors are arranged in a horizontal plane to allow the plants to be measured in the natural upright position. Two groups of four detector modules stand face-to-face and rotate around the field-of-view. This special system geometry requires dedicated image reconstruction and normalization procedures. We present the initial performance of the detector system and first phantom and plant measurements.
引用
收藏
页码:635 / 646
页数:12
相关论文
共 50 条
  • [21] High-resolution positron emission mammography in breast cancer
    Cheng, ZH
    Keen, R
    Narayanan, D
    Sawyer, K
    Watkins, S
    Tafra, L
    [J]. ANNALS OF SURGICAL ONCOLOGY, 2006, 13 (02) : 20 - 20
  • [22] A prototype high-resolution animal positron tomograph with avalanche photodiode arrays and LSO crystals
    Sibylle I. Ziegler
    Bernd J. Pichler
    Guido Boening
    Magdalena Rafecas
    Wendelin Pimpl
    Eckart Lorenz
    Norbert Schmitz
    Markus Schwaiger
    [J]. European Journal of Nuclear Medicine, 2001, 28 : 136 - 143
  • [23] DESIGN OF A VOLUME-IMAGING POSITRON EMISSION TOMOGRAPH
    ROGERS, JG
    HARROP, R
    COOMBES, GH
    WILKINSON, NA
    ATKINS, MS
    PATE, BD
    MORRISON, KS
    STAZYK, M
    DYKSTRA, CJ
    BARNEY, JS
    DOHERTY, PW
    SAYLOR, DP
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) : 993 - 997
  • [24] A NEW APPROACH TO THE MEASUREMENT OF RESOLUTION AND SAMPLING ON A POSITRON EMISSION TOMOGRAPH
    ROBESON, W
    DHAWAN, V
    BABCHYCK, B
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (04) : 1506 - 1513
  • [25] DESIGN OF THE NEURO-ECAT - A HIGH-RESOLUTION, HIGH-EFFICIENCY POSITRON TOMOGRAPH FOR IMAGING THE ADULT HEAD OR INFANT TORSO
    WILLIAMS, CW
    CRABTREE, MC
    BURKE, MR
    KEYSER, RM
    BURGISS, SG
    HOFFMAN, EJ
    PHELPS, ME
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (02) : 1736 - 1740
  • [26] HIGH-RESOLUTION CIRCULAR RING POSITRON TOMOGRAPH WITH DICHOTOMIC SAMPLING - DICHOTOM-I
    CHO, ZH
    HILAL, SK
    RA, JB
    HONG, KS
    BIGLER, RE
    YOSHIZUMI, T
    WOLF, AP
    FOWLER, JS
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (11): : 1219 - 1234
  • [27] A POSITRON EMISSION TOMOGRAPH DESIGNED FOR 3/4 MM RESOLUTION
    MCINTYRE, JA
    ALLEN, RD
    AGUIAR, J
    PAULSON, JT
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) : 1102 - 1106
  • [28] PERFORMANCE-CHARACTERISTICS OF THE MCMASTER POSITRON EMISSION TOMOGRAPH
    NAHMIAS, C
    FIRNAU, G
    GARNETT, ES
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1984, 31 (01) : 637 - 639
  • [29] Detector design of a LSO based positron emission tomograph with depth of interaction capability for high resolution brain imaging.
    Schmand, M
    Eriksson, L
    Casey, ME
    Newport, D
    Wienhard, K
    Heiss, WD
    Nutt, R
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1998, 39 (05) : 133P - 133P
  • [30] ANALYTICAL STUDY OF PERFORMANCE OF HIGH-RESOLUTION POSITRON EMISSION COMPUTED TOMOGRAPHS FOR ANIMAL STUDY
    NOHARA, N
    TANAKA, E
    TOMITANI, T
    YAMAMOTO, M
    MURAYAMA, H
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) : 818 - 821