Design and validation of a MR-compatible pneumatic manipulandum

被引:17
|
作者
Suminski, Aaron J.
Zimbelman, Janice L.
Scheidt, Robert A.
机构
[1] Marquette Univ, Dept Biomed Engn, Olin Engn Ctr, Milwaukee, WI 53201 USA
[2] Med Coll Wisconsin, Dept Neurol, Milwaukee, WI 53226 USA
[3] Northwestern Univ, Dept Phys Med & Rehabil, Feinberg Sch Med, Chicago, IL USA
关键词
fMRI; robotics; human; motor control; wrist;
D O I
10.1016/j.jneumeth.2007.03.014
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The combination of functional MR imaging and novel robotic tools may provide unique opportunities to probe the neural systems underlying motor control and learning. Here, we describe the design and validation of a MR-compatible, I degree-of-freedom pneumatic manipulandum along with experiments demonstrating its safety and efficacy. We first validated the robot's ability to apply computer-controlled loads about the wrist, demonstrating that it possesses sufficient bandwidth to simulate torsional spring-like loads during point-to-point flexion movements. Next, we verified the MR-compatibility of the device by imaging a head phantom during robot operation. We observed no systematic differences in two measures of MRI signal quality (signal/noise and field homogeneity) when the robot was introduced into the scanner environment. Likewise, measurements of joint angle and actuator pressure were not adversely affected by scanning. Finally, we verified device efficacy by scanning 20 healthy human subjects performing rapid wrist flexions against a wide range of spring-like loads. We observed a linear relationship between joint torque at peak movement extent and perturbation magnitude, thus demonstrating the robot's ability to simulate spring-like loads in situ. fMRI revealed task-related activation in regions known to contribute to the control of movement including the left primary sensorimotor cortex and right cerebellum. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:255 / 266
页数:12
相关论文
共 50 条
  • [21] Non-linear adaptive controllers for an over-actuated pneumatic MR-compatible stepper
    Hollnagel, Christoph
    Vallery, Heike
    Schaedler, Rainer
    Lopez, Isaac Gomez-Lor
    Jaeger, Lukas
    Wolf, Peter
    Riener, Robert
    Marchal-Crespo, Laura
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2013, 51 (07) : 799 - 809
  • [22] Design and development of an MR-Compatible PET scanner for imaging small animals
    Mackewn, JE
    Strul, D
    Hallett, WA
    Halsted, P
    Page, RA
    Keevil, SF
    Williams, SCR
    Cherry, SR
    Marsden, PK
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3271 - 3274
  • [23] Design and kinematics analysis of a novel MR-compatible robot for needle insertion
    School of Mechanical Engineering, Center for Advanced Mechanisms and Robotics, Tianjin University, 300072 Tianjin, China
    不详
    Adv. Reconfigurable Mech. Rob., (443-453):
  • [24] Design of MR-compatible robotic devices: magnetic and geometric compatibility aspects
    Keroglou, Christoforos
    Tsekos, Nikolaos V.
    Seimenis, Ioannis
    Eracleous, Eleni
    Christodoulou, Christodoulos G.
    Pitris, Constantinos
    Christoforou, Eftychios G.
    2009 9TH INTERNATIONAL CONFERENCE ON INFORMATION TECHNOLOGY AND APPLICATIONS IN BIOMEDICINE, 2009, : 486 - +
  • [25] The case for MR-compatible robotics: a review of the state of the
    Elhawary, Haytham
    Tse, Zion Tsz Ho
    Hamed, Abbi
    Rea, Marc
    Davies, Brian L.
    Lamperth, Michael U.
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2008, 4 (02): : 105 - 113
  • [26] A low-cost, MR-compatible olfactometer
    Lowen, Steven B.
    Lukas, Scott E.
    BEHAVIOR RESEARCH METHODS, 2006, 38 (02) : 307 - 313
  • [27] Gamma shielding materials for MR-compatible PET
    Strul, D
    Cash, D
    Keevil, SF
    Halsted, P
    Williams, SCR
    Marsden, PK
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (01) : 60 - 69
  • [28] Dodecapus: An MR-compatible system for somatosensory stimulation
    Huang, Ruey-Song
    Sereno, Martin I.
    NEUROIMAGE, 2007, 34 (03) : 1060 - 1073
  • [29] Development of an MR-compatible optical force sensor
    Tada, M
    Kanade, T
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 2022 - 2025
  • [30] PRECISION POSITION TRACKING OF MR-COMPATIBLE PNEUMATIC PISTON-CYLINDER USING SLIDING MODE CONTROL
    Comber, David
    Barth, Eric J.
    PROCEEDINGS OF THE ASME DYNAMIC SYSTEMS AND CONTROL CONFERENCE AND BATH/ASME SYMPOSIUM ON FLUID POWER AND MOTION CONTROL (DSCC 2011), VOL 1, 2012, : 45 - 51