Validation of the mechanical systems for a tomotherapy Hi-Art unit

被引:0
|
作者
Pavord, D. [1 ]
机构
[1] W Penn Allegheny Hlth Syst, Pittsburgh, PA USA
关键词
D O I
10.1118/1.2760348
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To validate the accuracy of the table and gantry motion of the Tomotherapy Hi‐Art system and to measure the range of table deflection and to assess the impact on dose delivery. Method and Materials: The Tomotherapy Hi‐Art system delivers radiation dose to a target using a helical tomotherapy delivery. Because of this fact, the mechanical accuracy of the table speed and gantry rotation speed, which determine the pitch of the delivery, are extremely important. The amount of table sag relative to the treatment plan is also critical. Table speed and gantry speed were measured by attaching a marker with three reflective spheres to both the table top and gantry structure. The position of the markers was monitored with a Polaris infra‐red camera system. Measurements were taken for 4 treatments of varying pitch. The table top sag for 0, 150, and 300 pounds weight was measured both at the end of the table and at the isocenter for different table positions. The impact of this on a delivered plan was checked by measuring a plan using film dosimetry with 0, 150, and 300 pounds of added weight. Results: Table speed was verified to be accurate within 0.5%. The average gantry rotation period was found to be within 0.02 seconds of the planned value. The maximum sag at the end of the table ranged from 1.25 to 3.35 cm. The sag at isocenter ranged from 0.5 to 1.4 cm. The results translate to a maximum deflection angle of 1.1 degrees. The resulting films showed a maximum shift in isodose curves of 1 mm in the vertical direction. Conclusion: The mechanical systems on the Tomotherapy unit are capable of producing the planned pitch values with high accuracy. The table deflection while substantial does not significantly affect the accuracy of the delivery. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
下载
收藏
页码:2331 / 2331
页数:1
相关论文
共 50 条
  • [41] Hypofractionated Radiotherapy for the Patients with Unresectable Primary Hepatocellular Carcinoma using TomoTherapy Hi-Art: Analysis of the Efficacy and Toxicity
    Jang, H.
    Son, S.
    Song, J.
    Choi, B.
    Kang, Y.
    Yoon, S.
    Kang, K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S361 - S362
  • [42] Stereotactic image-guided intensity modulated radiotherapy using the HI-ART II helical tomotherapy system
    Holmes, Timothy W.
    Hudes, Richard
    Dziuba, Sylwester
    Kazi, Abdul
    Hall, Mark
    Dawson, Dana
    MEDICAL DOSIMETRY, 2008, 33 (02) : 135 - 148
  • [43] Image-guided upper abdominal radiotherapy with the Tomotherapy® HI-ART® treatment system:: Daily set-up verification
    Dahele, M.
    Skinner, M.
    Ling, Y.
    Lightstone, A.
    Hunter, M.
    Kirby, R.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S66 - S67
  • [44] GEANT4 simulation of dose deposition in patients from TomoTherapy Hi-Art Megavoltage computed tomography (MVCT) imaging
    Brochu, F.
    Burnet, N.
    Jena, R.
    Thomas, S.
    RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S15 - S16
  • [45] Evaluation of mid-treatment tumor motion for hypo-fractionated lung radiosurgery using hi-art TomoTherapy system
    Jang, S.
    Wu, H.
    Sourivong, P.
    Katz, S.
    Rosen, L.
    MEDICAL PHYSICS, 2007, 34 (06) : 2451 - 2451
  • [46] Comparison of Hi-Art Tomotherapy Machine Outputs Using AAPM TG-148 and IAEA TRS 483 Codes of Practice
    Firmansyah, O. A.
    Firmansyah, A. F.
    Sunaryati, S., I
    Wibowo, W. E.
    2019 CONFERENCE ON FUNDAMENTAL AND APPLIED SCIENCE FOR ADVANCED TECHNOLOGY, 2019, 1373
  • [47] Impact of intra-fractional tumor motion for hypo-fractionated lung radiosurgery using Hi-Art TomoTherapy system
    Wu, H.
    Sourivong, P.
    Jang, S.
    Katz, S.
    Massey, J.
    Rosen, L.
    MEDICAL PHYSICS, 2007, 34 (06) : 2457 - 2457
  • [48] An independent dose validation technique for TomoTherapy Hi*Art II Tomo IMRT treatments
    Poole, D
    Kupelian, P
    Langen, K
    McNerney, G
    Willoughby, T
    Zhang, Y
    MEDICAL PHYSICS, 2004, 31 (06) : 1862 - 1862
  • [49] Stereotactic body radiotherapy using helical tomotherapy for hepatocellular carcinoma: clinical outcome and dosimetric comparison of Hi-ART vs. Radixact
    Bae, Sun Hyun
    Cho, Kwang Hwan
    Jung, Jae Hong
    Kim, Young Seok
    Kim, Sang Gyune
    Yoo, Jeong-Ju
    Lee, Jae Myung
    Lee, Min Hee
    Lim, Sung Hee
    TRANSLATIONAL CANCER RESEARCH, 2022, 11 (11) : 3964 - 3973
  • [50] Surface dose for five telecobalt machines, 6MV photon beam from four linear accelerators and a Hi-Art TomoTherapy
    Kinhikar, Rajesh A.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2008, 7 (05) : 381 - 384