Validation of the mechanical systems for a tomotherapy Hi-Art unit

被引:0
|
作者
Pavord, D. [1 ]
机构
[1] W Penn Allegheny Hlth Syst, Pittsburgh, PA USA
关键词
D O I
10.1118/1.2760348
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To validate the accuracy of the table and gantry motion of the Tomotherapy Hi‐Art system and to measure the range of table deflection and to assess the impact on dose delivery. Method and Materials: The Tomotherapy Hi‐Art system delivers radiation dose to a target using a helical tomotherapy delivery. Because of this fact, the mechanical accuracy of the table speed and gantry rotation speed, which determine the pitch of the delivery, are extremely important. The amount of table sag relative to the treatment plan is also critical. Table speed and gantry speed were measured by attaching a marker with three reflective spheres to both the table top and gantry structure. The position of the markers was monitored with a Polaris infra‐red camera system. Measurements were taken for 4 treatments of varying pitch. The table top sag for 0, 150, and 300 pounds weight was measured both at the end of the table and at the isocenter for different table positions. The impact of this on a delivered plan was checked by measuring a plan using film dosimetry with 0, 150, and 300 pounds of added weight. Results: Table speed was verified to be accurate within 0.5%. The average gantry rotation period was found to be within 0.02 seconds of the planned value. The maximum sag at the end of the table ranged from 1.25 to 3.35 cm. The sag at isocenter ranged from 0.5 to 1.4 cm. The results translate to a maximum deflection angle of 1.1 degrees. The resulting films showed a maximum shift in isodose curves of 1 mm in the vertical direction. Conclusion: The mechanical systems on the Tomotherapy unit are capable of producing the planned pitch values with high accuracy. The table deflection while substantial does not significantly affect the accuracy of the delivery. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2331 / 2331
页数:1
相关论文
共 50 条
  • [1] TomoTherapy HI-ART heterogeneity calculations
    McGary, J
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1989 - 1989
  • [2] Dosimetric validation of first helical tomotherapy Hi-Art II machine in India
    Kinhikar, Rajesh A.
    Jamema, Swamidas V.
    Reenadevi
    Pai, Rajeshri
    Zubin, Master
    Gupta, Tejpal
    Dhote, Deepak S.
    Deshpande, Deepak D.
    Shrivastava, Shyam K.
    Sarin, Rajiv
    [J]. JOURNAL OF MEDICAL PHYSICS, 2009, 34 (01) : 23 - 30
  • [3] Stereotactic radiosurgery using the TomoTherapy HI-ART II
    Holmes, T. W.
    Hu, X.
    Dawson, D.
    Hoban, P.
    Kazi, A.
    Hall, M.
    Dziuba, S. J.
    Hudes, R. S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S255 - S255
  • [4] Tomotherapy: Comparison of Hi-ART, Tomo-HD, and Radixact
    Kurosaki, Hiromasa
    Hirayama, Kenta
    Takahashi, Masaki
    Uematsu, Masahiro
    Tate, Etsuko
    [J]. CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (10)
  • [5] A comprehensive patient-specific IMRT quality assurance procedure on hi-art Tomotherapy® unit
    Yan, Y.
    Jiang, H.
    Weng, X.
    Penagaricano, J.
    Moros, E.
    Papanikolaou, N.
    Shi, C.
    Novak, P.
    Ratanatharathorn, V.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2250 - 2250
  • [6] Stereotactic cranial irradiation with HI-ART tomotherapy: A feasibility study
    Neybuch, Y.
    Vaandering, A.
    Tomsej, M.
    D'Hoore, W.
    Renard, L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S295 - S296
  • [7] The feasibility of using tomotherapy Hi-art machine for stereotactic radiosurgery
    Jaradat, H
    Tome, W
    Mehta, M
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2045 - 2046
  • [8] Adapting the Victoreen 7200(™) device as a tool for daily QA of the helical tomotherapy Hi-Art(™) Unit
    Detorie, NA
    Teslow, T
    Schultz, J
    Bell, R
    Ridenour, V
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2089 - 2089
  • [9] Technical Note: Output and energy fluctuations of the tomotherapy Hi-Art helical tomotherapy system
    Mahan, SL
    Chase, DJ
    Ramsey, CR
    [J]. MEDICAL PHYSICS, 2004, 31 (07) : 2119 - 2120
  • [10] To make two or more tomotherapy Hi-Art machines dosimetric twins
    Althof, V.
    Wispelweij, M.
    Westendorp, R.
    Minken, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S197 - S198