Mathematical method to calculate full-energy peak efficiency of detectors based on transfer technique

被引:32
|
作者
Gouda, M. M. [1 ]
Hamzawy, A. [2 ]
Badawi, M. S. [1 ]
El-Khatib, A. M. [1 ]
Thabet, A. A. [3 ]
Abbas, M. I. [1 ]
机构
[1] Univ Alexandria, Fac Sci, Dept Phys, Alexandria 21511, Egypt
[2] Umm Al Qura Univ, Univ Coll, Dept Phys, Makkah Al Mokrama, Saudi Arabia
[3] Pharos Univ Alexandria, Fac Allied Med Sci, Dept Med Equipment Technol, Alexandria, Egypt
关键词
High-purity germanium well-type detector; Cylindrical sources; Self-attenuation; Coincidence summing effect; Efficiency transfer method; COINCIDENCE SUMMING CORRECTIONS; GAMMA-RAY SPECTROMETRY; WELL-TYPE DETECTORS; SELF-ATTENUATION; NAI TL; COMPUTATION; CALIBRATE;
D O I
10.1007/s12648-015-0737-1
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The full-energy peak efficiency of high-purity germanium well-type detector is extremely important to calculate the absolute activities of natural and artificial radionuclides for samples with low radioactivity. In this work, the efficiency transfer method in an integral form is proposed to calculate the full-energy peak efficiency and to correct the coincidence summing effect for a high-purity germanium well-type detector. This technique is based on the calculation of the ratio of the effective solid angles subtended by the well-type detector with cylindrical sources measured inside detector cavity and an axial point source measured out the detector cavity including the attenuation of the photon by the absorber system. This technique can be easily applied in establishing the efficiency calibration curves of well-type detectors. The calculated values of the efficiency are in good agreement with the experimental calibration data obtained with a mixed gamma-ray standard source containing Co-60 and Y-88.
引用
收藏
页码:201 / 210
页数:10
相关论文
共 50 条
  • [1] Mathematical method to calculate full-energy peak efficiency of detectors based on transfer technique
    M. M. Gouda
    A. Hamzawy
    M. S. Badawi
    A. M. El-Khatib
    A. A. Thabet
    M. I. Abbas
    Indian Journal of Physics, 2016, 90 : 201 - 210
  • [2] Direct mathematical method for calculating full-energy peak efficiency and coincidence corrections of HPGe detectors for extended sources
    Abbas, Mahmoud I.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2007, 256 (01): : 554 - 557
  • [3] A method for establishing absolute full-energy peak efficiency and its confidence interval for HPGe detectors
    Rizwan, U.
    Chester, A.
    Domingo, T.
    Starosta, K.
    Williams, J.
    Voss, P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 802 : 102 - 112
  • [4] An empirical formula to calculate the full energy peak efficiency of scintillation detectors
    Badawi, Mohamed S.
    Abd-Elzaher, Mohamed
    Thabet, Abouzeid A.
    El-khatib, Ahmed M.
    APPLIED RADIATION AND ISOTOPES, 2013, 74 : 46 - 49
  • [5] A numerical approach to calculate the full-energy peak efficiency of HPGe well-type detectors using the effective solid angle ratio
    Badawi, M. S.
    Ruskov, I.
    Gouda, M. M.
    El-Khatib, A. M.
    Alotiby, M. F.
    Mohamed, M. M.
    Thabet, A. A.
    Abbas, M. I.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [6] A NEW MATHEMATICAL MODEL FOR DETERMINING FULL-ENERGY PEAK EFFICIENCY OF AN ARRAY OF TWO GAMMA DETECTORS COUNTING RECTANGULAR PARALLELEPIPED SOURCES
    Badawi, Mohamed S.
    Krar, Mohamed E.
    El-Khatib, Ahmed M.
    Jovanovic, Slobodan I.
    Dlabac, Aleksandar D.
    Mihaljevic, Nikola N.
    NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 2013, 28 (04): : 370 - 380
  • [7] Full-energy peak efficiency and response function of 1 cm3 CdZnTe detectors
    Tajudin, Suffian Mohamad
    Namito, Yoshihito
    Sanami, Toshiya
    Hirayama, Hideo
    MALAYSIAN JOURNAL OF FUNDAMENTAL AND APPLIED SCIENCES, 2019, 15 (04): : 580 - 584
  • [8] Analytical expression to calculate total and full-energy peak efficiencies for cylindrical phoswich and lanthanum bromide scintillation detectors
    Abbas, Mahmoud I.
    Noureddeen, Salam
    RADIATION MEASUREMENTS, 2011, 46 (04) : 440 - 445
  • [9] A NOVEL METHOD FOR COMPUTING THE FULL-ENERGY PEAK EFFICIENCY OF COAXIAL HIGH-PURITY GERMANIUM DETECTORS FOR CYLINDRICAL SOURCES
    Gurau, D.
    Stanga, D.
    Done, L.
    Sima, O.
    Ilie, G.
    ROMANIAN REPORTS IN PHYSICS, 2024, 76 (04)
  • [10] FULL-ENERGY PEAK EFFICIENCY OF ASYMMETRICAL POLYHEDRON GERMANIUM DETECTOR
    Abbas, Mahmoud I.
    Yoseph, Sara
    El-Khatib, Ahmed M.
    Badawi, Mohamed S.
    Gouda, Mona M.
    Thabet, Abouzeid A.
    NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 2018, 33 (02): : 150 - 158